Simulation Load Parameter on Bipolar Artificial Hip Joint Using Finite Element Method (FEM)

被引:0
作者
Al Hakim, Reza Azizul Nasa [1 ]
Edwan, Ahmad [1 ]
Jamari, Jamari [1 ]
Ismail, Rifky [1 ]
Bayuseno, Athanasius [1 ]
Tauviqirrahman, Mohammad [1 ]
Kurdi, Ojo [1 ]
Saputra, Eko [2 ]
Anwar, Iwan Budi [2 ]
Dahlan, Sakuri [3 ]
机构
[1] Diponegoro Univ, Fac Engn, Dept Mech Engn, Lab Engn Design & Tribol,UNDIP Tembalang, Jl Prof Sudharto Kampus, Semarang 50275, Indonesia
[2] Fac Engn Technol, Lab Surface Technol & Tribol, Drienerlolaan 5,Postbus 217, NL-7500 AE Enschede, Netherlands
[3] STT Wiworotomo, Dept Mech Engn, Jl Semingkir 1, Purwokerto 53134, Indonesia
关键词
hip joint; UHWPE; impingement; salat movement; finite element method (FEM);
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
UHMWPE is a biomaterial that is widely used in the orthopedic field. UHMWPE is used as a liner and is paired with a femoral head to replace the damaged hip joint. Currently, there is no research on parameters of salat biomechanics such as range of motion, joint reaction force. The results of hip joint reaction force during salat are used for simulation load parameter on UNDIP bipolar artificial hip joint simulation with those loads using finite element method (FEM). From the simulation results, impingement occurs during last sitting. Patient who using an artificial hip joint should avoid last sitting. (C) 2019 Elsevier Ltd. All rights reserved.
引用
收藏
页码:305 / 310
页数:6
相关论文
共 10 条
[1]  
Al-Bani M. H., 2003, RINGKASAN SIFAT SHOL
[2]  
Hamill J., 2006, Biomechanical Basis of Human Movement
[3]   Numerical Study of Salat Movements for Total Hip Replacement Patient [J].
Ismail, Rifky ;
Saputra, Eko ;
Tauviqirrahman, Mohammad ;
Legowo, A. B. ;
Anwar, Iwan Budiwan ;
Jamari, J. .
ADVANCES IN APPLIED MECHANICS AND MATERIALS, 2014, 493 :426-+
[4]  
Jamari J., 2014, Advanced Materials Research, V896, P272, DOI 10.4028/www.scientific.net/AMR.896.272
[5]   Current trends in total hip arthroplasty in Europe and experiences with the bicontact hip system [J].
Kiefer, Hartmuth .
TREATMENT OF OSTEOARTHRITIC CHANGE IN THE HIP: JOINT PRESERVATION OR JOINT REPLACEMENT?, 2007, :205-210
[6]   Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement [J].
Kluess, Daniel ;
Martin, Heiner ;
Mittelmeier, Wolfram ;
Schmitz, Klaus-Peter ;
Bader, Rainer .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (04) :465-471
[7]  
Narayan R., 2009, BIOMED MAT ENG
[8]  
Saputra Eko, 2017, Key Engineering Materials, V739, P96, DOI 10.4028/www.scientific.net/KEM.739.96
[9]  
Saputra Eko, 2015, Advanced Materials Research, V1123, P164, DOI 10.4028/www.scientific.net/AMR.1123.164
[10]   Finite Element Analysis of Artificial Hip Joint Movement during Human Activities [J].
Saputra, Eko ;
Anwar, Iwan Budiwan ;
Jamari, J. ;
van der Heide, Emile .
INTERNATIONAL TRIBOLOGY CONFERENCE MALAYSIA 2013, 2013, 68 :102-108