Posterior Capsule Opacification With Two Hydrophobic Acrylic Intraocular Lenses: 3-Year Results of a Randomized Trial

被引:22
作者
Leydolt, Christina [1 ]
Schartmuller, Daniel [1 ]
Schwarzenbacher, Luca [1 ]
Roggla, Veronika [1 ]
Schriefl, Sabine [1 ]
Menapace, Rupert [1 ]
机构
[1] Med Univ Vienna, Dept Ophthalmol, Waehringer Guertel 18-20, A-1090 Vienna, Austria
关键词
OPTIC EDGE DESIGN; CONDUCTIVE KERATOPLASTY; MODERATE HYPEROPIA; EPITHELIAL-CELLS; SILICONE; GLISTENINGS; POLYMETHYLMETHACRYLATE; ACRYSOF; BEND;
D O I
10.1016/j.ajo.2020.04.011
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the incidence and intensity of posterior capsule opacification (PCO) and neodymium- yttrium-aluminum-garnet (Nd:YAG) capsulotomy rates between 2 similar open-loop single-piece hydrophobic acrylic intraocular lenses (IOLs) that differ in the proprietary material characteristics and design features, over a period of 3 years. DESIGN: Randomized, prospective, patient- and examiner-masked clinical trial with intraindividual comparison. METHODS: SETTING: Department of Ophthalmology, Medical University of Vienna, Vienna, Austria. PATIENT POPULATION: Eighty patients (160 eyes) had bilateral cataract surgery and received a Vivinex XY1 IOL in 1 eye and an AcrySof SN60WFIOL in the other eye. OBSERVATION PROCEDURES: Follow-up examinations were performed 3 years after surgery. Digital retroillumination images were taken of each eye. The amount of PCO (score: 0-10) was assessed subjectively at the slit lamp and objectively using automated image analysis software (AQUA). MAIN OUTCOME MEASURE: PCO score (scale, 0-10). RESULTS: The mean objective PCO score of the Vivinex XY1 IOLs was 0.9 +/- 0.8 compared to the PCO score of 1.4 +/- 1.1 for the AcrySof SN60WF IOLs (P < .001). Three years postoperatively, 11.4% of patients had an Nd:YAG capsulotomy in the Vivinex XY1 eye and 18.6% had a capsulotomy in the AcrySof SN60WF eye (P = .23). CONCLUSION: The new hydrophobic acrylic Vivinex XY1 IOL showed significantly lower PCO rates and lower YAG rates compared to the AcrySof SN60WF IOL. The interaction of various factors such as hydrophobic material, smooth optic surface, and sharp posterior optic edge plays a key role in PCO development. (C) 2020 The Authors. Published by Elsevier Inc.
引用
收藏
页码:224 / 231
页数:8
相关论文
共 50 条
  • [21] Posterior capsular opacification and Nd:YAG capsulotomy rates with the iMics Y-60H and Micro AY intra-ocular lenses: 3-year results of a randomized clinical trial
    Schriefl, Sabine M.
    Leydolt, Christina
    Stifter, Eva
    Menapace, Rupert
    ACTA OPHTHALMOLOGICA, 2015, 93 (04) : 342 - 347
  • [22] Prevention of Posterior Capsule Opacification With 3 Intraocular Lens Models: a Prospective, Randomized, Long-Term Clinical Trial
    Zemaitiene, Reda
    Jasinskas, Vytautas
    MEDICINA-LITHUANIA, 2011, 47 (11): : 595 - 599
  • [23] Comparison of posterior capsule opacification development with 3 intraocular lens types Five-year prospective study
    Ronbeck, Margrethe
    Zetterstrom, Charlotta
    Wejde, Gisela
    Kugelberg, Maria
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (11) : 1935 - 1940
  • [24] Surface modification of intraocular lenses with hyaluronic acid and lysozyme for the prevention of endophthalmitis and posterior capsule opacification
    Wang, Bailiang
    Lin, Quankui
    Jin, Tingwei
    Shen, Chenghui
    Tang, Junmei
    Hana, Yuemei
    Chen, Hao
    RSC ADVANCES, 2015, 5 (05) : 3597 - 3604
  • [25] Laboratory analyses of two explanted hydrophobic acrylic intraocular lenses
    Dai, Yunhai
    Huang, Yusen
    Liu, Ting
    Xie, Lixin
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2014, 62 (06) : 737 - 739
  • [26] Posterior capsule opacification and Nd:YAG laser rates with two hydrophobic acrylic single-piece IOLs
    Schartmueller, Daniel
    Schriefl, Sabine
    Schwarzenbacher, Luca
    Leydolt, Christina
    Kundi, Michael
    Pieh, Stephan
    Menapace, Rupert
    Kriechbaum, Katharina
    EYE, 2020, 34 (05) : 857 - 863
  • [27] A comparison of posterior capsular opacification after implantation of three different hydrophobic square edge intraocular lenses
    Pai, H. Vijaya
    Pathan, Atiya
    Kamath, Yogish Subraya
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2019, 67 (09) : 1424 - 1427
  • [28] Posterior Capsule Opacification after Cataract Surgery via Implantation with Hydrophobic Acrylic Lens Compared with Silicone Intraocular Lens: A Systematic Review and Meta-Analysis
    Kwon, Ye Rin
    Hwang, Yoo Na
    Kim, Sung Min
    JOURNAL OF OPHTHALMOLOGY, 2022, 2022
  • [29] Intraindividual comparative analysis of capsule opacification after implantation of 2 single-piece hydrophobic acrylic intraocular lenses models: Three-year follow-up
    Kahraman, Guenal
    Amon, Michael
    Ferdinaro, Clara
    Nigl, Karl
    Walch, Michael
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (05) : 990 - 996
  • [30] Randomized intraindividual comparison of posterior capsule opacification between a microincision intraocular lens and a conventional intraocular lens
    Cleary, Georgia
    Spalton, David J.
    Hancox, Joanne
    Boyce, James
    Marshall, John
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (02) : 265 - 272