Development of a Faraday cup fast ion loss detector for keV beam ions

被引:16
作者
Lazerson, Samuel A. [1 ]
Ellis, Robert [1 ]
Freeman, Chris [1 ]
Ilagan, Jessica [1 ]
Wang, Tianyao [2 ]
Shao, Lin [2 ]
Allen, Nicole [1 ]
Gates, David [1 ]
Neilson, Hutch [1 ]
机构
[1] Princeton Plasma Phys Lab, Princeton, NJ 08542 USA
[2] Texas A&M Univ, Dept Nucl Engn, College Stn, TX 77845 USA
关键词
D O I
10.1063/1.5111714
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The development and testing of a Faraday cup fast-ion loss detector capable of measuring sub 100 keV particles is documented. Such measurement capabilities play an important role in the assessment of particle confinement of nuclear fusion experiments. The detector is manufactured using thin-film deposition techniques, building upon previous work using discrete foils. This new manufacturing method allows the form factor of the sensor to become that of essentially a microchip. Analysis of the diagnostic response is performed using Monte-Carlo particle simulations. These simulations show peaks in the detector response at 40 and 70 keV. The sensor is then tested in a tunable linear accelerator capable of accelerating protons from 20 to 120 keV. The detector response was found to be well matched to simulations. Improvements to the design to facilitate robustness are discussed.
引用
收藏
页数:5
相关论文
共 12 条
[1]  
[Anonymous], 1995, Python tutorial, technical report csr9526
[2]   Thin foil Faraday collectors as a radiation hard fast lost-ion diagnostic [J].
Cecil, FE ;
Aakhus-Witt, A ;
Hawbaker, J ;
Sayers, J ;
Bozek, A ;
Heidbrink, WW ;
Darrow, DS ;
Debey, TM ;
Marmar, E .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (03) :1747-1749
[3]   A thin film device as a low energy, high flux charged particle spectrometer [J].
Cecil, FE ;
Roy, B ;
Sutton, C ;
Wasinger, N .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (01) :792-795
[4]   Observation of alpha particle loss from JET plasmas during ion cyclotron resonance frequency heating using a thin foil Faraday cup detector array [J].
Darrow, D. S. ;
Cecil, F. E. ;
Kiptily, V. ;
Fullard, K. ;
Horton, A. ;
Murari, A. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2010, 81 (10)
[5]   Initial results from the lost alpha diagnostics on Joint European Torus [J].
Darrow, Doug ;
Baeumel, Stefan ;
Cecil, Ed ;
Ellis, Bob ;
Fullard, Keith ;
Hill, Ken ;
Horton, Alan ;
Kiptily, Vasily ;
Pedrick, Les ;
Reich, Matthias ;
Werner, Andreas .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (10)
[6]   Design and construction of a fast ion loss Faraday cup array diagnostic for Joint European Torus [J].
Darrow, DS ;
Bäumel, S ;
Cecil, FE ;
Kiptily, V ;
Ellis, R ;
Pedrick, L ;
Werner, A .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10) :3566-3568
[7]   ENERGETIC ION DISTRIBUTION RESULTING FROM NEUTRAL BEAM INJECTION IN TOKAMAKS [J].
GAFFEY, JD .
JOURNAL OF PLASMA PHYSICS, 1976, 16 (OCT) :149-169
[8]   Current status of the neutral beam heating system of W7-X [J].
McNeely, Paul ;
Barlak, Marek ;
Baldzuhn, Juergen ;
Bozhenkov, Sergey ;
Drevlak, Michael ;
Gawlik, Grzegorz ;
Heinemann, Bernd ;
Holtum, Dieter ;
Jagielski, Jacek ;
Kairys, Roland ;
Nocentini, Riccardo ;
Riedl, Rudolf ;
Rong, Peter ;
Rust, Norbert ;
Schroeder, Ralf ;
Speth, Eckehart ;
Staebler, Albrecht ;
Turos, Andrzej ;
Wolf, Robert .
FUSION ENGINEERING AND DESIGN, 2013, 88 (6-8) :1034-1037
[9]  
PyQT, 2012, PYQT REF GUID
[10]   Design of Faraday cup ion detectors built by thin film deposition [J].
Szalkowski, G. A. ;
Darrow, D. S. ;
Cecil, F. E. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 848 :87-90