Elimination of eddy current artifacts in diffusion-weighted echo-planar images: The use of bipolar gradients

被引:153
作者
Alexander, AL
Tsuruda, JS
Parker, DL
机构
[1] Department of Radiology, University of Utah, Salt Lake City
[2] AC 213 SOM Med. Imaging Res. Lab., Department of Radiology, University of Utah, Salt Lake City
关键词
magnetic resonance imaging; diffusion; echo planar; distortion correction;
D O I
10.1002/mrm.1910380623
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Small gradient fields resulting from incompletely canceled eddy currents can cause geometric distortion in echo-planar images. Although this distortion is negligible in most echo-planar applications, the large gradient pulses used in diffusion-weighted echo-planar imaging can result in significant image distortion. In this report, it is shown that this distortion can be significantly reduced by the application of bipolar gradient waveforms. Both bipolar diffusion-sensitizing gradients and an inverted gradient preparatory pulse were examined for minimizing the eddy currents responsible for these distortions.
引用
收藏
页码:1016 / 1021
页数:6
相关论文
共 23 条
  • [1] ANALYSIS OF EDDY CURRENTS IN NUCLEAR-MAGNETIC-RESONANCE IMAGING
    AHN, CB
    CHO, ZH
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 17 (01) : 149 - 163
  • [2] A MRI GRADIENT WAVE-FORM MODEL FOR AUTOMATED SEQUENCE CALIBRATION
    BARKER, BR
    ARCHER, BT
    ERDMAN, WA
    PESHOCK, RM
    [J]. MEDICAL PHYSICS, 1992, 19 (06) : 1483 - 1489
  • [3] Basser PJ, 1996, J MAGN RESON SER B, V111, P209, DOI 10.1016/j.jmr.2011.09.022
  • [4] NAVIGATED DIFFUSION IMAGING OF NORMAL AND ISCHEMIC HUMAN BRAIN
    DECRESPIGNY, AJ
    MARKS, MP
    ENZMANN, DR
    MOSELEY, ME
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) : 720 - 728
  • [5] THE APPARENT DIFFUSION CONSTANT MEASURED BY MRI CORRELATES WITH PO(2) IN A RIF-1 TUMOR
    DUNN, JF
    DING, SJ
    OHARA, JA
    LIU, KJ
    RHODES, E
    WEAVER, JB
    SWARTZ, HM
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (04) : 515 - 519
  • [6] EIS M, 1994, ACTA NEUROCHIR, P344
  • [7] A PFG NMR EXPERIMENT FOR ACCURATE DIFFUSION AND FLOW STUDIES IN THE PRESENCE OF EDDY CURRENTS
    GIBBS, SJ
    JOHNSON, CS
    [J]. JOURNAL OF MAGNETIC RESONANCE, 1991, 93 (02): : 395 - 402
  • [8] USE OF A PROJECTION RECONSTRUCTION METHOD TO DECREASE MOTION SENSITIVITY IN DIFFUSION-WEIGHTED MRI
    GMITRO, AF
    ALEXANDER, AL
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (06) : 835 - 838
  • [9] Correction for distortion of echo-planar images used to calculate the apparent diffusion coefficient
    Haselgrove, JC
    Moore, JR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (06) : 960 - 964
  • [10] Diffusion-weighted NMR imaging: Application to experimental focal cerebral ischemia
    HoehnBerlage, M
    [J]. NMR IN BIOMEDICINE, 1995, 8 (7-8) : 345 - 358