Influence of some experimental parameters on the results of differential scanning calorimetry - DSC.

被引:30
作者
Bernal, C [1 ]
Couto, AB [1 ]
Breviglieri, ST [1 ]
Cavalheiro, ETG [1 ]
机构
[1] Univ Fed Sao Carlos, Dept Quim, BR-13565905 Sao Carlos, SP, Brazil
来源
QUIMICA NOVA | 2002年 / 25卷 / 05期
关键词
saccharin; differential scanning calorimetry; results evaluation;
D O I
10.1590/S0100-40422002000500023
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A series of experiments were performed in order to demonstrate to undergraduate students or users of the differential scanning calorimetry (DSC), that several factors can influence the qualitative and quantitative aspects of DSC results. Saccharin, an artificial sweetner, was used as a probe and its thermal behavior is also discussed on the basis of thermogravimetric (TG) and DSC curves.
引用
收藏
页码:849 / 855
页数:7
相关论文
共 14 条
[1]  
[Anonymous], CERAMICA
[2]  
Brown M.E., 1988, Introduction to Thermal Analysis
[3]  
CAVALHEIRO ETG, 1995, QUIM NOVA, V18, P305
[4]  
GORDON S, 1963, J CHEM EDUC, V40, pA87
[5]  
Haines P.J., 1995, THERMAL METHODS ANAL
[6]  
HEMMINGER W, 1994, CALORIMETRY THERMAL, P18
[7]  
KERR PF, 1948, AM MINERAL, V33, P387
[8]   NOMENCLATURE IN THERMAL-ANALYSIS .4. [J].
MACKENZIE, RC .
THERMOCHIMICA ACTA, 1979, 28 (01) :1-6
[9]  
SKOOG DA, 1992, PRINCIPLES INSTRUMEN, P571
[10]  