Short-term side-effects of brain MR examination at 7 T: a single-centre experience

被引:29
作者
Cosottini, M. [1 ]
Frosini, D. [2 ]
Biagi, L. [3 ]
Pesaresi, I. [4 ]
Costagli, M. [5 ]
Tiberi, G. [5 ]
Symms, M. [6 ]
Tosetti, M. [3 ]
机构
[1] Univ Pisa, Dept Translat Res & New Technol Med & Surg, I-56124 Pisa, Italy
[2] Santa Chiara Hosp, Neurol Unit, Pisa, Italy
[3] IRCCS Stella Maris, Pisa, Italy
[4] Santa Chiara Hosp, Unit Neuroradiol, Pisa, Italy
[5] IMAGO7 Res Fdn, Pisa, Italy
[6] Gen Elect ASL Scientist EMEA, Pisa, Italy
关键词
Magnetic resonance imaging; Tolerability; Side effects; 7T; Prospective study; BLIND RANDOMIZED CROSSOVER; MAGNETIC STRAY FIELDS; HUMAN HEAD; TEMPERATURE; CHILDREN; SCANNER; SAR;
D O I
10.1007/s00330-014-3177-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To study patient tolerability of brain imaging that employs an ultrahigh field (7 T) MR system We examined 180 subjects that underwent brain MR examination at 7 T. A tolerability test consisting of two parts (during patient table motion and during the examination) was administered to all subjects in order to monitor their discomfort. The scores range from 0 to 5 for the first part, and from 0 to 10 for the second part, the total score of each subject therefore ranging from 0 (no side effects reported) to 15 (lowest tolerability) A total of 51 % of subjects reported at least one side effect but all were mild in intensity and did not require examination interruption. No serious adverse event was reported. The total score (mean +/- standard deviation) was 1.1 +/- 1.5 out of 15 (mean score 0.4 +/- 0.7 out of 5 during patient table motion and 0.7 +/- 1.1 out of 10 during MR). Patient discomfort was not related to gender or health status, but it was reduced with time after system installation with increasing operator experience in performing UHF MR examinations. Ultrahigh field MRI is well tolerated without excessive discomfort to subjects. aEuro cent 7-T MRI is well tolerated with low incidence of side effects aEuro cent The subjects' discomfort during 7-T MRI is reduced as the operators' experience increases aEuro cent 7-T MRI is practicable in healthy subjects and patients with neurodegenerative diseases.
引用
收藏
页码:1923 / 1928
页数:6
相关论文
共 24 条
[1]   Nonspecific medication side effects and the nocebo phenomenon [J].
Barsky, AJ ;
Saintfort, R ;
Rogers, MP ;
Borus, JF .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2002, 287 (05) :622-627
[2]   Brain magnetic resonance imaging increases core body temperature in sedated children [J].
Bryan, Yvon F. ;
Templeton, Thomas W. ;
Nick, Todd G. ;
Szafran, Martin ;
Tung, Avery .
ANESTHESIA AND ANALGESIA, 2006, 102 (06) :1674-1679
[3]   Temperature and SAR calculations for a human head within volume and surface coils at 64 and 300 MHz [J].
Collins, CM ;
Liu, WZ ;
Wang, JH ;
Gruetter, R ;
Vaughan, JT ;
Ugurbil, K ;
Smith, MB .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (05) :650-656
[4]   SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil [J].
Collins, CM ;
Li, SZ ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (06) :847-856
[5]   The Nocebo Effect and Its Relevance for Clinical Practice [J].
Colloca, Luana ;
Miller, Franklin G. .
PSYCHOSOMATIC MEDICINE, 2011, 73 (07) :598-603
[6]   Exposure, health complaints and cognitive performance among employees of an MRI scanners manufacturing department [J].
de Vocht, F ;
van Drooge, H ;
Engels, H ;
Kromhout, H .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 23 (02) :197-204
[7]  
Frosini D, 2013, P ISMRM WORKSH ULTR
[8]   Magnetic-field-induced vertigo: A theoretical and experimental investigation [J].
Glover, P. M. ;
Cavin, I. ;
Qian, W. ;
Bowtell, R. ;
Gowland, P. A. .
BIOELECTROMAGNETICS, 2007, 28 (05) :349-361
[9]   A Large-Scale Study on Subjective Perception of Discomfort During 7 and 1.5 T MRI Examinations [J].
Heilmaier, Christina ;
Theysohn, Jens M. ;
Maderwald, Stefan ;
Kraff, Oliver ;
Ladd, Mark E. ;
Ladd, Susanne C. .
BIOELECTROMAGNETICS, 2011, 32 (08) :610-619
[10]  
Hennel F, 1999, MAGNET RESON MED, V42, P6, DOI 10.1002/(SICI)1522-2594(199907)42:1<6::AID-MRM2>3.0.CO