The effect of multiple cycles of contamination, detergent washing, and disinfection on the development of biofilm in endoscope tubing

被引:43
作者
Vickery, Karen [1 ,2 ]
Ngo, Quan-D. [3 ]
Zou, Jean [3 ]
Cossart, Yvonne E. [3 ]
机构
[1] Liverpool Hosp, Dept Plast & Reconstruct Surg, Sydney, NSW, Australia
[2] Univ New S Wales, Surg Infect Res Grp, S Western Clin Sch, Sydney, NSW, Australia
[3] Univ Sydney, Dept Immunol & Infect Dis, Sydney, NSW 2006, Australia
关键词
GASTROINTESTINAL ENDOSCOPES; FLEXIBLE ENDOSCOPES; WASHER-DISINFECTOR; DECONTAMINATION; STERILIZATION; EFFICACY; FAILURE;
D O I
10.1016/j.ajic.2008.09.016
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Patient soil and bacterial biofilm in patient-ready endoscope channels can adversely impact the efficacy of detergent and disinfectant. thereby increasing the risk of nosocomial infection. Biofilm bacteria are firmly attached to one another and to the substrate by exopolysaccharide, making them difficult to remove. We analyzed the effect of 20 wash/contamination cycles on biofilm formation. Materials: Pseudomonas aeruginosa biofilm-covered endoscope tubing was soaked in water (control), an enzymatic cleaner, or a nonenzymatic cleaner (Matrix) for 10 minutes and decontaminated in a washer-disinfector machine. Media containing P aeruginosa was then recycled to simulate contamination in clinical practice. Results: SEM analysis showed that loosely attached biofilm was removed under the high flow rates in the washer-disinfector. The control tubing remained 100% covered with biofilm, which became thicker with increased recontamination cycles. Washing in the enzymatic detergent retarded the redevelopment of biofilm. The nonenzymatic cleaner (Matrix) continued to remove more biofilm with an increasing number of wash/contamination cycles. At the 20th cycle, 90% of the tubing was biofilm-free. Conclusion: Washing endoscopes under high now rates with some detergents removes established biofilm and retards biofilm generation, emphasizing the importance of cleaning before disinfection. Continued research into the physicochemical mechanisms of biofilm adherence and removal is needed to optimize detergents. Copyright (C) 2009 by the Association for Professionals in Infection Control and Epidemiology, Inc. (Am J Infect Control 2009;37:470-5.)
引用
收藏
页码:470 / 475
页数:6
相关论文
共 20 条
[1]   Automated washing with the Reliance Endoscope Processing System and its equivalence to optimal manual cleaning [J].
Alfa, Michelle J. ;
Olson, Nancy ;
DeGagne, Pat .
AMERICAN JOURNAL OF INFECTION CONTROL, 2006, 34 (09) :561-570
[2]   Worst-case soiling levels for patient-used flexible endoscopes before and after cleaning [J].
Alfa, MJ ;
Degagne, P ;
Olson, N .
AMERICAN JOURNAL OF INFECTION CONTROL, 1999, 27 (05) :392-401
[3]   A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure [J].
Bisset, Linda ;
Cossart, Yvonne E. ;
Selby, Warwick ;
West, Richard ;
Catterson, Denise ;
O'Hara, Kate ;
Vickery, Karen .
AMERICAN JOURNAL OF INFECTION CONTROL, 2006, 34 (05) :274-280
[4]  
Blob R, 2004, J HOSP INFECT, V56, P544
[5]   Evaluation of disinfection and sterilization of reusable angioscopes with the duck hepatitis B model [J].
Chaufour, X ;
Deva, AK ;
Vickery, K ;
Zou, J ;
Kumaradeva, P ;
White, GH ;
Cossart, YE .
JOURNAL OF VASCULAR SURGERY, 1999, 30 (02) :277-282
[6]   Introduction to biofilm [J].
Costerton, JW .
INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 1999, 11 (3-4) :217-221
[7]  
Das JR, 1998, J APPL MICROBIOL, V84, P852
[8]   Detection of persistent vegetative bacteria and amplified viral nucleic acid from in-use testing of gastrointestinal endoscopes [J].
Deva, AK ;
Vickery, K ;
Zou, J ;
West, RH ;
Selby, W ;
Benn, RAV ;
Harris, JP ;
Cossart, YE .
JOURNAL OF HOSPITAL INFECTION, 1998, 39 (02) :149-157
[9]   Endoscope decontamination incidents in England 2003-2004 [J].
Gamble, H. P. ;
Duckworth, G. J. ;
Ridgway, G. L. .
JOURNAL OF HOSPITAL INFECTION, 2007, 67 (04) :350-354
[10]   ELIMINATION OF HIGH TITER HIV FROM FIBEROPTIC ENDOSCOPES [J].
HANSON, PJV ;
GOR, D ;
JEFFRIES, DJ ;
COLLINS, JV .
GUT, 1990, 31 (06) :657-659