GeB flat fibre TL dosimeters for in-vivo measurements in radiosurgery

被引:3
作者
Alyahyawi, Amjad [1 ,2 ]
Dimitriadis, Alexis [3 ]
Nisbet, Andrew [4 ]
Bradley, D. A. [2 ,5 ]
机构
[1] Univ Hail, Dept Diagnost Radiol, Coll Appl Med Sci, Hail, Saudi Arabia
[2] Univ Surrey, Dept Phys, Guildford GU2 7XH, Surrey, England
[3] UCLH, Natl Hosp Neurol & Neurosurg, Queen Sq Radiosurg Ctr, London, England
[4] UCL, Dept Med Phys & Biomed Engn, London, England
[5] Sunway Univ, Ctr Biomed Phys, Jalan Univ, Subang Jaya 47500, Malaysia
关键词
Glass dosimeters; Stereotactic radiosurgery; GeB flat Fibres; DOPED OPTICAL-FIBERS; THERMOLUMINESCENCE RESPONSE; DOSE-RATE; PHOTON; RADIATION; VERIFICATION;
D O I
10.1016/j.radphyschem.2020.108973
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Noting an increase in demand for procedures involving clinical radiosurgery we seek to develop a high spatial resolution thermoluminescence dosimeter (TLD) to allow conduct of in vivo dose verification measurements. An associated need is for a dynamic dose range exceeding that of the well-established LiF (Mg,Ti) phosphor TLD-100, with in particular the latter being limited in performance at the elevated doses seen in radiotherapy. The work investigates the performance of a novel GeB co-doped Flat Fibre (GeB-FF) fabricated using the modified chemical vapour deposition (MCVD) process, the hollow capillary optical fibres (COF) produced from this being collapsed down into flat fibres (FF) to create strain-related defects. This process has already been demonstrated to increase the low dose sensitivity of optical fibres, notably at diagnostic x-ray potentials, with Minimum Detectable Dose (MDD) values of down to 0.1 mu Gy. The intent of present work, conducted as a component of a safety audit, part of the hospital periodic radiation protection quality assurance program, has been to examine and compare the performance of the two forms of TL dosimeter, GeB-FF and TLD-100, measuring scattered radiation resulting from cranial cavity radiosurgery procedures. The dosimeters were placed on the neck, chest and pelvis of 20 patients. Using both types of dosimeter, raw dose values at each site show general accord ( +/- 3 mGy at 1 sigma), covering mean doses ranging from some 10 mGy to less than 1 mGy, representing doses of < 1% to < 0.1% of prescribed dose at the treatment site. GeB-FF results uncorrected for energy response show absorbed doses greater than that using TLD-100, by factors of some 1.4, 1.2 and 1.5 for the pelvis, chest and neck respectively; energy corrections have been shown elsewhere to provide for much closer agreement.
引用
收藏
页数:4
相关论文
共 24 条
[1]  
Alyahyawi A., 2019, THESIS
[2]   Thermoluminescence measurements of eye-lens dose in a multi-centre stereotactic radiosurgery audit [J].
Alyahyawi, Amjad ;
Dimitriadis, A. ;
Jafari, S. M. ;
Lohstroh, A. ;
Alanazi, A. ;
Alsubaie, A. ;
Clark, C. H. ;
Nisbet, A. ;
Bradley, D. A. .
RADIATION PHYSICS AND CHEMISTRY, 2019, 155 :75-81
[3]   Tailor-made Ge-doped silica-glass for clinical diagnostic X-ray dosimetry [J].
Alyahyawi, Amjad ;
Jupp, T. ;
Alkhorayef, M. ;
Bradley, D. A. .
APPLIED RADIATION AND ISOTOPES, 2018, 138 :45-49
[4]   Comparison of thermoluminescence response of different sized Ge-doped flat fibers as a dosimeter [J].
Begum, Mahfuza ;
Rahman, A. K. M. Mizanur ;
Abdul-Rashid, H. A. ;
Yusoff, Z. ;
Mat-Sharif, K. A. ;
Zulkifli, M. I. ;
Muhamad-Yasin, S. Z. ;
Ung, N. M. ;
Kadir, A. B. A. ;
Amin, Y. M. ;
Bradley, D. A. .
RADIATION PHYSICS AND CHEMISTRY, 2015, 116 :155-159
[5]  
Dambul K. D., 2012, 2012 Photonics Global Conference (PGC 2012), DOI 10.1109/PGC.2012.6458091
[6]   Sensitive Fibre-Based Thermoluminescence Detectors for High Resolution In-Vivo Dosimetry [J].
Ghomeishi, Mostafa ;
Mahdiraji, G. Amouzad ;
Adikan, F. R. Mahamd ;
Ung, N. M. ;
Bradley, D. A. .
SCIENTIFIC REPORTS, 2015, 5
[7]   The thermoluminescence response of oxygen-doped optical fibres subjected to photon and electron irradiations [J].
Hashim, S. ;
Bradley, D. A. ;
Peng, N. ;
Ramli, A. T. ;
Wagiran, H. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3) :291-294
[8]   The thermoluminescence response of doped SiO2 optical fibres subjected to fast neutrons [J].
Hashim, S. ;
Bradley, D. A. ;
Saripan, M. I. ;
Ramli, A. T. ;
Wagiran, H. .
APPLIED RADIATION AND ISOTOPES, 2010, 68 (4-5) :700-703
[9]   Establishment of Ge-doped optical fibres as thermoluminescence dosimeters for brachytherapy [J].
Issa, Fatma ;
Rahman, A. T. Abdul ;
Hugtenburg, Richard P. ;
Bradley, David A. ;
Nisbet, Andrew .
APPLIED RADIATION AND ISOTOPES, 2012, 70 (07) :1158-1161
[10]   Small-field radiotherapy photon beam output evaluation: Detectors reviewed [J].
Lam, S. E. ;
Bradley, D. A. ;
Khandaker, M. U. .
RADIATION PHYSICS AND CHEMISTRY, 2021, 178