Evaluation of the B-Smart manometer and the CompuFlo computerized injection pump technology for accurate needle-tip injection pressure measurement during peripheral nerve blockade

被引:5
|
作者
Weisman, Robyn S. [1 ]
Bhavsar, Nirav P. [1 ]
Schuster, Kathleen A. [1 ]
Gebhard, Ralf E. [1 ]
机构
[1] Univ Miami, Miller Sch Med, Dept Anesthesiol, Miami, FL 33136 USA
关键词
NEUROLOGIC COMPLICATIONS; REGIONAL ANESTHESIA; CONTACT; RISK;
D O I
10.1136/rapm-2018-000026
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background and Objectives The exact mechanism of peripheral nerve blocks causing/leading to nerve injury remains controversial. Evidence from animal experiments suggests that intrafascicular injection resulting in high injection pressure has the potential to rupture nerve fascicles and may consequently cause permanent nerve injury and neurological deficits. The B-Smart (BS) in-line manometer and the CompuFlo (CF) computerized injection pump technology are two modalities used for monitoring pressure during regional anesthesia. This study sought to explore the accuracy of these two technologies in measuring needle-tip pressures in a simulated environment. Methods In seven simulated needle-syringe combinations, the BS and the CF devices were connected in series through a closed system and attached to a digital manometer at the tip of various needles. The pressures were evaluated in three trials per needle-syringe combination. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy (F1 Score) were determined for each needle type and overall. Results For pressures >= 15 psi and >= 20 psi, respectively, the CF device demonstrated a sensitivity of 100%, 100%; specificity of 96%, 98%; positive predictive value 93%, 93%; and negative predictive value of 100%, 100%. The BS device demonstrated a sensitivity of 60%, 100%; specificity of 99%, 95%; positive predictive value of 96%, 85%; and negative predictive value of 85%, 100%. Accuracy, as measured by the F1 Score, for detecting a pressure of >= 15 psi was 0.96 for the CF and 0.74 for the BS. Conclusions Future research is needed to explore invivo performance and evaluate whether either of these devices can impact on clinical outcomes.
引用
收藏
页码:86 / 90
页数:5
相关论文
共 1 条
  • [1] Combination of real-time needle-tip pressure sensing and minimal intensity stimulation limits unintentional intraneural injection during an ultrasound-guided peripheral nerve block procedure: A randomized, parallel group, controlled trial
    Coudray, Adrien
    Choquet, Olivier
    Swisser, Fabien
    Hochman, Mark
    Bringuier, Sophie
    Capdevila, Xavier
    JOURNAL OF CLINICAL ANESTHESIA, 2021, 74