The Effects of Transcatheter Valve Crimping on Pericardial Leaflets

被引:110
作者
Alavi, S. Hamed
Groves, Elliott M.
Kheradvar, Arash
机构
[1] Univ Calif Irvine, Edwards Lifesci Ctr Adv Cardiovasc Technol, Dept Biomed Engn, Irvine, CA 92697 USA
[2] Univ Calif Irvine, Div Cardiol, Irvine, CA 92697 USA
关键词
MEDICAL PROGRESS; AORTIC-STENOSIS; HEART-VALVES; IMPLANTATION; REPLACEMENT; INJURY; STENTS;
D O I
10.1016/j.athoracsur.2013.11.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Transcatheter aortic valve replacement has emerged as a promising therapy for treatment of severe aortic stenosis. Although it has been shown that these valves can be safely delivered and implanted, studies of valve longevity are lacking because of the infancy of the technology. Particularly, the effects of stent crimping on the valve's leaflets have not yet been sufficiently investigated. In this study, we have characterized the effects of crimping on pericardial leaflets in time and through the depth of the tissue. Methods. To test the structural changes at the surface and deep layers of bovine pericardial leaflets, scanning electron microscopy and second-harmonic generation microscopy were used. An uncrimped tissue sample was imaged, followed by imaging a segment of tissue after crimping in a stented transcatheter valve, immediately after, at 20 minutes, and 60 minutes after crimping. The crimping experiment was performed for multiple crimping sizes (ie, 14F, 16F, and 18F). We defined a damage index that quantifies the level of leaflet structural changes as a result of crimping. Results. Based on the calculated damage indices and analyses of the raw images, it was determined that crimping does measurable damage to the leaflet tissue that persists with time. Conclusions. Significant tissue damage was observed at the surface layers of the leaflets. In the deeper tissue layers, damage was substantial for 14F crimping; however, it became less significant but still visible for larger collapse profiles. Crimping may induce substantial structural damage to pericardial leaflets that does not improve with time. (C) 2014 by The Society of Thoracic Surgeons
引用
收藏
页码:1260 / 1266
页数:7
相关论文
共 20 条
[1]   Characterizing the Collagen Fiber Orientation in Pericardial Leaflets Under Mechanical Loading Conditions [J].
Alavi, S. Hamed ;
Ruiz, Victor ;
Krasieva, Tatiana ;
Botvinick, Elliot L. ;
Kheradvar, Arash .
ANNALS OF BIOMEDICAL ENGINEERING, 2013, 41 (03) :547-561
[2]   Traumatic leaflet injury during the use of percutaneous valves: a comparative study of balloon- and self-expandable valved stents [J].
Amahzoune, Brahim ;
Bruneval, Patrick ;
Allam, Bachir ;
Lafont, Antoine ;
Fabiani, Jean-Noel ;
Zegdi, Rachid .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2013, 43 (03) :488-493
[3]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[4]   Impairment of pericardial leaflet structure from balloon-expanded valved stents [J].
de Buhr, Wiebke ;
Pfeifer, Stefan ;
Slotta-Huspenina, Julia ;
Wintermantel, Erich ;
Lutter, Georg ;
Goetz, Wolfgang A. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2012, 143 (06) :1417-1421
[5]   Editorial Comment: Pericardial traumatic injury in trans-catheter aortic valve implantation [J].
Della Barbera, Mila ;
Valente, Marialuisa ;
Basso, Cristina ;
Thiene, Gaetano .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2013, 43 (03) :493-494
[6]   Severe intraprosthetic regurgitation following trans-catheter aortic valve implantation-to crimp or not to crimp? This might be the problem [J].
Ferrari, Enrico .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2011, 39 (04) :593-594
[7]   Crimping May Affect the Durability of Transcatheter Valves: An Experimental Analysis [J].
Kiefer, Philipp ;
Gruenwald, Felix ;
Kempfert, Joerg ;
Aupperle, Heike ;
Seeburger, Joerg ;
Mohr, Friedrich Wilhelm ;
Walther, Thomas .
ANNALS OF THORACIC SURGERY, 2011, 92 (01) :155-160
[8]   THE BOVINE PERICARDIAL XENOGRAFT .1. EFFECT OF FIXATION IN ALDEHYDES WITHOUT CONSTRAINT ON THE TENSILE VISCOELASTIC PROPERTIES OF BOVINE PERICARDIUM [J].
LEE, JM ;
HABERER, SA ;
BOUGHNER, DR .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1989, 23 (05) :457-475
[9]  
Lloyd MA., 2012, Cardiac Valvular Medicine, P103
[10]   Transcatheter Aortic-Valve Replacement for Inoperable Severe Aortic Stenosis [J].
Makkar, Raj R. ;
Fontana, Gregory P. ;
Jilaihawi, Hasan ;
Kapadia, Samir ;
Pichard, Augusto D. ;
Douglas, Pamela S. ;
Thourani, Vinod H. ;
Babaliaros, Vasilis C. ;
Webb, John G. ;
Herrmann, Howard C. ;
Bavaria, Joseph E. ;
Kodali, Susheel ;
Brown, David L. ;
Bowers, Bruce ;
Dewey, Todd M. ;
Svensson, Lars G. ;
Tuzcu, Murat ;
Moses, Jeffrey W. ;
Williams, Matthew R. ;
Siegel, Robert J. ;
Akin, Jodi J. ;
Anderson, William N. ;
Pocock, Stuart ;
Smith, Craig R. ;
Leon, Martin B. .
NEW ENGLAND JOURNAL OF MEDICINE, 2012, 366 (18) :1696-1704