Tricuspid valve malfunction and ventricular pacemaker lead: Case report and review of the literature

被引:33
作者
Iskandar, Said B.
Jackson, S. Ann
Fahrig, Stephen
Mechleb, Bassam K.
Garcia, Israel D.
机构
[1] E Tennessee State Univ, James H Quillen Coll Med, Dept Cardiol, Mountain Home, TN USA
[2] Vet Affairs Med Ctr, Mountain Home, TN USA
来源
ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES | 2006年 / 23卷 / 08期
关键词
tricuspid regurgitation; pacemaker; lead; leaflet adhesions;
D O I
10.1111/j.1540-8175.2006.00289.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Pacemaker implantation can be associated with several complications, including myocardial perforation with or without pericardial effusion, venous thrombosis, vegetations of the tricuspid valve (TV) or pacing lead, and tricuspid regurgitation (TR). The TR is thought to be derived from deformity or perforation of the TV by the pacing lead or secondary to atrioventricular discordance with asynchronous ventricular pacing. Severe TR can be deleterious to the patient because it raises the central venous pressure by increasing the right sided preload. Chronically, the increase in right sided blood volume can result in an increase in the right atrial pressure leading to a decrease in venous return and low cardiac output. Severe TR from leaflet adhesion to the pacemaker lead has not been reported before. With the aging of the population and the expanding use of pacemakers and implantable cardioverter defibrillators (ICD) in clinical practice, this complication may be seen more frequently. We present a patient diagnosed with severe TR, years after his pacemaker implantation. His TR was thought to be caused by adhesion of the tricuspid valve to his pacemaker lead.
引用
收藏
页码:692 / 697
页数:6
相关论文
共 38 条
[1]   AGE-RELATED VALVULAR REGURGITATION - A STUDY BY PULSED DOPPLER ECHOCARDIOGRAPHY [J].
AKASAKA, T ;
YOSHIKAWA, J ;
YOSHIDA, K ;
OKUMACHI, F ;
KOIZUMI, K ;
SHIRATORI, K ;
TAKAO, S ;
SHAKUDO, M ;
KATO, H .
CIRCULATION, 1987, 76 (02) :262-265
[2]   Partial rupture of the tricuspid valve after extraction of permanent pacemaker leads: Detection by transesophageal echocardiography [J].
Assayag, P ;
Thuaire, C ;
Benamer, H ;
Sebbah, J ;
Leport, C ;
Brochet, E .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (06) :971-974
[3]   PULSED AND CONTINUOUS WAVE DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF VALVULAR REGURGITATION IN NORMAL SUBJECTS [J].
BERGER, M ;
HECHT, SR ;
VANTOSH, A ;
LINGAM, U .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 13 (07) :1540-1545
[4]   PROBLEMS IN HEMODYNAMIC DIAGNOSIS OF TRICUSPID INSUFFICIENCY [J].
CAIRNS, KB ;
KLOSTER, FE ;
BRISTOW, JD ;
LEES, MH ;
GRISWOLD, HE .
AMERICAN HEART JOURNAL, 1968, 75 (02) :173-&
[5]   Complications arising after implantation of DDD pacemakers: The MOST experience [J].
Ellenbogen, KA ;
Hellkamp, AS ;
Wilkoff, BL ;
Camunas, JL ;
Love, JC ;
Hadjis, TA ;
Lee, KL ;
Lamas, GA .
AMERICAN JOURNAL OF CARDIOLOGY, 2003, 92 (06) :740-741
[6]   LACERATION OF TRICUSPID VALVE BY A PACEMAKER WIRE [J].
FISHENFELD, J ;
LAMY, Y .
CHEST, 1972, 61 (07) :697-+
[7]   ENTRAPMENT OF A TINED PACEMAKER ELECTRODE IN THE TRICUSPID-VALVE - A CASE-REPORT [J].
FRANDSEN, F ;
OXHOJ, H ;
NIELSEN, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (09) :1082-1083
[8]   ADHESIONS OF PACING CATHETER TO TRICUSPID VALVE - ADHESIVE ENDOCARDITIS [J].
FRIEDBER.HD ;
DCUNHA, GF .
THORAX, 1969, 24 (04) :498-&
[9]  
FURMAN S, 1958, Surg Forum, V9, P245
[10]   PRESUMPTIVE TRICUSPID-VALVE MALFUNCTION INDUCED BY A PACEMAKER LEAD - CASE-REPORT AND REVIEW OF THE LITERATURE [J].
GIBSON, TC ;
DAVIDSON, RC ;
DESILVEY, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1980, 3 (01) :88-95