Experimental designs in the optimisation of ultrasonic bath-acid-leaching procedures for the determination of trace elements in human hair samples by atomic absorption spectrometry

被引:40
作者
Bermejo-Barrera, P [1 ]
Muñiz-Naveiro, O [1 ]
Moreda-Piñeiro, A [1 ]
Bermejo-Barrera, A [1 ]
机构
[1] Univ Santiago Compostela, Dept Analyt Chem Nutr & Bromatol, Fac Chem, E-15706 Santiago, Spain
关键词
experimental designs; acid leaching; ultrasound energy; multi-element determination; human hair;
D O I
10.1016/S0379-0738(99)00155-3
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Experimental designs were used for the optimisation of acid-leaching procedures assisted by ultrasonic energy for the extraction of Ca, Cu, Fe, Mg, Mn and Zn from human hair samples. A Plackett-Burman 2(7)x3/32 design for seven factors ([HNO3], [HCl], [H2O2], acid/oxidant solution volume, exposure time to ultrasounds, temperature of the ultrasonic bath and hair particle size) was used in order to choose the variables affecting the acid-leaching process. The variables [HNO3], [HCl] and temperature of the ultrasonic bath were found to be the most important parameters for the acid-leaching procedure, and these variables were optimised by a response surface design (central composite design 2(3)+star) which involved 16 experiments. Optimum values in the 3.7-4.2 M range were found for [HNO3], while optimum values between 3.0 and 3.5 M were found for [HCl]. The optimum temperature of the ultrasonic bath was between 80 and 90 degrees C. An acid digestion induced by microwave energy (details given) was used to obtain the total metal concentration and also for comparative purposes. Ca, Cu, Fe, Mg and Zn were measured by flame atomic absorption spectrometry (FAAS) using a conventional air/acetylene flame, while Mn was determined by electrothermal atomic absorption spectrometry (ETAAS) under optimised conditions. Two different reference materials, IAEA-085 International Atomic Energy Agency, Monaco) and NIES No. 13 (National Institute for Environmental Studies, Japan), with certified metal contents for some of the elements investigated, were used in order to verify the accuracy of the methods. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:105 / 120
页数:16
相关论文
共 16 条
[1]   Acid predigestion as a slurry pretreatment for the determination of Ca, Cu, K, Mg, Na and Zn in human scalp hair by flame atomic absorption/emission spectrometry with a high-performance nebulizer [J].
Bermejo-Barrera, P ;
Moreda-Pineiro, A ;
Moreda-Pineiro, J ;
Bermejo-Barrera, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1998, 360 (06) :707-711
[2]   Slurry sampling electrothermal atomic absorption spectrometric determination of lead, cadmium and manganese in human hair samples using rapid atomizer programs [J].
BermejoBarrera, P ;
MoredaPineiro, A ;
MoredaPineiro, J ;
BermejoBarrera, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :301-306
[3]  
BERMEJOBARRERA P, 1999, IN PRESS ANAL CHIM A
[4]  
Box G.E.P., 1989, ESTADISTICA INVESTIG
[5]  
Burguera J L, 1987, J Trace Elem Electrolytes Health Dis, V1, P21
[6]  
CAPEL ID, 1981, CLIN CHEM, V27, P879
[7]   Literature study of microwave-assisted digestion using electrothermal atomic absorption spectrometry [J].
Chakraborty, R ;
Das, AK ;
Cervera, ML ;
delaGuardia, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1996, 355 (02) :99-111
[8]  
CORTESTORO E, 1993, J RADIOANNAL NUCL CH, V167, P413
[9]  
CREASON JP, 1975, CLIN CHEM, V21, P603
[10]   METHYLMERCURY DETERMINATION BY PURGE AND TRAP-GC-FTIR-AAS AFTER NABH4 DERIVATIZATION OF AN ENVIRONMENTAL THIOSULFATE EXTRACT [J].
FILIPPELLI, M .
APPLIED ORGANOMETALLIC CHEMISTRY, 1994, 8 (7-8) :687-691