An EM Approach to MAP Solution of Segmenting Tissue Mixtures: A Numerical Analysis

被引:42
作者
Liang, Zhengrong [1 ]
Wang, Su [1 ,2 ]
机构
[1] SUNY Stony Brook, Dept Radiol, Stony Brook, NY 11794 USA
[2] SUNY Stony Brook, Dept Comp Sci, Stony Brook, NY 11794 USA
基金
美国国家卫生研究院;
关键词
Expectation-maximization (EM) algorithm; maximum a posteriori (MAP) image segmentation; parameter estimation; partial volume effect; tissue mixture fraction; PARTIAL VOLUME SEGMENTATION; MAGNETIC-RESONANCE IMAGES; FINITE MIXTURES; MR-IMAGES; CLASSIFICATION; MODEL;
D O I
10.1109/TMI.2008.2004670
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This work presents an iterative expectation-maximization (EM) approach to the maximum a posteriori (MAP) solution of segmenting tissue mixtures inside each image voxel. Each tissue type is assumed to follow a normal distribution across the field-of-view (FOV). Furthermore, all tissue types are assumed to be independent from each other. Under these assumptions, the summation of all tissue mixtures inside each voxel leads to the image density mean value at that voxel. The summation of all the tissue mixtures' unobservable random processes leads to the observed image density at that voxel, and the observed image density value also follows a normal distribution (image data are observed to follow a normal distribution in many applications). By modeling the underlying tissue distributions as a Markov random field across the FOV the conditional expectation of the posteriori distribution of the tissue mixtures inside each voxel is determined, given the observed image data and the current-iteration estimation of the tissue mixtures. Estimation of the tissue mixtures at next iteration is computed by maximizing the conditional expectation. The iterative EM approach to a MAP solution is achieved by a finite number of iterations and reasonable initial estimate. This MAP-EM framework provides a theoretical solution to the partial volume effect, which has been a major cause of quantitative imprecision in medical image processing. Numerical analysis demonstrated its potential to estimate tissue mixtures accurately and efficiently.
引用
收藏
页码:297 / 310
页数:14
相关论文
共 23 条
  • [1] A spatially constrained mixture model for image segmentation
    Blekas, K
    Likas, A
    Galatsanos, NP
    Lagaris, IE
    [J]. IEEE TRANSACTIONS ON NEURAL NETWORKS, 2005, 16 (02): : 494 - 498
  • [2] PARTIAL VOLUME TISSUE CLASSIFICATION OF MULTICHANNEL MAGNETIC-RESONANCE IMAGES - A MIXEL MODEL
    CHOI, HS
    HAYNOR, DR
    KIM, YM
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) : 395 - 407
  • [3] MAXIMUM LIKELIHOOD FROM INCOMPLETE DATA VIA EM ALGORITHM
    DEMPSTER, AP
    LAIRD, NM
    RUBIN, DB
    [J]. JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-METHODOLOGICAL, 1977, 39 (01): : 1 - 38
  • [4] EREMINA D, 2006, P SPIE MED IMAGING, V6144, pD1
  • [5] THE INFORMATION-CONTENT OF MR IMAGES
    FUDERER, M
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (04) : 368 - 380
  • [6] STOCHASTIC RELAXATION, GIBBS DISTRIBUTIONS, AND THE BAYESIAN RESTORATION OF IMAGES
    GEMAN, S
    GEMAN, D
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1984, 6 (06) : 721 - 741
  • [7] Partial-volume Bayesian classification of material mixtures in MR volume data using voxel histograms
    Laidlaw, DH
    Fleischer, KW
    Barr, AH
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (01) : 74 - 86
  • [8] STATISTICAL APPROACH TO X-RAY CT IMAGING AND ITS APPLICATIONS IN IMAGE-ANALYSIS .1. STATISTICAL-ANALYSIS OF X-RAY CT IMAGING
    LEI, TH
    SEWCHAND, W
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) : 53 - 61
  • [9] MRI volumetric analysis of multiple sclerosis: Methodology and validation
    Li, LH
    Li, X
    Lu, HB
    Huang, W
    Christodoulou, C
    Tudorica, A
    Krupp, LB
    Liang, ZR
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1686 - 1692
  • [10] Partial volume segmentation of brain magnetic resonance images based on maximum a posteriori probability
    Li, X
    Li, LH
    Lu, HB
    Liang, ZR
    [J]. MEDICAL PHYSICS, 2005, 32 (07) : 2337 - 2345