Long-term equilibrium potential and electrochemical impedance study of Ag/AgCl electrodes used in Harned Cell measurements of pH

被引:12
作者
Brown, Richard J. C. [1 ]
Brewer, Paul J. [1 ]
Brett, Dan J. L. [2 ]
机构
[1] Natl Phys Lab, Analyt Sci Team, Teddington TW11 0LW, Middx, England
[2] UCL, Dept Chem Engn, London WC1E 7JE, England
关键词
pH; Ag/AgCl electrodes; Harned Cell; Electrochemical impedance; Long-term degradation; SILVER/SILVER CHLORIDE ELECTRODES; STABILITY;
D O I
10.1007/s00769-009-0487-x
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A long term study of the voltage and electrochemical impedance characteristics of Ag/AgCl electrodes used in Harned Cell measurement of pH is presented. By all the measures investigated the electrodes are shown to degrade only slowly until approximately 200 days after manufacture, after which time the rate of degradation and critical failure of the electrodes increases. The absolute voltage drift of the electrodes may not be easily measured, so parameters determined directed or indirectly by electrochemical impedance spectroscopy have been assessed as a method to produce an alternative indication of electrode integrity. In this respect, resistance to charge transfer has been shown to be a very sensitive measure of changes in the characteristics of the electrodes, and the most closely related to the observed changes in voltage. Evidence is presented to support the hypothesis that the majority of electrode degradation (excluding critical failure) comes from the increased blocking of the microporous structure of the electrodes.
引用
收藏
页码:139 / 145
页数:7
相关论文
共 12 条
[1]  
BATES RG, 1973, DETERMINATION PH
[2]  
Breitenbach M., 1983, METROLOGISCHE ABHAND, V3, P245
[3]   Effect of Structural Design of Silver/Silver Chloride Electrodes on Stability and Response Time and the Implications for Improved Accuracy in pH Measurement [J].
Brewer, Paul J. ;
Brown, Richard J. C. .
SENSORS, 2009, 9 (01) :118-130
[4]   Electrochemical study of biotin-modified self-assembled monolayers: Recommendations for robust preparation [J].
Brown, Richard J. C. ;
Brett, Dan J. L. .
THESCIENTIFICWORLDJOURNAL, 2006, 6 :20-29
[5]   The microporous structure of silver/silver chloride electrodes and the implications for Harned cell operation [J].
Brown, RJC ;
Milton, MJT .
ACCREDITATION AND QUALITY ASSURANCE, 2005, 10 (07) :352-355
[6]  
*BSI, 1979, 25861979 BSI
[7]   Measurement of pH. Definition, standards, and procedures [J].
Buck, RP ;
Rondinini, S ;
Covington, AK ;
Baucke, FGK ;
Brett, CMA ;
Camoes, MF ;
Milton, MJT ;
Mussini, T ;
Naumann, R ;
Pratt, KW ;
Spitzer, P ;
Wilson, GS .
PURE AND APPLIED CHEMISTRY, 2002, 74 (11) :2169-2200
[8]  
CHAMPION R, 2006, TECHNICAL PROBLEMS E
[9]   Hydrogen insertion reaction with restricted diffusion .1. Potential step-EIS theory and review for the direct insertion mechanism [J].
Chen, JS ;
Diard, JP ;
Durand, R ;
Montella, C .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1996, 406 (1-2) :1-13
[10]   OPTIMUM ELECTROLYTIC CHLORIDING OF SILVER ELECTRODES [J].
GEDDES, LA ;
BAKER, LE ;
MOORE, AG .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :49-&