Performance specification for lower limb orthotic devices

被引:15
作者
Johnson, GR
Ferrarin, M
Harrington, M
Hermens, H
Jonkers, I
Mak, P
Stallard, J [1 ]
机构
[1] Robert Jones & Agnes Hunt Hosp, ORLAU, Oswestry SY10 7AG, Shrops, England
[2] Univ Newcastle, CREST, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
[3] Politecn Milan, ONLUS, Ctr Bioingn, Fdn Don Gnocchi, I-20148 Milan, Italy
[4] Nuffield Orthopaed Ctr, Oxford OX3 7LD, England
[5] Roessingh Res & Dev, NL-7522 AH Enschede, Netherlands
[6] Katholieke Univ Leuven, B-3001 Heverlee, Belgium
关键词
orthosis; gait analysis; inverse dynamic approach; specification; clinical evaluation;
D O I
10.1016/j.clinbiomech.2003.10.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. To establish the range of forces and moments applied to lower limb orthoses during ambulation by routine users. Design. Well-established gait analysis techniques were used to determine the loading at the major joints. It was assumed that the joint moments were transmitted by the orthosis encompassing any particular joint. Two hundred and five assessments of 164 patients were successfully completed by a consortium of four gait laboratories in Europe. The orthosis specification and patient clinical data were also recorded. Background. The design and development of orthoses has occurred largely by evolution rather than by formal engineering methods. In particular, formal design has been hampered by a lack of information oil the forces and moments applied during ambulation. Methods. A standard gait analysis procedure was employed to capture the data. In-house biomechanical models were used to calculate the joint loading. Data were normalised with respect to patient weight and leg length. Results. It was found that the median maximum normalised ankle moment transmitted by an ankle foot orthosis was 0.15 and the maximum knee moment was 0.09. The greatest moment transmitted by the hip joint of a hip knee ankle foot orthosis was also 0.09. There was a wide variation in the data due to differences in the impairments of the test subjects. Conclusion. It is possible to estimate the loads transmitted by an orthosis using established gait analysis procedures Without the need for load measurement transducers. There is now a need both to collect a larger representative dataset and to perform validation Studies with transducers. Relevance The methodology developed in this project has formed an important step in the development of standards and the incorporation of new materials and technologies into orthotic design. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:711 / 718
页数:8
相关论文
共 25 条
[1]  
American Academy of Orthopaedic Surgeons, 1975, ATL ORTH BIOM PRINC
[2]  
BECK JC, 1997, P 23 AAOP ANN M SCI, P41
[3]  
Butler PB, 1991, PHYSIOTHERAPY, V77, P81, DOI DOI 10.1016/S0031-9406(10)63580-8
[4]   GAIT ANALYSIS METHODOLOGY [J].
CAPPOZZO, A .
HUMAN MOVEMENT SCIENCE, 1984, 3 (1-2) :27-50
[5]   POSITION AND ORIENTATION IN-SPACE OF BONES DURING MOVEMENT - ANATOMICAL FRAME DEFINITION AND DETERMINATION [J].
CAPPOZZO, A ;
CATANI, F ;
DELLA CROCE, U ;
LEARDINI, A .
CLINICAL BIOMECHANICS, 1995, 10 (04) :171-178
[6]   Orthotic management of gait in spastic diplegia [J].
Carlson, WE ;
Vaughan, CL ;
Damiano, DL ;
Abel, MF .
AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 1997, 76 (03) :219-225
[7]   A GAIT ANALYSIS DATA-COLLECTION AND REDUCTION TECHNIQUE [J].
DAVIS, RB ;
OUNPUU, S ;
TYBURSKI, D ;
GAGE, JR .
HUMAN MOVEMENT SCIENCE, 1991, 10 (05) :575-587
[8]  
Erdemir A., 1999, P AM SOC BIOM PITTSB
[9]  
Evans C, 1994, DEV MED CHILD NE S70, V36, P26
[10]   ESTIMATION OF DEFORMATION IN A WALKING ORTHOSIS FOR PARAPLEGIC PATIENTS [J].
FERRARIN, M ;
STALLARD, J ;
PALMIERI, R ;
PEDOTTI, A .
CLINICAL BIOMECHANICS, 1993, 8 (05) :255-261