CASTOR a VLSI CMOS mixed analog-digital circuit for low noise multichannel counting applications

被引:31
作者
Comes, G
Loddo, F
Hu, Y
Kaplon, J
Ly, F
Turchetta, R
Bonvicini, V
Vacchi, A
机构
[1] UNIV STRASBOURG 1,IN2P3,LEPSI,STRASBOURG,FRANCE
[2] IST NAZL FIS NUCL,TRIESTE,ITALY
关键词
D O I
10.1016/0168-9002(96)00031-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper we present the design and first experimental results of a VLSI mixed analog-digital 1.2 microns CMOS circuit (CASTOR) for multichannel radiation detectors applications demanding low noise amplification and counting of radiation pulses. This circuit is meant to be connected to pixel-like detectors. Imaging can be obtained by counting the number of hits in each pixel during a user-controlled exposure time. Each channel of the circuit features an analog and a digital part. In the former one, a charge preamplifier is followed by a CR-RC shaper with an output buffer and a threshold discriminator. In the digital part, a 16-bit counter is present together with some control logic. The readout of the counters is done serially on a common tri-state output. Daisy-chaining is possible. A 4-channel prototype has been built. This prototype has been optimised for use in the digital radiography Syrmep experiment at the Elettra synchrotron machine in Trieste (Italy): its main design parameters are: shaping time of about 850 ns, gain of 190 mV/fC and ENC (e(-) rms) = 60 + 17 C (pF). The counting rate per channel, limited by the analog part, can be as high as about 200 kHz. Characterisation of the circuit and first tests with silicon microstrip detectors are presented. They show the circuit works according to design specification and can be used for imaging applications.
引用
收藏
页码:440 / 445
页数:6
相关论文
共 12 条
  • [1] AKJETANOWICZ M, 1993, RD20 TN24
  • [2] SILICON X-RAY-DETECTOR - FOR SYNCHROTRON-RADIATION DIGITAL RADIOLOGY
    ARFELLI, F
    BARBIELLINI, G
    CANTATORE, G
    CASTELLI, E
    CRISTAUDO, P
    DALLAPALMA, L
    DIMICHIEL, M
    LONGO, R
    POROPAT, P
    ROSEI, R
    SESSA, M
    TOMASINI, F
    TROMBA, G
    VACCHI, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) : 366 - 370
  • [3] ACCURACY OF ABSORPTION METHODS IN IDENTIFICATION OF BETA-EMITTERS
    BALTAKMENS, T
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1977, 142 (03): : 535 - 538
  • [4] X-RAY-IMAGING TEST OF A MU-STRIP SILICON DETECTOR WITH A TRANSPUTER DAQ
    BENCIVELLI, W
    BERTOLUCCI, E
    BOTTIGLI, U
    CONTI, M
    DELGUERRA, A
    FANTACCI, ME
    GAMBACCINI, M
    LAROBINA, M
    MARZIANI, M
    RANDACCIO, P
    ROSSO, V
    RUSSO, P
    RUSSO, S
    SCARLATELLA, A
    STEFANINI, A
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1522 - 1525
  • [5] BERST JD, 1995, 9501 LEPSI
  • [6] BONVICINI V, 1996, 7 EUR S SEM DET SCHL, V377
  • [7] A NEW DESIGN OF A LOW-NOISE, LOW-POWER CONSUMPTION CMOS CHARGE AMPLIFIER
    HU, Y
    NYGARD, E
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 365 (01) : 193 - 197
  • [8] JINDAL RP, 1985, 4 INT C 1 F NOIS ROM
  • [9] A HIGH-RATE, LOW-NOISE, X-RAY SILICON STRIP DETECTOR SYSTEM
    LUDEWIGT, B
    JAKLEVIC, J
    KIPNIS, I
    ROSSINGTON, C
    SPIELER, H
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1037 - 1041
  • [10] CMOS LOW-NOISE AMPLIFIER FOR MICROSTRIP READOUT - DESIGN AND RESULTS
    NYGARD, E
    ASPELL, P
    JARRON, P
    WEILHAMMER, P