Ventralight ST and SorbaFix Versus Physiomesh and Securestrap in a Porcine Model

被引:29
作者
Deeken, Corey R. [1 ]
Matthews, Brent D. [1 ]
机构
[1] Washington Univ, Sch Med, Dept Surg, Sect Minimally Invas Surg, St Louis, MO 63110 USA
关键词
Absorbable fixation; Tensile strength; Adhesions; Ventral hernia; Tissue ingrowth; Absorbable barrier mesh; HERNIA REPAIR; ADHESION FORMATION; TISSUE INGROWTH; MESH; POLYPROPYLENE; COMPOSITE;
D O I
10.4293/108680813X13693422520125
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background and Objectives: The objective of this study was to compare mesh contracture, adhesion characteristics, tissue ingrowth, and histologic response of Ventralight ST/ SorbaFix (C. R. Bard/Davol, Warwick, RI, USA) with Physiomesh/ Securestrap (Ethicon, Somerville, NJ, USA) in a porcine model of laparoscopic ventral hernia repair. Methods: Standard laparoscopic technique was used to bilaterally implant meshes in 10 female Yorkshire swine. Each animal received either two Ventralight ST meshes (oval shaped, 10.2 ! 15.2 cm) or two Physiomesh meshes (oval shaped 10 ! 15 cm), one on either side of the midline. The meshes were fixated to the intact peritoneum with either SorbaFix (for animals receiving Ventralight ST) or Securestrap (for animals receiving Physiomesh). There were 5 animals in each group, yielding 10 of each meshfixation combination. Mesh contracture, adhesion characteristics, tissue ingrowth, and histologic response were evaluated after 14 days by image analysis, mechanical testing, and histologic staining (hematoxylin-eosin, Masson trichrome, picrosirius red, and von Willebrand factor). Results: Ventralight ST/SorbaFix and Physiomesh/Securestrap exhibited a similar percentage of mesh contracture, percentage of adhesion coverage, adhesion tenacity, collagen deposition, and levels of necrosis (P > .05 in all cases). However, Ventralight ST/SorbaFix exhibited significantly less inflammation (P = .0001), fibrosis (P = .0017), hemorrhage (P = .0001), and angiogenesis (P = .0032) and significantly greater strength of tissue ingrowth (P = .0003) than Physiomesh/Securestrap after the 14-day implantation period. Conclusions: Ventralight ST/SorbaFix exhibited more favorable strength of tissue ingrowth and histologic response and similar mesh contracture and adhesion characteristics compared with Physiomesh/Securestrap over a short-term 14-day implantation period in a preclinical porcine model.
引用
收藏
页码:549 / 559
页数:11
相关论文
共 20 条
[1]  
[Anonymous], 2011, PHYS INSTR US PACK I
[2]  
[Anonymous], 2005, SEPR IP COMP INSTR U
[3]  
[Anonymous], 2012, PRODUCT SHEET SECURE
[4]  
[Anonymous], 2001, J AM VET MED ASS, V218, P668
[5]  
[Anonymous], 2007, PROC SURG MESH INSTR
[6]  
[Anonymous], 2012, PRODUCT SHEET SORBAF
[7]  
[Anonymous], 2010, VENTR ST INSTR US PA
[8]   Comparison of contracture, adhesion, tissue ingrowth, and histologic response characteristics of permanent and absorbable barrier meshes in a porcine model of laparoscopic ventral hernia repair [J].
Deeken, C. R. ;
Matthews, B. D. .
HERNIA, 2012, 16 (01) :69-76
[9]   A review of the composition, characteristics, and effectiveness of barrier mesh prostheses utilized for laparoscopic ventral hernia repair [J].
Deeken, Corey R. ;
Faucher, Keith M. ;
Matthews, Brent D. .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2012, 26 (02) :566-575
[10]   Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair [J].
Deeken, Corey R. ;
Abdo, Michael S. ;
Frisella, Margaret M. ;
Matthews, Brent D. .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2011, 25 (05) :1541-1552