Simulation of long-term fatigue damage in bioprosthetic heart valves: effects of leaflet and stent elastic properties

被引:59
作者
Martin, Caitlin [1 ,2 ]
Sun, Wei [1 ,2 ,3 ]
机构
[1] Univ Connecticut, Tissue Mech Lab, Biomed Engn Program, Storrs, CT 06269 USA
[2] Univ Connecticut, Dept Mech Engn, Storrs, CT 06269 USA
[3] Georgia Inst Technol, Wallace H Coulter Dept Biomed Engn, Atlanta, GA 30313 USA
关键词
Bioprosthetic heart valve durability; Soft tissue fatigue; Finite element analysis; SHILEY PERICARDIAL XENOGRAFTS; CARPENTIER-EDWARDS; FOLLOW-UP; CONSTITUTIVE MODEL; MITRAL POSITION; PERMANENT SET; TISSUE; FAILURE; DETERIORATION; DEGENERATION;
D O I
10.1007/s10237-013-0532-x
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
One of the major failure modes of bioprosthetic heart valves (BHVs) is noncalcific structural deterioration due to fatigue of the tissue leaflets; yet, the mechanisms of fatigue are not well understood. BHV durability is primarily assessed based on visual inspection of the leaflets following accelerated wear testing. In this study, we developed a computational framework to simulate BHV leaflet fatigue, which is both efficient and quantitative, making it an attractive alternative to traditional accelerated wear testing. We utilize a phenomenological soft tissue fatigue damage model developed previously to describe the stress softening and permanent set of the glutaraldehyde-treated bovine pericardium leaflets in BHVs subjected to cyclic loading. A parametric study was conducted to determine the effects of altered leaflet and stent elastic properties on the fatigue of the leaflets. The simulation results show that heterogeneity of the leaflet elastic properties, poor leaflet coaptation, and little stent-tip deflection may accelerate leaflet fatigue, which agrees with clinical findings. Therefore, the developed framework may be an invaluable tool for evaluating leaflet durability in new tissue valve designs, including traditional BHVs as well as new transcatheter valves.
引用
收藏
页码:759 / 770
页数:12
相关论文
共 40 条
[21]   COMPARATIVE-STUDY OF PRIMARY TISSUE VALVE FAILURE BETWEEN IONESCU-SHILEY PERICARDIAL AND HANCOCK PORCINE VALVES IN THE AORTIC POSITION [J].
NISTAL, F ;
GARCIASATUE, E ;
ARTINANO, E ;
DURAN, CMG ;
GALLO, I .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (01) :161-164
[22]   VALVE FAILURE WITH THE IONESCU-SHILEY BOVINE PERICARDIAL BIOPROSTHESIS - ANALYSIS OF 2680 PATIENTS [J].
REUL, GJ ;
COOLEY, DA ;
DUNCAN, JM ;
FRAZIER, OH ;
HALLMAN, GL ;
LIVESAY, JJ ;
OTT, DA ;
WALKER, WE .
JOURNAL OF VASCULAR SURGERY, 1985, 2 (01) :192-204
[23]  
Schoen F., 1991, Cardiovascular pathology, V2nd, P1547
[24]  
Schoen FJ, 1999, J BIOMED MATER RES, V47, P439, DOI 10.1002/(SICI)1097-4636(19991215)47:4<439::AID-JBM1>3.0.CO
[25]  
2-O
[26]   CAUSES OF FAILURE AND PATHOLOGICAL FINDINGS IN SURGICALLY REMOVED IONESCU-SHILEY STANDARD BOVINE PERICARDIAL HEART-VALVE BIOPROSTHESES - EMPHASIS ON PROGRESSIVE STRUCTURAL DETERIORATION [J].
SCHOEN, FJ ;
FERNANDEZ, J ;
GONZALEZLAVIN, L ;
CERNAIANU, A .
CIRCULATION, 1987, 76 (03) :618-627
[27]   ON A FULLY 3-DIMENSIONAL FINITE-STRAIN VISCOELASTIC DAMAGE MODEL - FORMULATION AND COMPUTATIONAL ASPECTS [J].
SIMO, JC .
COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 1987, 60 (02) :153-173
[28]  
Singh P. J., 2013, ISRN BIOMATER, P1
[29]  
Smith DB, 1999, J HEART VALVE DIS, V8, P25
[30]   Finite element implementation of a generalized Fung-elastic constitutive model for planar soft tissues [J].
Sun, W ;
Sacks, MS .
BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2005, 4 (2-3) :190-199