Effect of Temperature of Sodium Hypochlorite on Cyclic Fatigue Resistance of Heat-treated Reciprocating Files

被引:16
|
作者
Keles, Ahmet [1 ]
Ozyurek, Emel Uzunoglu [1 ]
Uyanik, Mehmet Ozgur [1 ]
Nagas, Emre [1 ]
机构
[1] Hacettepe Univ, Fac Dent, Dept Endodont, TR-06100 Ankara, Turkey
关键词
Cyclic fatigue resistance; heat treatment; sodium hypochlorite; reciprocation; PHASE-TRANSFORMATION; FRACTURE PROPERTIES; BENDING PROPERTIES; WAVEONE GOLD; DISSOLUTION; CORROSION; EVOLUTION; CANALS; BLUE;
D O I
10.1016/j.joen.2018.11.003
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The aim of this study was to evaluate the effect of different temperatures of sodium hypochlorite (NaOCl) on the cyclic fatigue resistance of the heat treated file systems. Methods: The cyclic fatigue resistance of Reciproc 25 (VDW Dental, Munich, Germany), Reciproc Blue R25 (VDW), WaveOne Primary (Dentsply Maillefer, Ballaigues, Switzerland), WaveOne Gold Primary (Dentsply Maillefer), and One Shape (25.06; Micro Mega, Besancon, France) was tested in the following conditions: control (no immersion), immersion in distilled water (DW) at 37 degrees C +/- 1 degrees C, immersion in DW at 60 degrees C +/- 1 degrees C, immersion in NaOCl at 37 degrees C +/- 1 degrees C, and immersion in NaOCl at 60 degrees C +/- 1 degrees C. The immersion period was 5 minutes for all conditions. A stainless steel artificial canal with a curvature of 60 and a 5-mm radius was used. The time to failure and fragment lengths were recorded for each instrument, and data were subjected to statistical analysis. Results: Reciproc Blue showed greater cyclic fatigue resistance in all conditions compared with the other systems (P <.05). Immersion in NaOCl at 60 degrees C negatively affected all instruments' cyclic fatigue resistance except Reciproc Blue. Immersion in DW at 60 degrees C increased the cyclic fatigue resistance of Reciproc Blue. Heat-treated files presented higher fatigue resistance than traditional files. Conclusions: A high temperature of NaOCl affected the cyclic fatigue life of the file systems. Instruments produced with heat treatment presented a longer fatigue life.
引用
收藏
页码:205 / 208
页数:4
相关论文
共 50 条
  • [1] Effect of the temperature of sodium hypochlorite on the cyclic fatigue resistance of ProTaper Gold rotary files
    Mousavi, Seyed Amir
    Norouzi, Niloofar
    Memarzadeh, Bahareh
    Havaei, Seyed Rohollah
    Yousefshahi, Hazhir
    DENTAL AND MEDICAL PROBLEMS, 2021, 58 (04) : 533 - 537
  • [2] Bending resistance and cyclic fatigue of a new heat-treated reciprocating instrument
    Nogueira Leal Silva, Emmanuel Joao
    Rodrigues, Cristiane
    Leal Vieira, Victor Talarico
    Belladonna, Felipe Goncalves
    De-Deus, Gustavo
    Lopes, Helio Pereira
    SCANNING, 2016, 38 (06) : 837 - 841
  • [3] Effects of Sodium Hypochlorite Concentration and Temperature on the Cyclic Fatigue Resistance of Heat-treated Nickel-titanium Rotary Instruments
    Alfawaz, Hussam
    Alqedairi, Abdullah
    Alsharekh, Hala
    Almuzaini, Eman
    Alzahrani, Shand
    Jamleh, Ahmed
    JOURNAL OF ENDODONTICS, 2018, 44 (10) : 1563 - 1566
  • [4] Comparison of cyclic fatigue resistance between heat-treated and conventional retreatment files
    Ozkan, Hatice Buyukozer
    Surme, Koray
    Akman, Hayri
    Er, Kursat
    REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL, 2023, 64 (03): : 112 - 117
  • [5] Cyclic Fatigue Resistance of Heat-treated Nickel-titanium Instruments after Immersion in Sodium Hypochlorite and/or Sterilization
    Pedulla, Eugenio
    Benites, Angela
    La Rosa, Giusy M.
    Plotino, Gianluca
    Grande, Nicola M.
    Rapisarda, Ernesto
    Generali, Luigi
    JOURNAL OF ENDODONTICS, 2018, 44 (04) : 648 - 653
  • [6] Dynamic Cyclic Fatigue Resistance of Heat-treated Nickel Titanium Instruments in Reciprocating Motion
    Sobral, Thais Kauana Magalhaes
    Piasecki, Lucila
    Tomazinho, Flavia Sens Fagundes
    Kirchhoff, Alison Luis
    Gabardo, Marilisa Carneiro Leao
    Mattos, Natanael Henrique Ribeiro
    Filho, Flares Baratto
    Fariniuk, Luiz Fernando
    EUROPEAN ENDODONTIC JOURNAL, 2023, 8 (03): : 201 - 206
  • [7] Resistance to Cyclic Fatigue of Nickel-Titanium Files Immersed in Sodium Hypochlorite at Body Temperature
    Abuhulaibah, Hind F.
    AbuMostafa, Ammar
    INTERNATIONAL JOURNAL OF DENTISTRY, 2020, 2020
  • [8] Effect of temperature on the cyclic fatigue resistance of thermally treated reciprocating instruments
    Michel Espinosa Klymus
    Murilo Priori Alcalde
    Rodrigo Ricci Vivan
    Marcus Vinicius Reis Só
    Bruno Carvalho de Vasconselos
    Marco Antonio Hungaro Duarte
    Clinical Oral Investigations, 2019, 23 : 3047 - 3052
  • [9] Effect of temperature on the cyclic fatigue resistance of thermally treated reciprocating instruments
    Klymus, Michel Espinosa
    Alcalde, Murilo Priori
    Vivan, Rodrigo Ricci
    Reis So, Marcus Vinicius
    de Vasconselos, Bruno Carvalho
    Hungaro Duarte, Marco Antonio
    CLINICAL ORAL INVESTIGATIONS, 2019, 23 (07) : 3047 - 3052
  • [10] Effect of sodium hypochlorite and saline on cyclic fatigue resistance of WaveOne Gold and Reciproc reciprocating instruments
    Elnaghy, A. M.
    Elsaka, S. E.
    INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (10) : 991 - 998