Correlation analysis between radiation exposure and the image quality of cone-beam computed tomography in the dental clinical environment

被引:5
作者
Song, Chang-Ho [1 ,2 ]
Yeom, Han-Gyeol [3 ]
Kim, Jo-Eun [1 ,2 ]
Huh, Kyung-Hoe [1 ,2 ]
Yi, Won-Jin [1 ,2 ]
Heo, Min-Suk [1 ,2 ]
Lee, Sam-Sun [1 ,2 ]
机构
[1] Seoul Natl Univ, Sch Dent, Dept Oral & Maxillofacial Radiol, Seoul, South Korea
[2] Seoul Natl Univ, Sch Dent, Dent Res Inst, Seoul, South Korea
[3] Wonkwang Univ, Daejeon Dent Hosp, Coll Dent, Dept Oral & Maxillofacial Radiol, 77 Dunsan Ro, Daejeon 35233, South Korea
基金
新加坡国家研究基金会;
关键词
Radiation Exposure; Cone-Beam Computed Tomography; Diagnostic X-Ray; Quality Control;
D O I
10.5624/isd.20220129
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study was conducted to measure the radiation exposure and image quality of various cone-beam computed tomography (CBCT) machines under common clinical conditions and to analyze the correlation between them. Materials and Methods: Seven CBCT machines used frequently in clinical practice were selected. Because each machine has various sizes of fields of view (FOVs), 1 large FOV and 1 small FOV were selected for each machine. Radiation exposure was measured using a dose-area product (DAP) meter. The quality of the CBCT images was analyzed using 8 image quality parameters obtained using a dental volume tomography phantom. For statistical analysis, regression analysis using a generalized linear model was used. Results: Polymethyl-methacrylate (PMMA) noise and modulation transfer function (MTF) 10% showed statistically significant correlations with DAP values, presenting positive and negative correlations, respectively (P<0.05). Image quality parameters other than PMMA noise and MTF 10% did not demonstrate statistically significant correlations with DAP values. Conclusion: As radiation exposure and image quality are not proportionally related in clinically used equipment, it is necessary to evaluate and monitor radiation exposure and image quality separately.
引用
收藏
页码:283 / 288
页数:6
相关论文
共 21 条
  • [1] American Colleges of Radiology [Internet, COMPL ACCR INF CT
  • [2] [Anonymous], 2020, OPEN GOVT DATA
  • [3] Development of a compact computed tomographic apparatus for dental use
    Arai, Y
    Tammisalo, E
    Iwai, K
    Hashimoto, K
    Shinoda, K
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) : 245 - 248
  • [4] Diagnostic accuracy of imaging examinations for peri-implant bone defects around titanium and zirconium dioxide implants: A systematic review and meta-analysis
    Chagas, Mariana Murai
    Kobayashi-Velasco, Solange
    Gimenez, Thais
    Paraiso Cavalcanti, Marcelo Gusmao
    [J]. IMAGING SCIENCE IN DENTISTRY, 2021, 51 (04) : 363 - 372
  • [5] Clinical image quality evaluation for panoramic radiography in Korean dental clinics
    Choi, Bo-Ram
    Choi, Da-Hye
    Huh, Kyung-Hoe
    Yi, Won-Jin
    Heo, Min-Suk
    Choi, Soon-Chul
    Bae, Kwang-Hak
    Lee, Sam-Sun
    [J]. IMAGING SCIENCE IN DENTISTRY, 2012, 42 (03) : 183 - 190
  • [6] Destouet Judy M, 2005, J Am Coll Radiol, V2, P585
  • [7] Do K-H., 2016, Annals of the ICRP, V45, P113, DOI 10.1177/0146645316637783
  • [8] Principles of CT: Radiation Dose and Image Quality
    Goldman, Lee W.
    [J]. JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2007, 35 (04) : 213 - 225
  • [9] International Commission on Radiological Protection, 2004, Ann ICRP, V34, P1
  • [10] Clinical comparison of intraoral CMOS and PSP detectors in terms of time efficiency, patient comfort, and subjective image quality
    Kamburoglu, Kivanc
    Samunahmetoglu, Ercin
    Eratam, Nejlan
    Sonmez, Gul
    Karahan, Sevilay
    [J]. IMAGING SCIENCE IN DENTISTRY, 2022,