Feasibility of Pump Speed Modulation for Restoring Vascular Pulsatility with Rotary Blood Pumps

被引:37
|
作者
Ising, Mickey S. [1 ]
Sobieski, Michael A. [2 ]
Slaughter, Mark S. [2 ,3 ]
Koenig, Steven C. [2 ,3 ]
Giridharan, Guruprasad A. [3 ]
机构
[1] Univ Louisville, Sch Med, Louisville, KY 40202 USA
[2] Univ Louisville, Dept Cardiovasc & Thorac Surg, Louisville, KY 40202 USA
[3] Univ Louisville, Dept Bioengn, Louisville, KY 40202 USA
关键词
LVAD; vascular pulsatility; continuous flow; pulsatile flow; cardiovascular modeling; VENTRICULAR ASSIST DEVICE; ISCHEMIC-HEART FAILURE; CONTINUOUS-FLOW LVAD; BOVINE MODEL; ROTATIONAL SPEED; SUPPORT; NONPULSATILE; PRESSURE; SYSTEMS; VOLUME;
D O I
10.1097/MAT.0000000000000262
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Continuous flow (CF) left ventricular assist devices (LVAD) diminish vascular pressure pulsatility, which may be associated with clinically reported adverse events including gastrointestinal bleeding, aortic valve insufficiency, and hemorrhagic stroke. Three candidate CF LVAD pump speed modulation algorithms designed to augment aortic pulsatility were evaluated in mock flow loop and ischemic heart failure (IHF) bovine models by quantifying hemodynamic performance as a function of mean pump speed, modulation amplitude, and timing. Asynchronous and synchronous copulsation (high revolutions per minute [RPM] during systole, low RPM during diastole) and counterpulsation (low RPM during systole, high RPM during diastole) algorithms were tested for defined modulation amplitudes (+/- 300, +/- 500, +/- 800, and +/- 1,100 RPM) and frequencies (18.75, 37.5, and 60 cycles/minute) at low (2,900 RPM) and high (3,200 RPM) mean LVAD speeds. In the mock flow loop model, asynchronous, synchronous copulsation, and synchronous counterpulsation algorithms each increased pulse pressure (P = 931%, 210%, and 98% and reduced left ventricular external work (LVEW = 20%, 22%, 16%). Similar improvements in vascular pulsatility (1,142%) and LVEW (40%) were observed in the IHF bovine model. Asynchronous modulation produces the largest vascular pulsatility with the advantage of not requiring sensor(s) for timing pump speed modulation, facilitating potential clinical implementation.
引用
收藏
页码:526 / 532
页数:7
相关论文
共 23 条
  • [21] Feasibility of the Optical Imaging of Thrombus Formation in a Rotary Blood Pump by Near-Infrared Light
    Sakota, Daisuke
    Murashige, Tomotaka
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    ARTIFICIAL ORGANS, 2014, 38 (09) : 733 - 740
  • [22] Hemodynamics during Rotary Blood Pump Support with Speed Synchronization in Heart Failure Condition: A Modelling Study
    Htet, Zwe Lin
    Aye, Thin Pa Pa
    Singhavilai, Thamvarit
    Naiyanetr, Phornphop
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 3307 - 3310
  • [23] Estimation of pump flow rate and abnormal condition of implantable rotary blood pumps during long-term in vivo study
    Nakata, K
    Yosikawa, M
    Takano, T
    Sankai, Y
    Ohtsuka, G
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    ARTIFICIAL ORGANS, 2000, 24 (04) : 315 - 319