Improved tolerance to missing data in myocardial perfusion SPET using OSEM reconstruction

被引:13
作者
Hatton, RL [1 ]
Hutton, BF
Angelides, S
Choong, KKL
Larcos, G
机构
[1] Westmead Hosp, Dept Med Phys, Sydney, NSW 2145, Australia
[2] Westmead Hosp, Dept Nucl Med & Ultrasound, Sydney, NSW 2145, Australia
[3] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2500, Australia
[4] Univ Sydney, Div Med, Sydney, NSW 2006, Australia
关键词
emission computed tomography; SPET; filtered back-projection; OSEM;
D O I
10.1007/s00259-003-1450-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
When projection data are incomplete for various technical reasons, artefacts may occur in the reconstructed images. This study examines whether an iterative reconstruction method, the ordered subsets implementation of the EM algorithm (OSEM), can improve reconstruction and minimise the artefacts compared to filtered back-projection (FBP). We varied the number and location of projections removed to investigate when significant artefacts occur, and whether diagnosis is affected. Phantom studies were analysed with sequential orthogonal pairs of projection angles removed (as would typically occur when either data loss or severe motion is detected during acquisition with a right-angled, dual-head cardiac single-photon emission tomography system) and reconstructed with both FBP and OSEM. Twelve normal myocardial perfusion studies were also assessed to study the effect of missing projections on clinical diagnosis. Differences between reconstructions with intact versus missing data were measured. Also, reconstructed images were clinically assessed and scored on a five-point scale based on whether the artefacts would alter clinical interpretation. Although both reconstruction methods showed artefacts, the absolute differences between reconstructed phantom data with intact and missing projection sets were significantly greater (P<0.005) for FBP than for OSEM for all numbers of missing projections. The clinical data showed similar differences between FBP and OSEM reconstructions. The three observers noted superiority of OSEM compared to FBP, with reduced incidence of clinically significant artefacts. However, neither reconstruction method could tolerate six or more missing pairs from 32 projections. There was no significant dependence on the angular location of missing projections. In the absence of any attempt to correct for missing projections, OSEM reduced the influence of artefacts compared to FBP.
引用
收藏
页码:857 / 861
页数:5
相关论文
共 11 条
[1]   ALGEBRAIC RECONSTRUCTION IN CT FROM LIMITED VIEWS [J].
ANDERSEN, AH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (01) :50-55
[2]  
BAGHAEI H, 2000, NUCL SCI S, V3
[3]  
DEPUEY EG, 1994, J NUCL MED, V35, P699
[4]  
GERMANO G, 1994, J NUCL MED, V35, P1193
[5]   RECONSTRUCTION OF OBJECTS FROM INCOMPLETE PROJECTIONS [J].
GORE, JC ;
LEEMAN, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01) :129-136
[6]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609
[7]   CONSTRAINED FOURIER SPACE METHOD FOR COMPENSATION OF MISSING DATA IN EMISSION COMPUTED-TOMOGRAPHY [J].
KARP, JS ;
MUEHLLEHNER, G ;
LEWITT, RM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (01) :21-25
[8]   Statistical image reconstruction in PET with compensation for missing data [J].
Kinahan, PE ;
Fessler, JA ;
Karp, JS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (04) :1552-1557
[9]  
LANGE K, 1984, J COMPUT ASSIST TOMO, V8, P306
[10]   ITERATIVE RECONSTRUCTION-REPROJECTION AND THE EXPECTATION-MAXIMIZATION ALGORITHM [J].
OLLINGER, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (01) :94-98