Penetration of Celect Inferior Vena Cava Filters: Retrospective Review of CT Scans in 265 Patients

被引:28
|
作者
Zhou, Dayong [1 ]
Moon, Eunice [1 ]
Bullen, Jennifer [2 ]
Sands, Mark [1 ]
Levitin, Abraham [1 ]
Wang, Weiping [1 ]
机构
[1] Cleveland Clin, Sect Intervent Radiol, Imaging Inst, Cleveland, OH 44195 USA
[2] Cleveland Clin, Dept Quantitat Hlth Sci, Cleveland, OH 44195 USA
关键词
cavogram; CT scan; inferior vena cava filter; penetration; GUNTHER-TULIP; PERFORATION; EXPERIENCE; HEMORRHAGE; RETRIEVAL; MIGRATION;
D O I
10.2214/AJR.13.11097
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to retrospectively evaluate the prevalence and extent of Celect inferior vena cava (IVC) filter penetration. MATERIALS AND METHODS. All patients with Celect filters who underwent CT between 2007 and 2012 were assessed for penetration and related complications. RESULTS. Of the 690 patients with Celect filters placed at our institution, 265 underwent CT for various reasons. The mean (SD) interval between filter placement and last CT was 167.8 +/- 264.8 days (median, 56 days; range, 0- 1592 days). Penetration of primary leg was observed in 39% (95% CI, 29- 51%) of patients within 30 days and 80% (95% CI, 70- 87%) of patients within 90 days after placement. The mean number of penetrated legs per patient was 1.8 at 30 days and 2.1 at 90 days after placement. Penetration into adjacent organs occurred in 35 of 265 (13.2%) filters (in 45 IVC filter limbs); penetration into two structures occurred with 10 filters. The sites involved were the duodenum (n = 22), aorta (n = 9), psoas muscle (n = 4), vertebral body (n = 3), pancreas (n = 2), adrenal gland (n = 1), liver (n = 1), right kidney (n = 1), lymph node (n = 1), and diaphragm (n = 1). One patient presented with abdominal pain related to penetration. CONCLUSION. This study confirms a high penetration rate for Celect IVC filters and shows that CT can characterize the extent of leg penetration. Most cases of penetration were asymptomatic, but the likelihood of penetration increased over time.
引用
收藏
页码:643 / 647
页数:5
相关论文
共 50 条
  • [1] Retrospective Review of 120 Celect Inferior Vena Cava Filter Retrievals: Experience at a Single Institution
    Zhou, Dayong
    Spain, James
    Moon, Eunice
    Mclennan, Gordon
    Sands, Mark J.
    Wang, Weiping
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (12) : 1557 - 1563
  • [2] Fracture and Migration of Celect Inferior Vena Cava Filters: A Retrospective Review of 741 Consecutive Implantations
    Wang, Weiping
    Zhou, Dayong
    Obuchowski, Nancy
    Spain, James
    An, Tianzhi
    Moon, Eunice
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (11) : 1719 - 1722
  • [3] Symptomatic caval penetration by a Celect inferior vena cava filter
    Conor O. Bogue
    Philip R. John
    Bairbre L. Connolly
    David J. Rea
    Joao G. Amaral
    Pediatric Radiology, 2009, 39 : 1110 - 1113
  • [4] Symptomatic caval penetration by a Celect inferior vena cava filter
    Bogue, Conor O.
    John, Philip R.
    Connolly, Bairbre L.
    Rea, David J.
    Amaral, Joao G.
    PEDIATRIC RADIOLOGY, 2009, 39 (10) : 1110 - 1113
  • [5] Complications of inferior vena cava filters
    Sella, David M.
    Oldenburg, W. Andrew
    SEMINARS IN VASCULAR SURGERY, 2013, 26 (01) : 23 - 28
  • [6] Complications of inferior vena cava filters
    Grewal, Simer
    Chamarthy, Murthy R.
    Kalva, Sanjeeva P.
    CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2016, 6 (06) : 632 - 641
  • [7] Caval Penetration by Retrievable Inferior Vena Cava Filters: A Retrospective Comparison of Option and Gunther Tulip Filters
    Olorunsola, Olufoladare G.
    Kohi, Maureen P.
    Fidelman, Nicholas
    Westphalen, Antonio C.
    Kolli, Pallav K.
    Taylor, Andrew G.
    Gordon, Roy L.
    LaBerge, Jeanne M.
    Kerlan, Robert K.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (04) : 566 - 571
  • [8] Aortic Penetration by Temporary Inferior Vena Cava Filters: Report of an Interesting Case and Review of the Literature
    Caceres, Manuel
    Braud, Rebecca
    Weiman, Darryl
    VASCULAR AND ENDOVASCULAR SURGERY, 2012, 46 (02) : 181 - 186
  • [9] Endovascular Removal of Inferior Vena Cava Filters with Arterial Penetration
    Duncan, Christopher
    Trerotola, Scott O.
    Stavropoulos, S. William
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 29 (04) : 486 - 490
  • [10] Strut Penetration: Local Complications, Breakthrough Pulmonary Embolism, and Retrieval Failure in Patients with Celect Vena Cava Filters
    Bos, Aaron
    Thuong Van Ha
    van Beek, Darren
    Ginsburg, Michael
    Zangan, Steven
    Navuluri, Rakesh
    Lorenz, Jonathan
    Funaki, Brian
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2015, 26 (01) : 101 - 106