Impulse through-target X-ray tube spectrum

被引:4
作者
Kitov, BI [1 ]
Mukhachyov, YS [1 ]
机构
[1] Irkutsk State Univ, Sci Res Inst Appl Phys, Irkutsk 664003, Russia
来源
INTERNATIONAL CONFERENCE ON X-RAY AND NEUTRON CAPILLARY OPTICS | 2002年 / 4765卷
关键词
x-ray tube; pass-through anode; spectral intensity;
D O I
10.1117/12.489746
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
At present X-ray compact tubes with pass-through anodes operating either in the pulse mode or in the direct voltage one are applied to the equipment for the shady microscopy, and the local XRF analysis. The report presents the calculated spectral intensity distributions of the bremsstrahlung versus the pass-through anode thickness. The spectral function of the bremsstrahlung Mo tube with the anode thickness over 100 microns is demonstrated to contract to the narrow energy interval lying near the characteristic anode radiation range. However under the same conditions the spectrum of the pulse Cu-anode tube tends to be dichromatic. The spectral distributions of the tube bremsstrahlung operating at the direct current and pulse voltage are compared.
引用
收藏
页码:189 / 193
页数:5
相关论文
共 3 条
[1]  
FINKELSTEIN A, 1986, ZAVODSK LAB, V52, P25
[2]  
LEBED V, 1983, ZAVODSK LAB, V49, P26
[3]  
MORGOVSKY L, 1977, XRAY ANAL EQUIPMENT, P175