In Vitro Characterization of the OptiChamber Diamond Valved Holding Chamber

被引:15
作者
Hatley, Ross H. M. [1 ]
von Hollen, Dirk [2 ]
Sandell, Dennis [3 ]
Slator, Lois [1 ]
机构
[1] Respiron Resp Drug Delivery UK Ltd, Chichester PO20 2FT, W Sussex, England
[2] Respironics Inc, Murrysville, PA USA
[3] S5 Consulting, S-27562 Blentarp, Sweden
关键词
valved holding chamber; flow rate; in vitro equivalence; pMDI; aerodynamic particle size distribution; METERED-DOSE INHALERS; DEPOSITION; BECLOMETHASONE; DELIVERY; AEROSOLS; CHILDREN; SPACERS; DEVICES;
D O I
10.1089/jamp.2013.1067
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Use of a valved holding chamber (VHC) in conjunction with a pressurized metered dose inhaler (pMDI) can reduce issues relating to poor actuation-inhalation coordination and potentially improve the lung deposition of aerosol, compared with use of a pMDI alone. However, the performance of a VHC is influenced by different device-related factors, including the size and shape of the VHC and the material it is manufactured from (conventional versus antistatic). This study aimed to provide an in vitro characterization of an antistatic VHC, the OptiChamber Diamond VHC, comparing the aerodynamic particle size distribution of aerosol delivered via this VHC with results from a second antistatic VHC and a conventional VHC. Methods: The pMDI drug formulations (albuterol, suspension; beclomethasone dipropionate, solution) were connected to a Next Generation Impactor, either directly (pMDI alone tests) or via a VHC (VHC tests). The pMDIs were actuated (10 per product pair) and tested at extraction flow rates of 15 L/min and 30 L/min, without any time delay between actuation and inhalation. Dose delivery using the two pMDI drug formulations was compared, and is presented with reference to key aerodynamic particle size parameters. Results: Compared with tests on pMDIs alone, use of a VHC increased the dose of aerosol within the respirable range, particularly at a 15 L/min flow rate. Between-VHC comparisons indicated that the two antistatic VHCs were equivalent. Delivery of albuterol appeared to be influenced by the VHC used, but beclomethasone dipropionate seemed unaffected. Conclusions: The two antistatic VHCs were equivalent for both pMDI brands. Aerosol delivered from the antistatic VHCs at 15 L/min had a higher proportion of fine particles compared with the conventional VHC.
引用
收藏
页码:S24 / S36
页数:13
相关论文
共 15 条
[1]  
British Standards Institution, 2007, 1354412007A1200 BS E
[2]   The effect of spacers on the delivery of metered dose aerosols of nedocromil sodium and disodium cromoglycate [J].
Chew, NYK ;
Chan, HK .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2000, 200 (01) :87-92
[3]   Distribution of technetium-99m-labelled QVAR™ delivered using an Autohaler™ device in children [J].
Devadason, SG ;
Huang, T ;
Walker, S ;
Troedson, R ;
Le Souëf, PN .
EUROPEAN RESPIRATORY JOURNAL, 2003, 21 (06) :1007-1011
[4]  
European Medicines Agency/Committee for Medicinal Products for Human Use (CHMP), 2009, CPMPEWP415100REV1 EU
[5]  
Foss S.A., 1999, RESP CARE, V44, P1474
[6]   AN IN-VITRO EVALUATION OF VARIOUS SPACER DEVICES FOR METERED-DOSE INHALERS USING THE TWIN-IMPINGER [J].
HOLZNER, PM ;
MULLER, BW .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1994, 106 (01) :69-75
[7]   A Pilot Study to Assess Lung Deposition of HFA-Beclomethasone and CFC-Beclomethasone from a Pressurized Metered Dose Inhaler with and without Add-On Spacers and Using Varying Breathhold Times [J].
Leach, Chet L. ;
Colice, Gene L. .
JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2010, 23 (06) :355-361
[8]  
Mitchell JP, 2007, RESP CARE, V52, P283
[9]  
NEWMAN SP, 1982, EUR J RESPIR DIS, V63, P57
[10]   IMPROVEMENT OF PRESSURISED AEROSOL DEPOSITION WITH NEBUHALER SPACER DEVICE [J].
NEWMAN, SP ;
MILLAR, AB ;
LENNARDJONES, TR ;
MOREN, F ;
CLARKE, SW .
THORAX, 1984, 39 (12) :935-941