High precision direct analysis of magnesium isotope ratio by ion chromatography/multicollector-ICPMS using wet and dry plasma conditions

被引:23
|
作者
Karasinski, Jakub [1 ]
Bulska, Ewa [1 ]
Wojciechowski, Marcin [1 ]
Halicz, Ludwik [1 ,2 ]
Krata, Agnieszka Anna [1 ]
机构
[1] Univ Warsaw, Faulty Chem, Biol & Chem Res Ctr, Zwirki & Wigury 101, PL-02089 Warsaw, Poland
[2] Geol Survey Israel, 30 Malkhei Israel St, IL-95501 Jerusalem, Israel
关键词
MC-ICPMS; Wet and dry plasma mode; Isotope ratio; Magnesium; Ion chromatography; MASS SPECTROMETRY; FRACTIONATION; CHONDRULES; SEPARATION; METEORITES; SYSTEM;
D O I
10.1016/j.talanta.2016.12.033
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work the applicability of Ion Chromatography (IC) coupled to Multicollector Inductively Coupled Plasma Mass Spectrometry (MC-ICPMS) for on-line magnesium isotope ratio analysis was explored. Various instrumental setups were employed to enable continuous magnesium separation from the sample matrix by IC followed by MC-ICPMS. The performance of two separation columns IonPac CS16 (ID 5 mm or ID 3 mm) connected with appropriate CERS 500 suppressors (4 mm or 2 mm) using dry and wet plasma conditions was compared. With the use of ID 3 mm column and 2 mm suppressor it was possible to apply dry plasma mode with Aridus II desolvation system. Mass discrimination and instrument drift were corrected by sample-standard bracketing method using the Mg-26/Mg-24 isotope ratio of DSM-3 as standard. Good accuracy and high precision of the magnesium isotope ratio (generally 0.15%0 (2 SD)) were achieved for wet and dry plasma modes; both were comparable to off-line Mg separation and continuous measurement. The sensitivity of MC-ICPMS measurements with dry plasma was 25 times higher in comparison to wet plasma conditions. Robustness and applicability of the method was demonstrated for matrix-rich natural water and rock samples magnesium isotope analysis.
引用
收藏
页码:64 / 68
页数:5
相关论文
共 50 条
  • [1] Isotope Analysis of Sulfur, Bromine, and Chlorine in Individual Anionic Species by Ion Chromatography/Multicollector-ICPMS
    Zakon, Yevgeni
    Halicz, Ludwik
    Gelman, Faina
    ANALYTICAL CHEMISTRY, 2014, 86 (13) : 6495 - 6500
  • [2] Direct determination of δ44/42Ca isotope ratio by ion chromatography/low-resolution multicollector ICPMS
    Karasinski, Jakub
    Bulska, Ewa
    Halicz, Ludwik
    Wojciechowski, Marcin
    Krata, Agnieszka Anna
    JOURNAL OF MASS SPECTROMETRY, 2018, 53 (01): : 78 - 82
  • [3] On-line separation of strontium from a matrix and determination of the 87Sr/86Sr ratio by Ion Chromatography/Multicollector-ICPMS
    Karasinski, Jakub
    Bulska, Ewa
    Wojciechowski, Marcin
    Krata, Agnieszka Anna
    Halicz, Ludwik
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2016, 31 (07) : 1459 - 1463
  • [4] High-precision magnesium isotope analysis of carbonates by laser ablation MC-ICP-MS using wet and dry conditions
    Lu, Jue
    Chen, Wei
    Sun, Jian
    Lin, Jie
    Luo, Tao
    Liu, Yong-Hong
    Zhao, Kui-Dong
    Jiang, Shao-Yong
    Liu, Yong-Sheng
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (08) : 1665 - 1674
  • [5] High precision Ag isotope measurements in geologic materials by multiple-collector ICPMS:: An evaluation of dry versus wet plasma
    Schoenbaechler, Maria
    Carlson, Richard W.
    Horan, Mary E.
    Mock, Timothy D.
    Hauri, Erik H.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2007, 261 (2-3) : 183 - 191
  • [6] High precision PU isotope ratio measurements using multicollector ICP-MS
    Croudace, IW
    Warneke, T
    Warwick, PE
    Taylor, RN
    Milton, JA
    ENVIRONMENTAL RADIOCHEMICAL ANALYSIS II, 2003, (291): : 104 - 112
  • [7] Species-specific Antimony Isotope Analysis by High Performance Liquid Chromatography Coupled with Multicollector ICPMS Using Hydride Generation as an Interface
    Chen, Linjie
    Du, Yuanyuan
    Li, Shuyang
    Liu, Xing
    Shen, Jun
    Zhu, Zhenli
    ANALYTICAL CHEMISTRY, 2025, 97 (04) : 2264 - 2272
  • [8] Ion Exchange Chromatography and Multicollector-Inductively Coupled Plasma-Mass Spectrometry for High Precision Isotopic Measurements of Titanium Isotope Ratios
    Tang Suo-Han
    Zhu Xiang-Kun
    Zhao Xin-Miao
    Li Jin
    Yan Bin
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 39 (12) : 1830 - 1835
  • [9] High-precision isotope ratio analysis of inorganic bromide by continuous flow MC-ICPMS
    Gelman, Faina
    Halicz, Ludwik
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2011, 307 (1-3) : 211 - 213
  • [10] Detection of counterfeit antiviral drug Heptodin™ and classification of counterfeits using isotope amount ratio measurements by multicollector inductively coupled plasma mass spectrometry (MC-ICPMS) and isotope ratio mass spectrometry (IRMS)
    Santamaria-Fernandez, Rebeca
    Hearn, Ruth
    Wolff, Jean-Claude
    SCIENCE & JUSTICE, 2009, 49 (02) : 102 - 106