Biomechanical effects of calcar screws and bone block augmentation on medial support in locked plating of proximal humeral fractures

被引:74
作者
Katthagen, Jan Christoph [1 ]
Schwarze, Michael [2 ]
Meyer-Kobbe, Josefin [1 ]
Voigt, Christine [1 ]
Hurschler, Christof [2 ]
Lill, Helmut [1 ]
机构
[1] Diakoniekrankenhaus Friederikenstift gGmbH, Dept Trauma & Reconstruct Surg, D-30169 Hannover, Germany
[2] MHH, Lab Biomech & Biomat, D-30625 Hannover, Germany
关键词
Proximal humerus; Fracture; Medial support; Calcar screw; Dynamic locking screw; Bone augmentation; INTRAMEDULLARY FIBULAR ALLOGRAFT; LOCKING-PLATE; ENDOSTEAL IMPLANT; UNSTABLE FRACTURES; MINERAL DENSITY; OPEN REDUCTION; JOINT LOADS; IN-VITRO; FIXATION; GRAFT;
D O I
10.1016/j.clinbiomech.2014.06.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The objective of this study was to investigate the biomechanical effects of medial fracture gap augmentation in locked plating of an unstable 2-part proximal humeral fracture with calcar screws and insertion of a corticocancellous bone block Furthermore the mechanical behavior of dynamic locking screws in the non-parallel arrangement of a proximal humeral plate was of interest. Methods: Thirty-two fresh frozen humeri were randomized in four equal groups. An unstable 2-part fracture was fixed by locked plating in all specimens. The basic screw setup was supplemented by additional calcar screws in one group. Humeral head screws were replaced by dynamic locking screws in a second group. The third group featured an additional corticocancellous femoral head allograft. Assessment of stiffness was followed by cyclic loading and load to failure tests. Resulting stiffness, fracture gap deflection and ultimate load were compared utilizing Bonferroni corrected t-test for independent samples. Findings: The mechanical effect of additional calcar screws was non-significant as compared to the basic screw configuration whereas bone block insertion significantly increased construct stiffness and failure load. The use of dynamic locking screws did not significantly reduce construct stiffness when compared to conventional locking screws. Interpretation: Additional calcar screws alone did not improve the initial biomechanical properties of an unstable 2-part proximal humeral fracture model. However bone block augmentation appeared to be a reliable alternative of additional bony support by raising stiffness and failure load. Dynamic locking screws did not show their expected dynamic component when used in a non-parallel arrangement. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:735 / 741
页数:7
相关论文
共 50 条
[41]   Are Polyaxially Locked Screws Advantageous in the Plate Osteosynthesis of Proximal Humeral Fractures in the Elderly? A Prospective Randomized Clinical Observational Study [J].
Voigt, Christine ;
Geisler, Andreas ;
Hepp, Pierre ;
Schulz, Arndt P. ;
Lill, Helmut .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (10) :596-602
[42]   Helical Plating Compared with Straight Plating and Nailing for Treatment of Proximal Third Humeral Shaft Fractures-A Biomechanical Study [J].
Pastor, Torsten ;
Zderic, Ivan ;
Pastor, Tatjana ;
Drenchev, Ludmil ;
Skulev, Hristo Kostov ;
van Knegsel, Kenneth P. ;
Lenz, Mark ;
Link, Bjorn-Christian ;
Gueorguiev, Boyko ;
Beeres, Frank J. P. .
MEDICINA-LITHUANIA, 2023, 59 (11)
[43]   Arthroscopy After Locked Plating of Proximal Humeral Fractures: Implant Removal, Capsular Release, and Intra-articular Findings [J].
Katthagen, Jan Christoph ;
Hennecke, Daniel ;
Jensen, Gunnar ;
Ellwein, Alexander ;
Voigt, Christine ;
Lill, Helmut .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2014, 30 (09) :1061-1067
[44]   Biomechanical comparison of bone-screw-fasteners versus traditional locked screws in plating female geriatric bone [J].
DeBaun, Malcolm R. ;
Swinford, Steven T. ;
Chen, Michael J. ;
Thio, Timothy ;
Behn, Anthony A. ;
Lucas, Justin F. ;
Bishop, Julius A. ;
Gardner, Michael J. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2020, 51 (02) :193-198
[45]   Medial support for treatment of fractures of the proximal humerus by retrograde nail osteosynthesis: biomechanical and clinical aspects [J].
Dietz, Sven Oliver ;
Sohlbach, D. ;
Kuechle, R. ;
Enders, A. ;
Hartmann, F. ;
Nowak, T. E. ;
Hofmann, A. ;
Isemer, F. E. ;
Rommens, P. M. .
OBERE EXTREMITAET-SCHULTER-ELLENBOGEN-HAND-UPPER EXTREMITY-SHOULDER ELBOW HAND, 2012, 7 (04) :221-227
[46]   Contradictory working length effects in locked plating of the distal and middle femoral fractures a biomechanical study [J].
Chao, Ching-Kong ;
Chen, You-Lin ;
Wu, Jian-Ming ;
Lin, Chen-Huei ;
Chuang, Ting-Yu ;
Lin, Jinn .
CLINICAL BIOMECHANICS, 2020, 80
[47]   Cancellous bone allograft is comparable to fibular strut allograft for augmentation in three- or four-part proximal humeral fractures [J].
Sheng, Ning ;
Wang, Qiuke ;
Chu, Guangyu ;
Wang, Lei ;
Cheng, Mengqi ;
Weng, Zhenjun ;
Wang, Yanmao ;
Rui, Biyu ;
Chen, Yunfeng .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2021, 30 (09) :2065-2072
[48]   Effect of Fibular Allograft Augmentation in Medial Column Comminuted Proximal Humeral Fractures A Randomized Controlled Trial [J].
Wang, Qiuke ;
Sheng, Ning ;
Huang, Jen-Tai ;
Zhu, Hongyi ;
Tuerxun, Maimaitiaili ;
Ruan, Zesong ;
Shi, Tingwang ;
Zhu, Yu ;
Zhang, Yunlong ;
Rui, Biyu ;
Wang, Lei ;
Chen, Yunfeng .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2023, 105 (04) :302-311
[49]   Comparison of interfragmentary compression in conventional and locked plating of proximal unicondylar tibia fractures : A biomechanical cadaver study [J].
Rahm, Stefan ;
Ebneter, Lukas ;
Schultz, Birte ;
Spross, Christian ;
Kulling, Fabrice ;
Erhardt, Johannes B. .
ACTA ORTHOPAEDICA BELGICA, 2016, 82 (03) :586-592
[50]   Biomechanical analysis of dual versus lateral locked plating in elderly bicondylar tibial plateau fractures: Does medial comminution matter? [J].
Velez, Dencel A. Garcia ;
Headford, Mikel ;
Suresh, Krishna Vangipuram ;
Liberatos, Paraskevi Marousa ;
Bledsoe, Gary ;
Revak, Thomas .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (10) :3109-3114