Penetration of Celect Inferior Vena Cava Filters: Retrospective Review of CT Scans in 265 Patients

被引:29
作者
Zhou, Dayong [1 ]
Moon, Eunice [1 ]
Bullen, Jennifer [2 ]
Sands, Mark [1 ]
Levitin, Abraham [1 ]
Wang, Weiping [1 ]
机构
[1] Cleveland Clin, Sect Intervent Radiol, Imaging Inst, Cleveland, OH 44195 USA
[2] Cleveland Clin, Dept Quantitat Hlth Sci, Cleveland, OH 44195 USA
关键词
cavogram; CT scan; inferior vena cava filter; penetration; GUNTHER-TULIP; PERFORATION; EXPERIENCE; HEMORRHAGE; RETRIEVAL; MIGRATION;
D O I
10.2214/AJR.13.11097
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to retrospectively evaluate the prevalence and extent of Celect inferior vena cava (IVC) filter penetration. MATERIALS AND METHODS. All patients with Celect filters who underwent CT between 2007 and 2012 were assessed for penetration and related complications. RESULTS. Of the 690 patients with Celect filters placed at our institution, 265 underwent CT for various reasons. The mean (SD) interval between filter placement and last CT was 167.8 +/- 264.8 days (median, 56 days; range, 0- 1592 days). Penetration of primary leg was observed in 39% (95% CI, 29- 51%) of patients within 30 days and 80% (95% CI, 70- 87%) of patients within 90 days after placement. The mean number of penetrated legs per patient was 1.8 at 30 days and 2.1 at 90 days after placement. Penetration into adjacent organs occurred in 35 of 265 (13.2%) filters (in 45 IVC filter limbs); penetration into two structures occurred with 10 filters. The sites involved were the duodenum (n = 22), aorta (n = 9), psoas muscle (n = 4), vertebral body (n = 3), pancreas (n = 2), adrenal gland (n = 1), liver (n = 1), right kidney (n = 1), lymph node (n = 1), and diaphragm (n = 1). One patient presented with abdominal pain related to penetration. CONCLUSION. This study confirms a high penetration rate for Celect IVC filters and shows that CT can characterize the extent of leg penetration. Most cases of penetration were asymptomatic, but the likelihood of penetration increased over time.
引用
收藏
页码:643 / 647
页数:5
相关论文
共 31 条
[1]   Chronic pain syndrome caused by a bird's nest filter: First case report [J].
Al-Basheer M.A. ;
Hamilton M. ;
Holdaway C. .
CardioVascular and Interventional Radiology, 2008, 31 (Suppl 2) :S182-S184
[2]  
ALZAHRANI HA, 1995, J ENDOVASC SURG, V2, P372, DOI 10.1583/1074-6218(1995)002<0372:BSNIVC>2.0.CO
[3]  
2
[4]   Inferior vena caval filters: Review of a 26-year single-center clinical experience [J].
Athanasoulis, CA ;
Kaufman, JA ;
Halpern, EF ;
Waltman, AC ;
Geller, SC ;
Fan, CM .
RADIOLOGY, 2000, 216 (01) :54-66
[5]   Duodenocaval Fistula From Inferior Vena Cava Filter Penetration Masquerading as Lower Gastrointestinal Bleeding [J].
Bathla, Lokesh ;
Panwar, Aru ;
Fitzgibbons, Robert J., Jr. ;
Balters, Marcus .
ANNALS OF VASCULAR SURGERY, 2011, 25 (08) :1140.e7-1140.e11
[6]   Symptomatic hydronephrosis caused by inferior vena cava penetration by a Greenfield filter [J].
Berger, BD ;
Jafri, SZH ;
Konczalski, M .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (01) :99-101
[7]   Symptomatic caval penetration by a Celect inferior vena cava filter [J].
Bogue, Conor O. ;
John, Philip R. ;
Connolly, Bairbre L. ;
Rea, David J. ;
Amaral, Joao G. .
PEDIATRIC RADIOLOGY, 2009, 39 (10) :1110-1113
[8]   Quality Improvement Guidelines for the Performance of Inferior Vena Cava Filter Placement for the Prevention of Pulmonary Embolism [J].
Caplin, Drew M. ;
Nikolic, Boris ;
Kalva, Sanjeeva P. ;
Ganguli, Suvranu ;
Saad, Wael E. A. ;
Zuckerman, Darryl A. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (11) :1499-1506
[9]   LATE COMPLICATION OF A GREENFIELD FIL TER ASSOCIATING CAUDAL MIGRATION AND PERFORATION OF THE ABDOMINAL-AORTA BY A RUPTURED STRUT [J].
DABBAGH, A ;
CHAKFE, N ;
KRETZ, JG ;
DEMRI, B ;
NICOLINI, P ;
FUENTES, C ;
METTAUER, B ;
EPAILLY, E ;
MUSTER, D ;
EISENMANN, B .
JOURNAL OF VASCULAR SURGERY, 1995, 22 (02) :182-187
[10]   Initial experience in 115 patients with the retrievable Cook Celect vena cava filter [J].
Doody, O. ;
Given, M. F. ;
Kavnoudias, H. ;
Street, M. ;
Thomson, K. R. ;
Lyon, S. M. .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2009, 53 (01) :64-68