Reliability of wireless monitoring using a wearable patch sensor in high-risk surgical patients at a step-down unit in the Netherlands: a clinical validation study

被引:88
作者
Breteler, Martine J. M. [1 ,2 ]
Huizinga, Erik [1 ]
van Loon, Kim [1 ]
Leenen, Luke P. H. [3 ]
Dohmen, Daan A. J. [2 ]
Kalkman, Cor J. [1 ]
Blokhuis, Taco J. [3 ,4 ]
机构
[1] Univ Utrecht, Univ Med Ctr Utrecht, Dept Anesthesiol, Utrecht, Netherlands
[2] FocusCura, Driebergen Rijsenburg, Netherlands
[3] Univ Med Ctr Utrecht, Dept Trauma Surg, Utrecht, Netherlands
[4] Maastricht Univ, Dept Surg, Med Ctr, Maastricht, Netherlands
来源
BMJ OPEN | 2018年 / 8卷 / 02期
关键词
IN-HOSPITAL MORTALITY; INTENSIVE-CARE-UNIT; EARLY WARNING SCORE; POSTOPERATIVE COMPLICATIONS; COLORECTAL SURGERY; CARDIAC-ARREST; FAILURE; ADMISSION; RESCUE; WARDS;
D O I
10.1136/bmjopen-2017-020162
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and objectives Intermittent vital signs measurements are the current standard on hospital wards, typically recorded once every 8 hours. Early signs of deterioration may therefore be missed. Recent innovations have resulted in 'wearable' sensors, which may capture patient deterioration at an earlier stage. The objective of this study was to determine whether a wireless 'patch' sensor is able to reliably measure respiratory and heart rate continuously in high-risk surgical patients. The secondary objective was to explore the potential of the wireless sensor to serve as a safety monitor. Design In an observational methods comparisons study, patients were measured with both the wireless sensor and bedside routine standard for at least 24 hours. Setting University teaching hospital, single centre. Participants Twenty-five postoperative surgical patients admitted to a step-down unit. Outcome measures Primary outcome measures were limits of agreement and bias of heart rate and respiratory rate. Secondary outcome measures were sensor reliability, defined as time until first occurrence of data loss. Results 1568 hours of vital signs data were analysed. Bias and 95% limits of agreement for heart rate were -1.1 (-8.8 to 6.5) beats per minute. For respiration rate, bias was -2.3 breaths per minute with wide limits of agreement (-15.8 to 11.2 breaths per minute). Median filtering over a 15 min period improved limits of agreement of both respiration and heart rate. 63% of the measurements were performed without data loss greater than 2 min. Overall data loss was limited (6% of time). Conclusions The wireless sensor is capable of accurately measuring heart rate, but accuracy for respiratory rate was outside acceptable limits. Remote monitoring has the potential to contribute to early recognition of physiological decline in high-risk patients. Future studies should focus on the ability to detect patient deterioration on low care environments and at home after discharge.
引用
收藏
页数:9
相关论文
共 38 条
[1]  
Ben-Ari J., 2010, Journal of Medical Engineering & Technology, V34, P393, DOI 10.3109/03091902.2010.503308
[2]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[3]   Association between clinically abnormal observations and subsequent in-hospital mortality: a prospective study [J].
Buist, M ;
Bernard, S ;
Nguyen, TV ;
Moore, G ;
Anderson, J .
RESUSCITATION, 2004, 62 (02) :137-141
[4]   The impact of Rapid Response System on delayed emergency team activation patient characteristics and outcomes-A follow-up study [J].
Calzavacca, Paolo ;
Licari, Elisa ;
Tee, Augustine ;
Egi, Moritoki ;
Downey, Andrew ;
Quach, Jon ;
Haase-Fielitz, Anja ;
Haase, Michael ;
Bellomo, Rinaldo .
RESUSCITATION, 2010, 81 (01) :31-35
[5]  
Chan AM, 2013, IEEE ENG MED BIO, P4058, DOI 10.1109/EMBC.2013.6610436
[6]  
Chan AM, 2013, IEEE ENG MED BIO, P6115, DOI 10.1109/EMBC.2013.6610948
[7]   The value of vital sign trends for detecting clinical deterioration on the wards [J].
Churpek, Matthew M. ;
Adhikari, Richa ;
Edelson, Dana P. .
RESUSCITATION, 2016, 102 :1-5
[8]   Predicting Cardiac Arrest on the Wards A Nested Case-Control Study [J].
Churpek, Matthew M. ;
Yuen, Trevor C. ;
Huber, Michael T. ;
Park, Seo Young ;
Hall, Jesse B. ;
Edelson, Dana P. .
CHEST, 2012, 141 (05) :1170-1176
[9]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[10]   Can physiological variables and early warning scoring systems allow early recognition of the deteriorating surgical patient? [J].
Cuthbertson, Brian H. ;
Boroujerdi, Massoud ;
Mckie, Laurin ;
Aucott, Lorna ;
Prescott, Gordon .
CRITICAL CARE MEDICINE, 2007, 35 (02) :402-409