Study of standard thermometers and their reproduction of the λ point in liquid helium

被引:3
作者
Astrov, DN
Korostin, SV
机构
关键词
Liquid Helium; Drift Rate; Phase Transition Point; Sensor Coil; Radio Engineer Measurement;
D O I
10.1007/BF02512093
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Self-calibrated resistance thermometers to be used as standards have been investigated. The heat exchange of the sensor coil with liquid helium per unit coil surface area at 0.3 K is 12 times better than that of a conventional gas-filled thermometer. The exchange is due to Kapitza surface thermal resistance. The lambda point in liquid helium filling of the thermometer is reproducible:to within 0.1 mK.