Vascular Enhancement and Image Quality of MDCT Pulmonary Angiography in 400 Cases: Comparison of Standard and Low Kilovoltage Settings

被引:61
作者
Matsuoka, Shin [1 ]
Hunsaker, Andetta R.
Gill, Ritu R.
Oliva, Isabel B.
Trotman-Dickenson, Beatrice
Jacobson, Francine L.
Hatabu, Hiroto
机构
[1] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
关键词
CT; lung; pulmonary angiography; pulmonary embolism; radiation; MULTIDETECTOR-ROW CT; TUBE CURRENT MODULATION; DOSE REDUCTION; SPIRAL CT; COMPUTED-TOMOGRAPHY; RADIATION-EXPOSURE; PEDIATRIC CT; CHEST CT; EMBOLISM; OPTIMIZATION;
D O I
10.2214/AJR.08.1730
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to investigate the vascular enhancement and image quality of pulmonary CT angiography performed with lower peak kilovoltage settings in a large patient sample. MATERIALS AND METHODS. This retrospective study was approved by our institutional review board, which waived the requirement for informed consent. Four hundred patients believed to have a pulmonary embolism were studied. All patients underwent 16- or 64-MDCT with automatic tube current modulation. The 200 patients in the standard peak kilovoltage group (mean age, 57 years; range, 22-95 years) underwent MDCT at 130 or 120 kVp. The 200 patients in the low peak kilovoltage group (mean age, 56 years; range, 21-92 years) underwent MDCT at 110 or 100 kVp. Vascular enhancement was evaluated by measurement of the attenuation value in the main pulmonary artery and segmental and subsegmental arteries. Image noise was quantified by measurement of the SD of the attenuation value in the main pulmonary artery. One blinded radiologist assessed image quality using visual scores. Wilcoxon's rank test was used to evaluate differences between the groups. RESULTS. Mean vascular enhancement in the main pulmonary artery had significantly higher attenuation values in the low peak kilovoltage group (376.1 +/- 102.9 HU) than in the standard peak kilovoltage group (309.2 +/- 94.8 HU) (p < 0.0001). Mean attenuation values in all measured segmental and subsegmental arteries were significantly higher in the low peak kilovoltage group than in the standard peak kilovoltage group (p < 0.0001). Image noise in the low peak kilovoltage group was significantly higher than in the standard peak kilovoltage group (p < 0.0001). There was no significant difference in the image quality scores of the two groups (p = 0.116). CONCLUSION. Lowering kilovoltage improved vascular enhancement without deterioration of image quality. The results of our study confirm previously reported preliminary findings.
引用
收藏
页码:1651 / 1656
页数:6
相关论文
共 28 条
[1]   Dose reduction in pediatric CT: A rational approach [J].
Boone, JM ;
Geraghty, EM ;
Seibert, JA ;
Wootton-Gorges, SL .
RADIOLOGY, 2003, 228 (02) :352-360
[2]   Pulmonary embolism: Radiation dose with multi-detector row CT and digital angiography for diagnosis [J].
Coche, Emmanuel ;
Vynckier, Stefaan ;
Octave-Prignot, Michelle .
RADIOLOGY, 2006, 240 (03) :690-697
[3]   Utilization patterns and diagnostic yield of 3421 consecutive multidetector row computed tomography pulmonary angiograms in a busy emergency department [J].
Donohoo, Jay H. ;
Mayo-Smith, William W. ;
Pezzullo, John A. ;
Egglin, Thomas K. .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2008, 32 (03) :421-425
[4]   Suspected and incidental pulmonary embolism on multidetector-row CT: Analysis of technical and morphological factors influencing the diagnosis in a cross-sectional cancer centre patient cohort [J].
Engelke, C ;
Manstein, P ;
Rummeny, EJ ;
Marten, K .
CLINICAL RADIOLOGY, 2006, 61 (01) :71-80
[5]   In-plane bismuth breast shields for pediatric CT: Effects on radiation dose and image quality using experimental and clinical data [J].
Fricke, BL ;
Donnelly, LF ;
Frush, DP ;
Yoshizumi, T ;
Varchena, V ;
Poe, SA ;
Lucaya, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (02) :407-411
[6]   Peripheral pulmonary arteries: How far in the lung does multi-detector row spiral CT allow analysis? [J].
Ghaye, B ;
Szapiro, D ;
Mastora, I ;
Delannoy, V ;
Duhamel, A ;
Remy, J ;
Remy-Jardin, M .
RADIOLOGY, 2001, 219 (03) :629-636
[7]   Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions [J].
Greess, H ;
Wolf, H ;
Baum, U ;
Lell, M ;
Pirkl, M ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2000, 10 (02) :391-394
[8]   CT pulmonary angiography versus ventilation-perfusion scintigraphy in pregnancy: Implications from a UK survey of doctors' knowledge of radiation exposure [J].
Groves, Ashley M. ;
Yates, Stuart J. ;
Win, Thida ;
Kayani, Irfan ;
Gallagher, Ferdia A. ;
Syed, Rizwan ;
Bomanji, Jamshed ;
Ell, Peter J. .
RADIOLOGY, 2006, 240 (03) :765-770
[9]   Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol: prospective randomized study [J].
Heyer, Christoph M. ;
Mohr, Patrick S. ;
Lemburg, Stefan P. ;
Peters, Soeren A. ;
Nicolas, Volkmar .
RADIOLOGY, 2007, 245 (02) :577-583
[10]   Technique factors and image quality as functions of patient weight at abdominal CT [J].
Huda, W ;
Scalzetti, EM ;
Levin, G .
RADIOLOGY, 2000, 217 (02) :430-435