Navigation surgery for dental implants: Assessment of accuracy of the image guided implantology system

被引:85
作者
Casap, N
Wexler, A
Persky, N
Schneider, A
Lustmann, J
机构
[1] Hebrew Univ Jerusalem, Hadassah Fac Dent Med, Dept Oral & Maxillofacial Surg, IL-91120 Jerusalem, Israel
[2] Hebrew Univ Jerusalem, Hadassah Fac Dent Med, Dept Prosthodont, IL-91120 Jerusalem, Israel
[3] DenX Adv Dent Syst LTD, Res & Dev, Moshav Ora, Israel
关键词
D O I
10.1016/j.joms.2004.06.028
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study is to assess the accuracy of the navigation provided by the Image Guided Implantology system (DenX Advanced Dental Systems, Moshav Ora, Israel), which was designed to guide the surgeon in the placement of dental implants. Materials and Methods: Seven jaw models incorporating special ceramic spheres were imaged by dental computerized tomography. The coordinates of these reference ceramic markers were calculated by the Image Guided Implantology and compared with their actual position coordinates as identified on the computed tomography image. Results: The overall mean spatial navigation error was 0.35 +/- 0.14 mm. The 1-tailed probability of any single measurement exceeding 0.75 mm was less than 0.003, and that exceeding 1 mm was less than 0.0001. Conclusion: The Image Guided Implantology system provides highly accurate navigation with less than 0.73 mm error, which is acceptable in dental implantology. The accurate reporting of the exact position of the drilling bur should minimize the potential risk of damage to critical anatomic structures. The accurate intraoperative navigation allows the surgeon to precisely transfer the presurgical plan to the patient. (C) 2004 American Association of Oral and Maxillofacial Surgeons.
引用
收藏
页码:116 / 119
页数:4
相关论文
共 6 条
[1]  
BRIEF J, 2001, INT C SERIES, V1230, P739
[2]   Computer assisted oral and maxillofacial surgery -: a review and an assessment of technology [J].
Hassfeld, S ;
Mühling, J .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2001, 30 (01) :2-13
[3]  
Koele W, 2002, RHINOLOGY, V40, P1
[4]   Emerging technologies for surgery in the 21st century [J].
Satava, RM .
ARCHIVES OF SURGERY, 1999, 134 (11) :1197-1202
[5]   Computer-aided placement of endosseous oral implants in patients after ablative tumour surgery: assessment of accuracy [J].
Wagner, A ;
Wanschitz, F ;
Birkfellner, W ;
Zauza, K ;
Klug, C ;
Schicho, K ;
Kainberger, F ;
Czerny, C ;
Bergmann, H ;
Ewers, R .
CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (03) :340-348
[6]   Evaluation of accuracy of computer-aided intraoperative positioning of endosseous oral implants in the edentulous mandible [J].
Wanschitz, F ;
Birkfellner, W ;
Watzinger, F ;
Schopper, C ;
Patruta, S ;
Kainberger, F ;
Figl, M ;
Kettenbach, J ;
Bergmann, H ;
Ewers, R .
CLINICAL ORAL IMPLANTS RESEARCH, 2002, 13 (01) :59-64