A novel automated discontinuous venous blood monitoring system for ex vivo glucose determination in humans

被引:6
作者
Schaller, R. [1 ]
Feichtner, F. [1 ]
Koehler, H. [1 ]
Bodenlenz, M. [1 ]
Plank, J. [2 ]
Wutte, A. [2 ]
Mader, J. K. [2 ]
Ellmerer, M. [2 ]
Hellmich, R.
Wedig, H.
Hainisch, R.
Pieber, T. R. [2 ]
Schaupp, L. [1 ,2 ]
机构
[1] Joanneum Res Forsch Gesell mbH, Inst Med Technol & Hlth Management, A-8010 Graz, Austria
[2] Med Univ Graz, Dept Internal Med, Div Endocrinol & Nucl Med, Graz, Austria
关键词
Glucose; Sensors; Monitoring; Blood; Automated; Ex vivo; CLINICAL-EVALUATION; ACCURACY;
D O I
10.1016/j.bios.2008.11.029
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
intensive insulin therapy reduces mortality and morbidity in critically ill patients but imposes great demands on medical staff who must take frequent blood samples for the determination of glucose levels. A solution to this resourcing problem would be provided by an automated blood monitoring system. The aim of the present clinical study was to evaluate such a system comprising an automatic blood sampling unit linked to a glucose biosensor. Our approach was to determine the correlation and system error of the sampling unit alone and of the combined system with respect to reference levels over 12 h in humans. Two venous cannulae were inserted to connect the automatic and reference systems to the subjects. Blood samples were taken at 15 and 30 min intervals. The median Pearson coefficient of correlation between manually and automatically withdrawn blood samples was 0.982 for the sampling unit alone and 0.950 for the complete system. The biosensor had a linear range up to 20 mmol l(-1) and a 95% response, time of <2 min. Clark Error Grid analysis showed that 96.93% of the data (228 data pairs) was in zone A and 3.07% in zone B. Insulin Titration Error Grid analysis suggested an acceptable treatment in 99.56% of cases. Implementation of a "Keep Vein Open" saline infusion into the automated blood sampling system reduced blood withdrawal failures through occluded catheters fourfold. in summary, automated blood sampling from a peripheral vein coupled with automatic glucose determination is a promising alternative to frequent manual blood sampling. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:2239 / 2245
页数:7
相关论文
共 20 条
[1]   Interaction of DAP3 and FADD only after cellular disruption [J].
Berger, T ;
Kretzler, M .
NATURE IMMUNOLOGY, 2002, 3 (01) :3-4
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   First human experiments with a novel non-invasive, non-optical continuous glucose monitoring system [J].
Caduff, A ;
Hirt, E ;
Feldman, Y ;
Ali, Z ;
Heinemann, L .
BIOSENSORS & BIOELECTRONICS, 2003, 19 (03) :209-217
[4]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[5]   Clinical evaluation of alternative-site glucose measurements in patients after major cardiac surgery [J].
Ellmerer, M ;
Haluzik, M ;
Blaha, J ;
Kremen, J ;
Svacina, S ;
Toller, W ;
Mader, J ;
Schaupp, L ;
Plank, J ;
Pieber, T .
DIABETES CARE, 2006, 29 (06) :1275-1281
[6]  
Ganesh Arjunan, 2008, J Diabetes Sci Technol, V2, P182
[7]   Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor [J].
Garg, S ;
Zisser, H ;
Schwartz, S ;
Bailey, T ;
Kaplan, R ;
Ellis, S ;
Jovanovic, L .
DIABETES CARE, 2006, 29 (01) :44-50
[8]   Novel system for real-time ex vivo lactate monitoring in human whole blood [J].
Gfrerer, RJ ;
Brunner, GA ;
Trajanoski, Z ;
Schaupp, L ;
Sendlhofer, G ;
Skrabal, F ;
Jobst, G ;
Moser, I ;
Urban, G ;
Pieber, TR ;
Wach, P .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (12) :1271-1278
[9]   Bedside monitoring of subcutaneous interstitial glucose in healthy individuals using microdialysis and infrared spectrometry [J].
Heise, H. Michael ;
Damm, Uwe ;
Bodenlenz, Manfred ;
Kondepati, Venkata Radhakrishna ;
Koehler, Gerd ;
Ellmerer, Martin .
JOURNAL OF BIOMEDICAL OPTICS, 2007, 12 (02)
[10]   Nonlinear model predictive control of glucose concentration in subjects with type 1 diabetes [J].
Hovorka, R ;
Canonico, V ;
Chassin, LJ ;
Haueter, U ;
Massi-Benedetti, M ;
Federici, MO ;
Pieber, TR ;
Schaller, HC ;
Schaupp, L ;
Vering, T ;
Wilinska, ME .
PHYSIOLOGICAL MEASUREMENT, 2004, 25 (04) :905-920