Performance specification of areal surface texture instruments exemplified by a self-built WLI

被引:0
|
作者
Bulun, Georgis [1 ]
Eifler, Matthias [1 ]
Hering, Julian [2 ,3 ]
von Freymann, Georg [2 ,3 ,4 ]
Seewig, Joerg [1 ]
机构
[1] Tech Univ Kaiserslautern, Inst Measurement & Sensor Technol MTS, D-67663 Kaiserslautern, Germany
[2] Tech Univ Kaiserslautern, Phys Dept, D-67663 Kaiserslautern, Germany
[3] Tech Univ Kaiserslautern, State Res Ctr OPTIMAS, D-67663 Kaiserslautern, Germany
[4] Fraunhofer Inst Ind Math ITWM, D-67663 Kaiserslautern, Germany
来源
INTERFEROMETRY XX | 2020年 / 11490卷
关键词
calibration; performance specification; white light interference microscope; optical topography measurement; CALIBRATION;
D O I
10.1117/12.2567666
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The calibration of optical surface topography measuring instruments ensures the comparability of the resulting surface topographies. Such comparability is also desirable for self-built optical microscopes in order to optimize these instruments with regard to their systematic and stochastic measurement uncertainty. However, there is not yet a fixed set of standards in the field of interference microscopy as in the field of stylus instruments that describes methods for their calibration, verification, adjustment and performance specification [1] [2]. The associated standard for optical measuring instruments is still in preparation in the corresponding ISO working group. There is the guideline according to VDI/VDE 2655 Sheet 1.1, for the calibration of interference microscopes for surface texture measurement. This guideline is essentially a transfer of methods that origin from the calibration of stylus instruments [1]. Thus, it describes a profile-like calibration and does specifically consider areal properties of interference microscopes. The field of areal calibration is still work in progress in the international standardization. The ISO 25178-600 that was recently published defines the basic metrological characteristics of areal surface topography measuring instruments [3]. In our study, we describe the optimization and performance specification of a self-built white light interference microscope based on methods that were developed for instrument calibration. For the performance specification we use the metrological characteristics of ISO 25178-600 that are imaged with the help of a previously developed Universal Calibration Artefact for areal calibration [4]. The results of the metrological characteristics are compared with a commercially available white light interferometer (WLI). In order to improve the self-built interference microscope, the measurements are repeated with different configurations and a final setup is chosen that allows a specification of the performance of the self-built interference microscope for measuring areal surface texture.
引用
收藏
页数:17
相关论文
共 19 条
  • [1] Performance verification of areal surface texture measuring instruments with the Sk-parameters
    Eifler, Matthias
    Klauer, Katja
    Kirsch, Benjamin
    Aurich, Jan C.
    Seewig, Joerg
    MEASUREMENT, 2021, 173 (173)
  • [2] Performance verification of areal surface texture measuring instruments with the Sk-parameters
    Eifler, Matthias
    Klauer, Katja
    Kirsch, Benjamin
    Aurich, Jan C.
    Seewig, Joerg
    MEASUREMENT, 2021, 173
  • [3] Calibration and verification of areal surface texture measuring instruments
    Leach, R. K.
    Giusca, C. L.
    Haitjema, H.
    Evans, C.
    Jiang, X.
    CIRP ANNALS-MANUFACTURING TECHNOLOGY, 2015, 64 (02) : 797 - 813
  • [4] A study on surface material measures for areal surface texture measuring instruments - Measuring conditions for the areal profiling
    Nemoto, K.
    Yanagi, K.
    Aketagawa, M.
    Kanda, D.
    Yoshida, I.
    Uchidate, M.
    Key Engineering Materials, 2008, 381-382 : 241 - 244
  • [5] Effect of masonry strength on aseismic performance of self-built dwellings
    Yu, Rui-Fang
    Xie, Zhi-Qiang
    Peng, Ling-Yun
    Yu, Yan-Xiang
    Zhendong yu Chongji/Journal of Vibration and Shock, 2014, 33 (11): : 6 - 11
  • [6] Progress in the specification of optical instruments for the measurement of surface form and texture
    de Groot, Peter J.
    DIMENSIONAL OPTICAL METROLOGY AND INSPECTION FOR PRACTICAL APPLICATIONS III, 2014, 9110
  • [7] Investigation on Seismic Performance of Self-Built House in Henan Rural Areas
    Su, Congcong
    CIVIL ENGINEERING, ARCHITECTURE AND SUSTAINABLE INFRASTRUCTURE II, PTS 1 AND 2, 2013, 438-439 : 1520 - 1523
  • [8] The bridges of Barrio Nuevo: performance criteria as a tool for the urban improvement of a self-built neighborhood
    Paniagua Arguedas, Laura
    CIUDADES POPULARES EN DISPUTA: ACCESO A SUELO URBANO PARA TODOS?, VOL IV, 2016, 4 : 319 - 346
  • [9] A study on surface material measures for areal surface texture measuring instruments - 1st report: Design of the roughness measurement standard for areal profiling
    Nemoto K.
    Yanagi K.
    Aketagawa M.
    Uchidate M.
    Yoshida I.
    Seimitsu Kogaku Kaishi/Journal of the Japan Society for Precision Engineering, 2010, 76 (03): : 349 - 353
  • [10] Development of low-cost material measures for calibration of the metrological characteristics of areal surface texture instruments
    Leach, Richard
    Giusca, Claudiu
    Guttmann, Markus
    Jakobs, Peter-Juergen
    Rubert, Paul
    CIRP ANNALS-MANUFACTURING TECHNOLOGY, 2015, 64 (01) : 545 - 548