A PCI DAQ board for MWPC detectors with high-rate 2D delay line position readout

被引:16
作者
Levchanovski, FV
Gebauer, B
Litvinenko, EI
Nikiforov, AS
Prikhodko, VI
Schulz, C
Wilpert, T
机构
[1] Hahn Meitner Inst Berlin GmbH, D-14109 Berlin, Germany
[2] Joint Nucl Res Inst, Frank Lab Neutron Phys, Dubna 141980, Russia
关键词
DAQ board; thermal neutron/X-ray detector; delay line readout; high-rate multihit readout; 2D position; time-of-flight and pulse-height readout; histogram and list-mode data recording;
D O I
10.1016/j.nima.2004.05.023
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
For high rate thermal neutron (and e.g. also synchrotron radiation) detection a PCI data acquisition (DAQ) board for readout of two-dimensional position-sensitive multi-wire proportional chamber (MWPC) detectors with delay line position encoding was developed, and a first series of boards was produced and tested. The MWPC board comprises a deadtime-less 8-channel multihit TDC chip of the F1 type with similar to 120 ps LSB, a 40 MHz flash ADC, a FPGA, a FIFO, a 0.9 GFLOP DSP chip TMS320C6711 and a 256 MB histogram memory. For position encoding a position-independent start and four position-dependent stop signals are recorded from the MWPC anode and from either end of X and Y delay lines connected to the MWPC cathodes, respectively. Time-of-flight (TOF) can be measured with reference to a pulsed beam or chopper signal either with greater than or equal to 25 ns bin size for TOF ranges of up to similar to 420 ms or alternatively with similar to 120 ps LSB for ranges less than or equal to 16 ms. The pulse height of the MWPC anode signal is digitized in the flash ADC for neutron-gamma and background discrimination. Hardware event filtering in the FPGA allows event rates up to 1 Mevents/s by disentangling of two events occurring within the delay line time length of e.g. 250 ns. Whilst list mode data are recorded with full rate and full resolution on the host PC disk, 1D-3D histograms can be accumulated online in the on-board HM for a fraction of the data and with reduced resolution. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:413 / 416
页数:4
相关论文
共 5 条
[1]   Detector requirements for the European spallation neutron source ESS [J].
Gebauer, B .
ADVANCES IN NEUTRON SCATTERING INSTRUMENTATION, 2002, 4785 :182-196
[2]   Development of a hybrid MSGC detector for thermal neutron imaging with a MHz data acquisition and histogramming system [J].
Gebauer, B ;
Schulz, C ;
Richter, G ;
Levchanovsky, FV ;
Nikiforov, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 471 (1-2) :249-253
[3]  
LEVCHANOVSKI FV, 2001, P DANEF 2000 WORKSH, P85
[4]   Development of hybrid MSGC detectors with high position and time-of-flight resolution for neutron scattering experiments at ESS [J].
Schulz, C ;
Gebauer, B ;
Richter, G ;
Namaschk, B ;
Balykov, LN ;
Levchanovski, FV ;
Nikiforov, A ;
Shashkin, VI ;
Klimov, AY ;
Rogov, VV .
ADVANCES IN NEUTRON SCATTERING INSTRUMENTATION, 2002, 4785 :203-213
[5]  
*TEX INSTR, TMS320C6000 PER REF