A First Evaluation of OMNI®, A New Device for Continuous Renal Replacement Therapy

被引:14
|
作者
Schlaepfer, Pierre [1 ,2 ]
Durovray, Jean-Daniel [2 ]
Plouhinec, Valery [1 ]
Chiappa, Cristiano [1 ]
Bellomo, Rinaldo [3 ]
Schneider, Antoine [1 ]
机构
[1] CHU Vaudois, Adult Intens Care Unit, 21 Ave Bugnon, CH-1011 Lausanne, Switzerland
[2] CHU Vaudois, Dept Anesthesiol, Lausanne, Switzerland
[3] Austin Hlth, Intens Care Unit, Heidelberg, Vic, Australia
关键词
Acute kidney injury; Renal replacement therapy; Blood purification; Renal dose; CRITICALLY-ILL PATIENTS; ACUTE KIDNEY INJURY; CONTINUOUS VENOVENOUS HEMOFILTRATION; REGIONAL CITRATE ANTICOAGULATION; RANDOMIZED CONTROLLED-TRIAL; HEPARIN ANTICOAGULATION; INTENSITY; EFFICACY; PROTOCOL;
D O I
10.1159/000451053
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Omni((R)) (B. Braun, Germany) is a new-generation, continuous renal replacement therapy (CRRT) machine designed to improve user interface, minimize downtime and optimize renal dose delivery. It was never tested in humans. Methods: We used Omni((R)) to provide CRRT in 10 critically ill patients. We collected therapy data, metabolic parameters and evaluated user's satisfaction with a survey. Results: CRRT was delivered using Omni((R)) in CVVH-heparin (6 patients) and CVVHD-citrate (4 patients) modes for a total duration of 617.7 h. No adverse event was observed. The mean filter life was 22.8 (CVVH-heparin) and 33.5 (CVVHD-citrate) h. Alarms-related downtime corresponded to 5.9% of total therapy time. Delivered renal dose was 96.6% of prescribed. Satisfactory metabolic control and fluid removal were achieved. Overall, users evaluated interface, design and usability as excellent. Conclusion: CRRT in CVVH-heparin and CVVHD-citrate modes was provided using Omni((R)) in a safe and efficient way for 10 critically ill patients.
引用
收藏
页码:11 / 17
页数:7
相关论文
共 50 条
  • [31] Principles and modalities in continuous renal replacement therapy
    Samuel Chavez-Iniguez, Jonathan
    Cerda, Jorge
    GACETA MEDICA DE MEXICO, 2018, 154 : S31 - S39
  • [32] Clinical impact of regional citrate anticoagulation in continuous renal replacement therapy in critically ill patients
    Huguet, Maria
    Rodas, Lida
    Blasco, Miquel
    Quintana, Luis F.
    Mercadal, Jordi
    Ortiz-Perez, Jose T.
    Rovira, Irene
    Poch, Esteban
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2017, 40 (12) : 676 - 682
  • [33] Hyperlactatemia Predicts Citrate Intolerance With Regional Citrate Anticoagulation During Continuous Renal Replacement Therapy
    Tan, Jia-Neng
    Haroon, Sabrina Wong Peixin
    Mukhopadhyay, Amartya
    Lau, Titus
    Murali, Tanusya M.
    Phua, Jason
    Tan, Zong-Yao
    Lee, Nicholas
    Chua, Horng-Ruey
    JOURNAL OF INTENSIVE CARE MEDICINE, 2019, 34 (05) : 418 - 425
  • [34] Vascular access and extracorporeal circuit patency in continuous renal replacement therapy
    Gonzalez de Molina, F. J.
    Galindo, M.
    Gonzalez, C.
    Broch, M. J.
    del Bano, L.
    Roglan, A.
    MEDICINA INTENSIVA, 2016, 40 (09) : 572 - 585
  • [35] Anticoagulation Strategies for Continuous Renal Replacement Therapy in France: A Survey of Practices
    Pible, Justine
    Bidar, Frank
    Chardon, Nicolas
    Cerro, Valerie
    Ichai, Carole
    Monard, Celine
    Schneider, Antoine
    Joannes-Boyau, Olivier
    Constantin, Jean-Michel
    Rimmele, Thomas
    BLOOD PURIFICATION, 2025, 54 (01) : 1 - 8
  • [36] Continuous renal replacement therapy with regional citrate anticoagulation: do we really know the details?
    Lanckohr, Christian
    Hahnenkamp, Klaus
    Boschin, Matthias
    CURRENT OPINION IN ANESTHESIOLOGY, 2013, 26 (04) : 428 - 437
  • [37] Actually Delivered Dose of Continuous Renal Replacement Therapy Is Underestimated in Hemofiltration
    Zhang, Zhongheng
    Ni, Hongying
    Fan, Haozhe
    Li, Danyu
    Xu, Xiao
    ASAIO JOURNAL, 2013, 59 (06) : 622 - 626
  • [38] Quality measures for acute kidney injury and continuous renal replacement therapy
    Rewa, Oleksa
    Mottes, Theresa
    Bagshaw, Sean M.
    CURRENT OPINION IN CRITICAL CARE, 2015, 21 (06) : 490 - 499
  • [39] Iohexol Transmembrane Clearance during Modeled Continuous Renal Replacement Therapy
    Yardman-Frank, J. Michael
    Mercier, Renee-Claude
    Wong, Craig S.
    Vilay, A. Mary
    BLOOD PURIFICATION, 2015, 39 (1-3) : 188 - 192
  • [40] Predicting and Managing Complications of Renal Replacement Therapy in the Critically Ill
    Selby, Nicholas M.
    McIntyre, Christopher W.
    BLOOD PURIFICATION, 2012, 34 (02) : 171 - 176