A Noninvasive Accurate Measurement of Blood Glucose Levels with Raman Spectroscopy of Blood in Microvessels

被引:70
作者
Li, Nan [1 ,2 ]
Zang, Hang [1 ,2 ]
Sun, Huimin [1 ,2 ]
Jiao, Xianzhi [1 ,2 ]
Wang, Kangkang [1 ,2 ]
Liu, Timon Cheng-Yi [1 ,2 ]
Meng, Yaoyong [1 ]
机构
[1] South China Normal Univ, Coll Biophoton, MOE Key Lab Laser Life Sci, Guangzhou 510631, Guangdong, Peoples R China
[2] South China Normal Univ, Coll Biophoton, Lab Photon Chinese Med, Guangzhou 510631, Guangdong, Peoples R China
基金
美国国家科学基金会;
关键词
Raman spectroscopy; microvessels; blood glucose; non-invasive; PCA; BP-ANN; Clarke error grid; NEAR-INFRARED SPECTROSCOPY; IN-VIVO; DIABETES MANAGEMENT; DIAGNOSTICS; TECHNOLOGY; SCATTERING; PROBE; SKIN;
D O I
10.3390/molecules24081500
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Raman spectra of human skin obtained by laser excitation have been used to non-invasively detect blood glucose. In previous reports, however, Raman spectra thus obtained were mainly derived from the epidermis and interstitial fluid as a result of the shallow penetration depth of lasers in skin. The physiological process by which glucose in microvessels penetrates into the interstitial fluid introduces a time delay, which inevitably introduces errors in transcutaneous measurements of blood glucose. We focused the laser directly on the microvessels in the superficial layer of the human nailfold, and acquired Raman spectra with multiple characteristic peaks of blood, which indicated that the spectra obtained predominantly originated from blood. Incorporating a multivariate approach combining principal component analysis (PCA) and back propagation artificial neural network (BP-ANN), we performed noninvasive blood glucose measurements on 12 randomly selected volunteers, respectively. The mean prediction performance of the 12 volunteers was obtained as an RMSEP of 0.45 mmol/L and R-2 of 0.95. It was no time lag between the predicted blood glucose and the actual blood glucose in the oral glucose tolerance test (OGTT). We also applied the procedure to data from all 12 volunteers regarded as one set, and the total predicted performance was obtained with an RMSEP of 0.27 mmol/L and an R-2 of 0.98, which is better than that of the individual model for each volunteer. This suggested that anatomical differences between volunteer fingernails do not reduce the prediction accuracy and 100% of the predicted glucose concentrations fall within Region A and B of the Clarke error grid, allowing acceptable predictions in a clinically relevant range. The Raman spectroscopy detection of blood glucose from microvessels is of great significance of non-invasive blood glucose detection of Raman spectroscopy. This innovative method may also facilitate non-invasive detection of other blood components.
引用
收藏
页数:14
相关论文
共 57 条
[1]  
Advances A., 2011, AIP ADV, V1
[2]   Raman Spectroscopy of Blood and Blood Components [J].
Atkins, Chad G. ;
Buckley, Kevin ;
Blades, Michael W. ;
Turner, Robin F. B. .
APPLIED SPECTROSCOPY, 2017, 71 (05) :767-793
[3]   Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus [J].
Baca, Justin T. ;
Finegold, David N. ;
Asher, Sanford A. .
OCULAR SURFACE, 2007, 5 (04) :280-293
[4]   Accurate Spectroscopic Calibration for Noninvasive Glucose Monitoring by Modeling the Physiological Glucose Dynamics [J].
Barman, Ishan ;
Kong, Chae-Ryon ;
Singh, Gajendra P. ;
Dasari, Ramachandra R. ;
Feld, Michael S. .
ANALYTICAL CHEMISTRY, 2010, 82 (14) :6104-6114
[5]   Noninvasive fluorescent study in situ and in real time of glucose effects on the pharmacokinetic of calcein [J].
Bégu, S ;
Devoisselle, JM ;
Mordon, S .
JOURNAL OF BIOMEDICAL OPTICS, 2002, 7 (04) :609-612
[6]   Effect of l-arginine on energy metabolism, skeletal muscle and brown adipose tissue in South Asian and Europid prediabetic men: a randomised double-blinded crossover study [J].
Boon, Mariette R. ;
Hanssen, Mark J. W. ;
Brans, Boudewijn ;
Hulsman, Cindy J. M. ;
Hoeks, Joris ;
Nahon, Kimberly J. ;
Bakker, Charlotte ;
van Klinken, Jan B. ;
Havekes, Bas ;
Schaart, Gert ;
Jazet, Ingrid M. ;
Rensen, Patrick C. N. ;
Lichtenbelt, Wouter D. van Marken .
DIABETOLOGIA, 2019, 62 (01) :112-122
[7]   A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring [J].
Cengiz, Eda ;
Tamborlane, William V. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 :S11-S16
[8]   Instrument for near infrared emission spectroscopic probing of human fingertips in vivo [J].
Chaiken, J. ;
Deng, Bin ;
Bussjager, Rebecca J. ;
Shaheen, George ;
Rice, David ;
Stehlik, Dave ;
Fayos, John .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (03)
[9]   Noninvasive blood analysis by tissue modulated NIR Raman spectroscopy [J].
Chaiken, J ;
Finney, WF ;
Knudson, PE ;
Peterson, KP ;
Peterson, CM ;
Yang, X ;
Weinstock, RS .
VISUALIZATION OF TEMPORAL AND SPATIAL DATA FOR CIVILIAN AND DEFENSE APPLICATIONS, 2001, 4368 :134-145
[10]   Metabolic Contrasts Between Youth and Adults With Impaired Glucose Tolerance or Recently Diagnosed Type 2 Diabetes: I. Observations Using the Hyperglycemic Clamp [J].
Kahn, Steven E. ;
Arslanian, Silva A. ;
Buchanan, Thomas A. ;
Edelstein, Sharon L. ;
Ehrmann, David A. ;
Nadeau, Kristen J. ;
Palmer, Jerry P. ;
Utzschneider, Kristina M. .
DIABETES CARE, 2018, 41 (08) :1696-1706