摘要
Purpose of ReviewThe purpose of this review is to clarify the risks associated with MRI exams for patients with cardiac implantable electronic devices (CIEDs) and to provide information regarding the MRI examination protocol for patients with CIEDs.Recent FindingsSeveral prospective studies evaluated the feasibility of MRI exams for patients with CIEDs and reported no adverse events. These studies suggest that by following a specific MRI examination protocol and monitoring both CIED parameters and the patient's symptoms, an MRI exam can be performed by appropriately trained personnel with an acceptable benefit-to-risk ratio.SummaryBoth MR unsafe and MR conditional CIEDs are commercially available, but there are no MR safe CIEDs. The potential risks faced by patients with CIEDs during an MRI exam are always present and warrant careful monitoring. Three magnetic fields in the MRI scanner interact with the device in ways that can damage the CIED or harm the patient. Due to safety concerns and out of an abundance of caution, the majority of MRI exams for patients with CIEDs are currently denied. However, when following a specific MRI exam protocol, these risks can be mitigated.
引用
收藏
页数:9
相关论文
共 37 条
[1]  
Acikel V., 2016, J Cardiovasc Magn Reson, V18, pO121, DOI DOI 10.1186/1532-429X-18-S1-O121
[2]  
Allison Jerry, 2015, AJR Am J Roentgenol, V205, pW140, DOI 10.2214/AJR.14.14173
[3]   In vitro investigation of pacemaker lead heating induced by magnetic resonance imaging: Role of implant geometry [J].
Calcagnini, Giovanni ;
Triventi, Michele ;
Censi, Federica ;
Mattei, Eugenio ;
Bartolini, Pietro ;
Kainz, Wolfgang ;
Bassen, Howard I. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (04) :879-886
[4]   Access to magnetic resonance imaging of patients withmagnetic resonance-conditional pacemaker and implantable cardioverter-defibrillator systems: results from the Really ProMRI study [J].
Celentano, Eduardo ;
Caccavo, Vincenzo ;
Santamaria, Matteo ;
Baiocchi, Claudia ;
Melissano, Donato ;
Pisano, Ennio ;
Gallo, Paolo ;
Polcino, Antonio ;
Arena, Giuseppe ;
Patane, Santina ;
Senatore, Gaetano ;
Licciardello, Giovanni ;
Padeletti, Luigi ;
Vado, Antonello ;
Giorgi, Davide ;
Pecora, Domenico ;
Stella, Prospero ;
Anaclerio, Matteo ;
Guastaferro, Ciro ;
Giovannini, Tiziana ;
Giacopelli, Daniele ;
Gargaro, Alessio ;
Maglia, Giampiero .
EUROPACE, 2018, 20 (06) :1001-1009
[5]  
Epstein AE, 2008, CIRCULATION, V117, P2820, DOI 10.1161/CIRCUALTIONAHA.108.189741
[6]   MRI-conditional pacemakers: current perspectives [J].
Ferreira, Antonio M. ;
Costa, Francisco ;
Tralhao, Antonio ;
Marques, Hugo ;
Cardim, Nuno ;
Adragao, Pedro .
MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2014, 7 (01) :115-124
[7]   Incidence and predictors of MRI scan utilization in MRI-conditional pacemaker recipients: A multicenter experience [J].
Gopinathannair, Rakesh ;
Mar, Philip L. ;
Gandhi, Gaurang ;
Leiserowitz, Amy ;
Tripuraneni, Ajay ;
Lakkireddy, Dhanunjaya ;
Chen, Guanhua ;
Kreps, Eric ;
Botting, Luke ;
Copeland, Stephanie ;
Firsich, Nancy ;
Kioussopoulos, Kathy ;
Granato, Jerome E. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (11) :1519-1525
[8]   Cardiac troponin T in patients with cardiac implantable electronic devices undergoing magnetic resonance imaging [J].
Higgins, John V. ;
Watson, Robert E., Jr. ;
Jaffe, Allan S. ;
Dalzell, Connie ;
Acker, Nancy ;
Felmlee, Joel P. ;
Asirvatham, Samuel J. ;
Cha, Yong-Mei ;
Friedman, Paul A. ;
Kapa, Suraj .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2016, 45 (01) :91-97
[9]   Safety and Outcomes of Magnetic Resonance Imaging in Patients with Abandoned Pacemaker and Defibrillator Leads [J].
Higgins, John V. ;
Gard, Joseph J. ;
Sheldon, Seth H. ;
Espinosa, Raul E. ;
Wood, Christopher P. ;
Felmlee, Joel P. ;
Cha, Yong-Mei ;
Asirvatham, Samuel J. ;
Dalzell, Connie ;
Acker, Nancy ;
Watson, Robert E., Jr. ;
Friedman, Paul A. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (10) :1284-1290
[10]  
Horwood L, 2016, EUROPACE