Increased total knee arthroplasty ultra-high molecular weight polyethylene wear using a clinically relevant hyaluronic acid simulator lubricant

被引:33
作者
DesJardins, J.
Aurora, A.
Tanner, S. L.
Pace, T. B.
Acampora, K. B.
LaBerge, M.
机构
[1] Clemson Univ, Dept Bioengn, Clemson, SC 29634 USA
[2] Carolina Orthoped Res Fdn, Greenville, SC USA
关键词
biomaterials wear; hyaluronic acid; total knee replacement; cobalt chromium femoral component pitting; force controlled simulation;
D O I
10.1243/09544119JEIM30
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, osteoarthritic and periprosthetic synovial fluid samples were rheologicalty and biochemically compared to develop a hyaluronic acid (HA) supplemented bovine serum (BS) lubricant that mimicked the properties of human joint synovial fluid. The effect of this BS + HA lubricant (50 per cent bovine calf serum + 1.5 g/l HA) on the wear rate of ultra-high molecular weight polyethylene (UHMWPE) during a total knee replacement wear test was then investigated. In conjunction with biochemical similarities, the rheological analysis showed that the BS + HA lubricant viscosity was not statistically different to aspirated total knee arthroplasty (TKA) revision joint fluid viscosity over a range of physiologic shear rates. Gravimetric results at 5 million wear testing cycles showed that the BS + HA lubricant produced an average of 6.88 times more UHMWPE wear than 50 per cent bovine serum lubricant alone. The BS + HA lubricated CoCr femoral component surfaces revealed pitting and surface roughening that was not observed using standard bovine serum only lubricants, but that was similar to the metallic surface corrosion observed on in vivo CoCr femoral component retrievals. These findings support the hypothesis that the addition of HA to simulator lubricant is capable of producing CoCr femoral component surface damage similar to that observed in vivo.
引用
收藏
页码:609 / 623
页数:15
相关论文
共 55 条
[1]  
*AM NAT STAND I, 1982, ANN BOOK ASTM STAND, P212
[2]  
American Society for Testing and Materials, 1996, ANN BOOK ASTM STAND
[3]   Effect of different metal ions on the oxidative damage and antioxidant capacity of hyaluronic acid [J].
Balogh, GT ;
Illés, J ;
Székely, Z ;
Forrai, E ;
Gere, A .
ARCHIVES OF BIOCHEMISTRY AND BIOPHYSICS, 2003, 410 (01) :76-82
[4]  
BENSON LC, 2003, ADV BIOENG, P42955
[5]   Development of a six station knee wear simulator and preliminary wear results [J].
Burgess, IC ;
Kolar, M ;
Cunningham, JL ;
Unsworth, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :37-47
[6]   Analysis of products diffused into UHMWPE prosthetic components in vivo [J].
Costa, L ;
Bracco, P ;
del Prever, EB ;
Luda, MP ;
Trossarelli, L .
BIOMATERIALS, 2001, 22 (04) :307-315
[7]   Ultra-high molecular weight polyethylene .1. Mechano-oxidative degradation [J].
Costa, L ;
Luda, MP ;
Trossarelli, L .
POLYMER DEGRADATION AND STABILITY, 1997, 55 (03) :329-338
[8]   Oxidation in orthopaedic UHMWPE sterilized by gamma-radiation and ethylene oxide [J].
Costa, L ;
Luda, MP ;
Trossarelli, L ;
del Prever, EMB ;
Crova, M ;
Gallinaro, P .
BIOMATERIALS, 1998, 19 (7-9) :659-668
[9]  
DELECRIN J, 1994, CLIN ORTHOP RELAT R, V307, P240
[10]   The use of a force-controlled dynamic knee simulator to quantify the mechanical performance of total knee replacement designs during functional activity [J].
DesJardins, JD ;
Walker, PS ;
Haider, H ;
Perry, J .
JOURNAL OF BIOMECHANICS, 2000, 33 (10) :1231-1242