Potentiometric flow injection determination of serum bromide in patients with epilepsy

被引:13
作者
Katsu, T [1 ]
Mori, Y [1 ]
Matsuka, N [1 ]
Gomita, Y [1 ]
机构
[1] OKAYAMA UNIV,SCH MED,DEPT HOSP PHARM,OKAYAMA 700,JAPAN
关键词
flow injection analysis; ion-selective electrode; bromide; therapeutic drug monitoring; epilepsy;
D O I
10.1016/S0731-7085(96)02017-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection system was constructed using a bromide-selective electrode and used to determine serum bromide in patients with epilepsy. A 10-mu l serum sample was injected into a carrier stream flowing at 0.12 mi min(-1). Potential changes and bromide concentrations were linearly related in the range 3-50 mM. The lower limit of detection for serum bromide was 1 mM and this electrode sensitivity spanned the entire concentration range required for bromide therapy (9-24 mM). The results compared favourably with those obtained by colorimetry. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:1829 / 1832
页数:4
相关论文
共 50 条
  • [21] FLOW-INJECTION POTENTIOMETRIC DETERMINATION OF HARMINE AND HARMALINE HALLUCINOGENS
    HASSAN, SSM
    ELNEMMA, EM
    HAMADA, MA
    ELECTROANALYSIS, 1995, 7 (07) : 656 - 659
  • [22] Potentiometric Determination of Sibutramine Using Batch and Flow Injection Analysis
    El Gohary, Nesrine A.
    El Nashar, Rasha M.
    Aboul-Enien, Hassan Y.
    ANALYTICAL LETTERS, 2011, 44 (1-3) : 241 - 257
  • [23] Flow-Injection Potentiometric Method for the Routine Determination of Chloride: Application to Bread Analysis
    Almeida, Sofia A. A.
    Rebelo, Tania S. C. R.
    Heitor, Ana M.
    Oliveira, M. Beatriz P. P.
    Sales, M. Goreti F.
    CURRENT ANALYTICAL CHEMISTRY, 2010, 6 (04) : 277 - 287
  • [24] Potentiometric flow injection analysis of dicyclomine hydrochloride in serum, urine and milk
    Ibrahim, H
    Issa, YM
    Abu-Shawish, HM
    ANALYTICA CHIMICA ACTA, 2005, 532 (01) : 79 - 88
  • [25] Potentiometric Determination of Imatinib under Batch and Flow Injection Analysis Conditions
    Karim, Shereen E. Abdel
    El-Nashar, Rasha M.
    Abadi, Ashraf H.
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2012, 7 (10): : 9668 - 9681
  • [26] Polymeric Membrane Sensors for Batch and Flow Injection Potentiometric Determination of Procainamide
    Alrabiah, Haitham
    Homoda, Atef M. A.
    Radowan, Abdellatef A.
    Ezzeldin, Essam
    Mostafa, Gamal A. E.
    IEEE SENSORS JOURNAL, 2021, 21 (04) : 4198 - 4208
  • [27] Flow injection differential potentiometric determination of lysine by using a lysine biosensor
    García-Villar, N
    Saurina, J
    Hernández-Cassou, S
    ANALYTICA CHIMICA ACTA, 2003, 477 (02) : 315 - 324
  • [28] Carbon paste and PVC electrodes for the flow injection potentiometric determination of dextromethorphan
    Khaled, Elmorsy
    Hassan, H. N. A.
    Mohamed, Gehad G.
    Seleim, A. A.
    TALANTA, 2010, 81 (1-2) : 510 - 515
  • [29] Determination of the titratable acidity and the pH of wine based on potentiometric flow injection analysis
    Vahl, Katja
    Kahlert, Heike
    von Muehlen, Lisandro
    Albrecht, Anja
    Meyer, Gabriele
    Behnert, Juergen
    TALANTA, 2013, 111 : 134 - 139
  • [30] Determination of nitrate in carbon black by flow injection analysis (FIA) with potentiometric detection
    Canal, IG
    Lima, JLFC
    Montenegro, MCBSM
    PrezOlmos, R
    ANALUSIS, 1997, 25 (02) : 32 - 35