Ventricular Fibrillation and Tachycardia detection from surface ECG using time-frequency representation images as input dataset for machine learning

被引:44
作者
Mjahad, A. [1 ]
Rosado-Munoz, A. [1 ]
Bataller-Mompean, M. [1 ]
Frances-Villora, J. V. [1 ]
Guerrero-Martinez, J. F. [1 ]
机构
[1] Univ Valencia, GDDP, ETSE, Elect Engn Dept, Av Univ S-N, E-46100 Valencia, Spain
关键词
VF Detection; VT Detection; Time-frequency representation; Pseudo Wigner-Ville; ECG; Classification algorithms; SUPPORT VECTOR MACHINES; CARDIAC-ARRHYTHMIAS; FEATURE-SELECTION; CLASSIFICATION; FEATURES; DEFIBRILLATION; ASSOCIATION; CONTRACTION; INTERVALS; RATES;
D O I
10.1016/j.cmpb.2017.02.010
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Background and objective: To safely select the proper therapy for Ventricullar Fibrillation (VF) is essential to distinct it correctly from Ventricular Tachycardia (VT) and other rhythms. Provided that the required therapy would not be the same, an erroneous detection might lead to serious injuries to the patient or even cause Ventricular Fibrillation (VF). The main novelty of this paper is the use of time-frequency (t-f) representation images as the direct input to the classifier. We hypothesize that this method allow to improve classification results as it allows to eliminate the typical feature selection and extraction stage, and its corresponding loss of information. Methods: The standard AHA and MIT-BIH databases were used for evaluation and comparison with other authors. Previous to t-f Pseudo Wigner-Ville (PWV) calculation, only a basic preprocessing for denoising and signal alignment is necessary. In order to check the validity of the method independently of the classifier, four different classifiers are used: Logistic Regression with L2 Regularization (L2 RLR), Adaptive Neural Network Classifier (ANNC), Support Vector Machine (SSVM), and Bagging classifier (BAGG). Results: The main classification results for VF detection (including flutter episodes) are 95.56% sensitivity and 98.8% specificity, 88.80% sensitivity and 99.5% specificity for ventricular tachycardia (VT), 98.98% sensitivity and 97.7% specificity for normal sinus, and 96.87% sensitivity and 99.55% specificity for other rhythms. Conclusion: Results shows that using t-f data representations to feed classifiers provide superior performance values than the feature selection strategies used in previous works. It opens the door to be used in any other detection applications. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:119 / 127
页数:9
相关论文
共 64 条
[1]  
Alonso-Atienza F, 2012, COMPUT CARDIOL CONF, V39, P385
[2]   Feature selection using support vector machines and bootstrap methods for ventricular fibrillation detection [J].
Alonso-Atienza, Felipe ;
Luis Rojo-Alvarez, Jose ;
Rosado-Munoz, Alfredo ;
Vinagre, Juan J. ;
Garcia-Alberola, Arcadi ;
Camps-Valls, Gustavo .
EXPERT SYSTEMS WITH APPLICATIONS, 2012, 39 (02) :1956-1967
[3]   Electrocardiogram Classification Using Reservoir Computing With Logistic Regression [J].
Angel Escalona-Moran, Miguel ;
Soriano, Miguel C. ;
Fischer, Ingo ;
Mirasso, Claudio R. .
IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2015, 19 (03) :892-898
[4]  
[Anonymous], 1947, VENTRICULAR FIBRILLA, V135, P985, DOI [10.1001/jama.1947.62890150005007a, DOI 10.1001/JAMA.1947.62890150005007A]
[5]  
[Anonymous], 2015, 2015 ANN IEEE IND C, DOI [10.1109/INDICON.2015.7443220, DOI 10.1109/INDICON.2015.7443220]
[6]  
[Anonymous], P IEEE INT FUZZ SYST
[7]   Detection of ventricular fibrillation using empirical mode decomposition and Bayes decision theory [J].
Arafat, Muhammad Abdullah ;
Sieed, Jubair ;
Hasan, Md. Kamrul .
COMPUTERS IN BIOLOGY AND MEDICINE, 2009, 39 (11) :1051-1057
[8]   Support vector machine-based arrhythmia classification using reduced features of heart rate variability signal [J].
Asl, Babak Mohammadzadeh ;
Setarehdan, Seyed Kamaledin ;
Mohebbi, Maryam .
ARTIFICIAL INTELLIGENCE IN MEDICINE, 2008, 44 (01) :51-64
[9]   Bagging predictors [J].
Breiman, L .
MACHINE LEARNING, 1996, 24 (02) :123-140
[10]  
CLAASEN TACM, 1980, PHILIPS J RES, V35, P276