Introduction of a flexible polymeric heart valve prosthesis with special design for aortic position

被引:39
作者
Daebritz, SH
Fausten, B
Hermanns, B
Schroeder, J
Groetzner, J
Autschbach, R
Messmer, BJ
Sachweh, JS
机构
[1] Univ Munich, Univ Hosp Grosshadern, Dept Cardiac Surg, D-81377 Munich, Germany
[2] Univ Hosp, Dept Thorac & Cardiovasc Surg, Aachen, Germany
[3] Univ Hosp, Dept Pathol, Aachen, Germany
[4] Univ Hosp, Dept Cardiol, Aachen, Germany
关键词
heart valve prosthesis; aortic valve surgery; polyurethane; biodegradation;
D O I
10.1016/j.ejcts.2004.02.040
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Current prosthetic heart valves necessitate permanent anticoagulation or have limited durability and impaired hemodynamic performance compared to natural valves. Recently a polymeric valve prostheses with special design for mitral position demonstrated excellent in vitro and in vivo results with improved durability and no need for permanent anti coagulation. In this study, a respective flexible polymeric aortic valve is presented and in vitro and in vivo results are reported. Methods: The aortic prosthesis (ADIAM(R) lifescience AG, Erkelenz, Germany) is entirely made of polycarbonaturethane. The tri-leaflet flexible prosthesis mimicks the natural aortic valve and has a diminished pressure loss and reduced stress and strain peaks at the commissures. The valve underwent long-term in vitro testing and in vivo-testing in a growing calve animal model (20 weeks, 7 aortic valves) and was compared to two different commercial bioprostheses. Results: The polymeric aortic heart valve substitute demonstrated excellent in vitro and in vivo hemodynamics. Five/seven animals with aortic PCU-prostheses had an excellent clinical long-term course. The explanted valves showed a variable degree of calcification. Two of the seven animals died at 27 and 77 days due to pannus overgrowth causing severe LVOTO without degeneration of the valve itself. Both animals with commercial bioprostheses had to be sacrificed because of congestive heart failure related to structural degeneration of the bioprosthesis after 10 and 30 days of implantation. There was no increased thrombogenity of the PCU valves compared to bioprostheses. Conclusion: The new flexible polymeric aortic valve prosthesis is superior to current bioprostheses in animal testing. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:946 / 952
页数:7
相关论文
共 26 条
[1]   POLYURETHANE ARTIFICIAL HEART VALVES IN ANIMALS [J].
AKUTSU, T ;
DREYER, B ;
KOLFF, WJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1959, 14 (06) :1045-1048
[3]   DESIGN, FABRICATION AND EVALUATION OF A TRILEAFLET PROSTHETIC HEART-VALVE [J].
CHETTA, GE ;
LLOYD, JR .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1980, 102 (01) :34-41
[4]   Introduction of a flexible polymeric heart valve prosthesis with special design for mitral position [J].
Daebritz, SH ;
Sachweh, JS ;
Hermanns, B ;
Fausten, B ;
Franke, A ;
Groetzner, J ;
Klosterhalfen, B ;
Messmer, BJ .
CIRCULATION, 2003, 108 (10) :134-139
[5]   Late hemodynamic and clinical outcomes of aortic valve replacement with the Carpentier-Edwards Perimount pericardial bioprosthesis [J].
Dellgren, G ;
David, TE ;
Raanani, E ;
Armstrong, S ;
Ivanov, J ;
Rakowski, H .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 124 (01) :146-154
[6]  
Edmunds LH, 1996, TEX HEART I J, V23, P24
[7]   European experience with the Mosaic bioprosthesis [J].
Eichinger, WB ;
Botzenhardt, F ;
Gunzinger, R ;
Kemkes, BM ;
Sosnowski, A ;
Maïza, D ;
Coto, EO ;
Bleese, N .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 124 (02) :333-339
[8]  
*FDA, 1994, REPL HEART VALV GUID
[9]  
HEROLD M, 1987, POLYURETHANES BIOMED, V2
[10]  
HILBERT SL, 1987, J THORAC CARDIOV SUR, V94, P419