Impingement of metal-polyethylene hip prostheses Potential cause of high systemic titanium levels?

被引:0
|
作者
Lutter, Christoph [1 ]
Kluess, Daniel [1 ]
Enz, Andreas [1 ]
Mittelmeier, Wolfram [1 ]
机构
[1] Univ Med Ctr Rostock, Dept Orthoped, Doberaner Str 142, D-18057 Rostock, Germany
来源
ORTHOPADE | 2020年 / 49卷 / 12期
关键词
Cobalt; Chromium; Nickel; Metal ions; EXERCISE-RELATED RISE; ION LEVELS; ARTHROPLASTY; BLOOD; SERUM; BEARINGS; DEBRIS;
D O I
10.1007/s00132-020-03998-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
High blood metal levels have been described in the past, primarily in metal-on-metal bearings. Other possible causes are often underestimated. This report presents the case of a 70-year-old female patient who suffered from pronounced neurological symptoms (especially fatigue and concentration problems) 13 years after implantation of a hip endoprosthesis with metal-polyethylene (ME-PE) bearing. An osteolysis in the pelvis and loosening of the acetabular component were detected. In addition, large quantities of metallic black discolored granulomas were detected in the periarticular environment during surgery. A ventral impingement with destruction of the titanium cup and the PE insert was identified as the suspected cause of this condition. The postoperative course of the blood metal levels was unexpected as titanium levels increased massively in the blood. Anamnesis, course of the disease and the surgical procedure as well as especially the course of the metal values in the blood of the patient are presented. Possible causes for the excessive occurrence of metal abrasion, the systemic distribution and potential toxic effects of titanium are explained and discussed in detail. In addition, the currently available literature on the subject is critically examined.
引用
收藏
页码:1066 / 1071
页数:6
相关论文
共 50 条
  • [41] Vibrational spectroscopy of ultra-high molecular weight polyethylene hip prostheses: influence of the sterilisation method on crystallinity and surface oxidation
    Taddei, P
    Affatato, S
    Fagnano, C
    Bordini, B
    Tinti, A
    Toni, A
    JOURNAL OF MOLECULAR STRUCTURE, 2002, 613 (1-3) : 121 - 129
  • [42] Adverse Reactions to Metal Debris in Metal-On-Polyethylene Total Hip Arthroplasty Using a Titanium-Molybdenum-Zirconium-Iron Alloy Stem
    Kiran, Manish
    Boscainos, Petros J.
    JOURNAL OF ARTHROPLASTY, 2015, 30 (02): : 277 - 281
  • [43] INCREASED METAL RELEASE FROM CEMENTED FEMORAL COMPONENTS MADE OF TITANIUM-ALLOY - 19 HIP PROSTHESES FOLLOWED WITH RADIOSTEREOMETRY (RSA)
    KARRHOLM, J
    FRECH, W
    NILSSON, KG
    SNORRASON, F
    ACTA ORTHOPAEDICA SCANDINAVICA, 1994, 65 (06): : 599 - 604
  • [44] Evaluation of the Method for Analyzing Chromium, Cobalt and Titanium Ion Levels in the Blood Following Hip Replacement with a Metal-on-Metal Prosthesis
    Barry, Janie
    Lavigne, Martin
    Vendittoli, Pascal-Andre
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2013, 37 (02) : 90 - 96
  • [45] Modification of ultra-high-molecular-weight polyethylene with nanoparticles of titanium subgroup metal oxides
    Nemeryuk, A. M.
    Lylina, M. M.
    Retivov, V. M.
    Volkov, P. A.
    Zhdanovich, O. A.
    RUSSIAN JOURNAL OF INORGANIC CHEMISTRY, 2015, 60 (12) : 1548 - 1555
  • [46] Modification of ultra-high-molecular-weight polyethylene with nanoparticles of titanium subgroup metal oxides
    A. M. Nemeryuk
    M. M. Lylina
    V. M. Retivov
    P. A. Volkov
    O. A. Zhdanovich
    Russian Journal of Inorganic Chemistry, 2015, 60 : 1548 - 1555
  • [47] Psychiatric Disorders and Genotoxicity Following Primary Metal on Polyethylene Total Hip Arthroplasty and Their Correlation to Cobalt/Chromium Levels
    Hamid, Omaima I. Abdel
    Attia, Mohamed E.
    Hirshon, Jon M.
    El-Shinawi, Mohamed
    El-Hussaini, Moustafa
    El-Setouhy, Maged
    DRUG HEALTHCARE AND PATIENT SAFETY, 2022, 14 : 97 - 111
  • [48] Ten years on: increased metal ion levels in a cohort of patients who underwent uncemented metal-on-polyethylene total hip arthroplasty
    Dover, C.
    Kuiper, J. H.
    Craig, P.
    Shaylor, P.
    BONE & JOINT JOURNAL, 2020, 102B (07): : 832 - 837
  • [49] Systemic metal ion levels in patients with modular neck total hip arthroplasties: A prospective cohort study
    Laurencon, Jonathan
    Gkagkalis, Georgios
    Augsburger, Marc
    Hassani, Hasson
    Becce, Fabio
    Rudiger, Hannes A.
    SWISS MEDICAL WEEKLY, 2014, 144 : 36S - 36S
  • [50] The Relation Between Titanium Taper Corrosion and Cobalt-Chromium Bearing Wear in Large-Head Metal-on-Metal Total Hip Prostheses A Retrieval Study
    Witt, F.
    Bosker, B. H.
    Bishop, N. E.
    Ettema, H. B.
    Verheyen, C. C. P. M.
    Morlock, M. M.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A (18): : e157(1)