Experimental evaluation of the deadtime phenomenon for GM detector: deadtime dependence on operating voltages

被引:4
|
作者
Almutairi, Bader [1 ,3 ,4 ]
Alam, Syed [1 ]
Akyurek, Tayfun [2 ]
Goodwin, Cameron S. [3 ]
Usman, Shoaib [1 ]
机构
[1] Missouri Univ Sci & Technol, Min & Nucl Engn, Rolla, MO 65401 USA
[2] Marmara Univ, Fac Art & Sci, Dept Phys, TR-34722 Istanbul, Turkey
[3] Rhode Isl Atom Energy Commiss, 16 Reactor Rd, Narragansett, RI 02882 USA
[4] Kuwait Inst Sci Res, Environm & Life Sci Ctr, Kuwait 13109, Kuwait
关键词
D O I
10.1038/s41598-020-75310-3
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
A detailed analysis of Geiger Mueller counter deadtime dependence on operating voltage is presented in the manuscript using four pairs of radiation sources. Based on two-source method, detector deadtime is calculated for a wide range of operating voltages which revealed a peculiar relationship between the operating voltage and the detector deadtime. In the low voltage range, a distinct drop in deadtime was observed where deadtime reached a value as low as a few microseconds (22 mu s for Tl-204, 26 mu s for Cs-137, 9 mu s for Na-22). This sharp drop in the deadtime is possibly due to reduced recombination with increasing voltage. After the lowest point, the deadtime generally increased rapidly to reach a maximum (292 mu s for Tl-204, 277 mu s for Cs-137, 258 mu s for Na-22). This rapid increase in the deadtime is mainly due to the on-set of charge multiplication. After the maximum deadtime values, there was an exponential decrease in the deadtime reaching an asymptotic low where the manufacturer recommended voltage for operation falls. This pattern of deadtime voltage dependence was repeated for all sources tested with the exception of Mn-54. Low count rates leading to a negative deadtime suggested poor statistical nature of the data collected for Mn-54 and the data while being presented here is not used for any inference.
引用
收藏
页数:15
相关论文
共 4 条
  • [1] Experimental evaluation of the deadtime phenomenon for GM detector: deadtime dependence on operating voltages
    Bader Almutairi
    Syed Alam
    Tayfun Akyurek
    Cameron S. Goodwin
    Shoaib Usman
    Scientific Reports, 10
  • [2] Simultaneous experimental evaluation of pulse shape and deadtime phenomenon of GM detector
    Bader Almutairi
    Syed Alam
    Cameron S. Goodwin
    Shoaib Usman
    Tayfun Akyurek
    Scientific Reports, 11
  • [3] Simultaneous experimental evaluation of pulse shape and deadtime phenomenon of GM detector
    Almutairi, Bader
    Alam, Syed
    Goodwin, Cameron S.
    Usman, Shoaib
    Akyurek, Tayfun
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [4] GM counter deadtime dependence on applied voltage, operating temperature and fatigue
    Akyurek, T.
    Yousaf, M.
    Liu, X.
    Usman, S.
    RADIATION MEASUREMENTS, 2015, 73 : 26 - 35