Spatial resolution of microwave tomography for detection of myocardial ischemia and infarction-experimental study on two-dimensional models

被引:42
作者
Semenov, SY [1 ]
Svenson, RH
Bulyshev, AE
Souvorov, AE
Nazarov, AG
Sizov, YE
Posukh, VG
Pavlovsky, AV
Repin, PN
Tatsis, GP
机构
[1] Carolinas Med Ctr, Carolinas Heart Inst, Laser & Appl Technol Lab, Charlotte, NC 28203 USA
[2] IV Kurchatov Atom Energy Inst, Biophys Res Lab, Moscow 123182, Russia
[3] Inst Laser Phys, Novosibirsk, Russia
关键词
microwave tomography; myocardial infarction; spatial resolution;
D O I
10.1109/22.842025
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, the experimental study of spatial resolution of microwave tomography was performed. Our microwave tomographic system with operational frequencies of 0.9 and 2.36 GHz and with signal-to-noise ratio of 30 dB allowed us to achieve a spatial resolution between 7.3-9.5 mm and 6.3-7.8 mm at the former and latter frequencies, respectively, It was shown in experiments, with structurally complicated objects, that the spatial resolution of about the same distances can be expected in a practical application of microwave tomography to detect areas of myocardial ischemia and infarction.
引用
收藏
页码:538 / 544
页数:7
相关论文
共 12 条
[1]  
[Anonymous], 1986, Medical Applications of Microwave Imaging
[2]  
BOLOMEY JC, 1992, MALVERN PHY, P319
[3]   Experimental verification of super resolution in nonlinear inverse scattering [J].
Chen, FC ;
Chew, WC .
APPLIED PHYSICS LETTERS, 1998, 72 (23) :3080-3082
[4]   A COMPUTER-SIMULATION STUDY OF DIFFRACTION TOMOGRAPHY [J].
DEVANEY, AJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (07) :377-386
[5]  
DEVANEY AJ, 1992, MALVERN PHY, P47
[6]   Microwave imaging for tissue assessment: Initial evaluation in multitarget tissue-equivalent phantoms [J].
Meaney, PM ;
Paulsen, KD ;
Hartov, A ;
Crane, RK .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (09) :878-890
[7]  
SEMENOV S, 1997, P IEEE 19 ENG MED BI
[8]   Three-dimensional microwave tomography: Experimental prototype of the system and vector born reconstruction method [J].
Semenov, SY ;
Svenson, RH ;
Bulyshev, AE ;
Souvorov, AE ;
Nazarov, AG ;
Sizov, YE ;
Pavlovsky, AV ;
Borisov, VY ;
Voinov, BA ;
Simonova, GI ;
Starostin, AN ;
Posukh, VG ;
Tatsis, GP ;
Baranov, VY .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (08) :937-946
[9]   Microwave tomography: Theoretical and experimental investigation of the iteration reconstruction algorithm [J].
Semenov, SY ;
Bulyshev, AE ;
Souvorov, AE ;
Svenson, RH ;
Sizov, YE ;
Borisov, VY ;
Posukh, VG ;
Kozlov, IM ;
Nazarov, AG ;
Tatsis, GP .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1998, 46 (02) :133-141
[10]   Microwave tomography: Two-dimensional system for biological imaging [J].
Semenov, SY ;
Svenson, RH ;
Boulyshev, AE ;
Souvorov, AE ;
Borisov, VY ;
Sizov, Y ;
Starostin, AN ;
Dezern, KR ;
Tatsis, GP ;
Baranov, VY .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (09) :869-877