Numerical calculation of relative dose rates from spherical 106Ru beta sources used in ophthalmic brachytherapy

被引:6
作者
de Paiva, Eduardo [1 ]
机构
[1] Inst Radioprotecao & Dosimetria IRD CNEN MCTI, Div Fish Med, Ave Salvador Allende S-N, BR-22783127 Rio De Janeiro, RJ, Brazil
关键词
Brachytherapy; Ophthalmic applicator; Ru-106/Rh-106; DISTRIBUTIONS; RADIOTHERAPY;
D O I
10.1016/j.rinp.2015.02.006
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Concave beta sources of Ru-106/Rh-106 are used in radiotherapy to treat ophthalmic tumors. However, a problem that arises is the difficult determination of absorbed dose distributions around such sources mainly because of the small range of the electrons and the steep dose gradients. In this sense, numerical methods have been developed to calculate the dose distributions around the beta applicators. In this work a simple code in Fortran language is developed to estimate the dose rates along the central axis of Ru-106/Rh-106 curved plaques by numerical integration of the beta point source function and results are compared with other calculated data. (C) 2015 The Author. Published by Elsevier B.V. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
引用
收藏
页码:76 / 77
页数:2
相关论文
共 7 条
[1]  
[Anonymous], RAD DOSIMETRY, DOI 10.1016/B978-1-4832-3257-7.50024-X
[2]  
Barbosa N.A., 2014, INT J CANC THER ONCO, V2, P1, DOI [10.14319/ijcto.0203.8, DOI 10.14319/IJCTO.0203.8]
[3]   Calculation of dose distributions for 12 106Ru/106Rh ophthalmic applicator models with the PENELOPE Monte Carlo code [J].
Hermida-Lopez, M. .
MEDICAL PHYSICS, 2013, 40 (10)
[4]   Theoretical calculations of dose distributions for beta-ray eye applicators [J].
Hokkanen, J ;
Heikkonen, J ;
Holmberg, P .
MEDICAL PHYSICS, 1997, 24 (02) :211-213
[5]   106Ru/106Rh Plaque and proton radiotherapy for ocular melanoma:: A comparative dosimetric study [J].
Mourtada, F ;
Koch, N ;
Newhauser, W .
RADIATION PROTECTION DOSIMETRY, 2005, 116 (1-4) :454-460
[6]   Monte Carlo calculation of the dose distributions of two 106Ru eye applicators [J].
Sánchez-Reyes, A ;
Tello, JJ ;
Guix, B ;
Salvat, F .
RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) :191-196
[7]   DOSIMETRY OF BETA SOURCES IN RADIOTHERAPY .1. THE BETA POINT-SOURCE DOSE FUNCTION [J].
VYNCKIER, S ;
WAMBERSIE, A .