A novel ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS-MS) method has been established for the determination of a newly synthesized epothilone D analog (AV-EPO-106) in human plasma. The plasma samples were prepared by liquid-liquid extraction with cold tert-butyl methyl ether. The chromatographic separation was achieved within 5 min on a C-18 column with water-methanol (10:90, v/v) as mobile phase at a flow-rate of 0.8 mL/min. Mass transition of m/z 568.2 to 386.1 was measured for AV-EPO-106 in positive atmospheric pressure chemical ionization mode. A detailed validation of the method was performed as per the USFDA guidelines. For AV-EPO-106 at the concentrations of 1.0, 5.0 and 10.0 mu g/mL in human plasma, the absolute extraction recoveries were 86.17, 85.24 and 85.69%, respectively. The linear quantification range of the method was 0.10-20.0 mu g/mL in human plasma with linear correlation coefficients greater than 0.999. The intra-day and inter-day accuracy for AV-EPO-106 at the levels of 1.0, 5.0 and 10.0 mu g/mL in human plasma fell in the ranges of 98.25-100.47 and 94.19-97.25%, and the intra- and inter-day precision were in the ranges of 4.75-6.30% and 8.89-10.45%, respectively. The method was successfully applied to quantify AV-EPO-106 in human plasma to determine the half-life of this compound in human plasma. Copyright (C) 2008 John Wiley & Sons, Ltd.
引用
收藏
页码:302 / 307
页数:6
相关论文
共 50 条
  • [41] Determination of tandospirone in human plasma by a liquid chromatography-tandem mass spectrometry method
    Jin, Fengdan
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (09) : 2881 - 2887
  • [42] Simultaneous stereoselective determination of seven β-agonists in pork meat samples by ultra-performance liquid chromatography-tandem mass spectrometry
    Zhu, Bolin
    Cai, Liangzhao
    Jiang, Zhen
    Li, Qing
    Guo, Xingjie
    MICROCHEMICAL JOURNAL, 2019, 150
  • [43] Determination of loperamide in human plasma and saliva by liquid chromatography-tandem mass spectrometry
    Arafat, Tawfiq
    Arafat, Basil
    Awad, Riad
    Abu Awwad, Ahmad
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 972 : 81 - 88
  • [44] Determination of thalidomide concentration in human plasma by liquid chromatography-tandem mass spectrometry
    Bai, Nan
    Cui, Xiang-Yong
    Wang, Jin
    Sun, Chun-Guang
    Mei, He-Kun
    Liang, Bei-Bei
    Cai, Yun
    Song, Xiu-Jie
    Gu, Jing-Kai
    Wang, Rui
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2013, 5 (02) : 626 - 630
  • [45] A rapid and sensitive ultra-performance liquid chromatography-tandem mass spectrometry method for determination of phytohormones in the medicinal plant saffron
    Chen, Jing
    Tao, Yuanyuan
    Yang, Shuhui
    Jiang, Fengqin
    Zhou, Guifen
    Qian, Xiaodong
    Zhu, Yuehong
    Li, Liqin
    ANALYTICAL METHODS, 2024, 16 (09) : 1347 - 1356
  • [46] Trace-level determination of eight cholesterol oxidation products in human plasma by dispersive liquid-liquid microextraction and ultra-performance liquid chromatography-tandem mass spectrometry
    Oh, Jin-Aa
    Shin, Ho-Sang
    JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2017, 40 (13) : 691 - 698
  • [47] Pre-analytical and analytical variation of drug determination in segmented hair using ultra-performance liquid chromatography-tandem mass spectrometry
    Nielsen, Marie Katrine Klose
    Johansen, Sys Stybe
    Linnet, Kristian
    FORENSIC SCIENCE INTERNATIONAL, 2014, 234 : 16 - 21
  • [48] Determination of tiapride in human plasma using hydrophilic interaction liquid chromatography-tandem mass spectrometry
    Ya Moon
    Bok Paek
    Hui-Hyun Kim
    Hye Young Ji
    Hye Won Lee
    Hyoung-Geun Park
    Hye Suk Lee
    Archives of Pharmacal Research, 2004, 27 : 901 - 905
  • [49] Determination of tiapride in human plasma using hydrophilic interaction liquid chromatography-tandem mass spectrometry
    Moon, Y
    Paek, IB
    Kim, HH
    Ji, HY
    Lee, HW
    Park, HG
    Lee, HS
    ARCHIVES OF PHARMACAL RESEARCH, 2004, 27 (09) : 901 - 905
  • [50] Liquid chromatography-tandem mass spectrometry method of loxoprofen in human plasma
    Lee, Hye Won
    Ji, Hye Young
    Sohn, Dong Hwan
    Kim, Se-Mi
    Lee, Yong Bok
    Lee, Hye Suk
    BIOMEDICAL CHROMATOGRAPHY, 2009, 23 (07) : 714 - 718