Application of nafion/cobalt hexacyanoferrate chemically modified electrodes for the determination of electroinactive cations by ion chromatography

被引:0
|
作者
Xu, JM [1 ]
Xian, YZ [1 ]
Shi, GY [1 ]
Li, JH [1 ]
Jin, LT [1 ]
机构
[1] E China Normal Univ, Dept Chem, Shanghai 200062, Peoples R China
关键词
ion chromatography; electrochemical detection; water analysis;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
An amperometric detector based on the chemical modification of Nafion and cobalt (II) hexacyanoferrate (II, III) thin Mm (Nafion/Co-CN-Fe) onto a glassy carbon (GC) electrode was firstly developed for the determination of electroinactive cations (Li+, Na+, K+, Rb+, Cs+ and NH4+) in single column ion chromatography. A set of well-defined peaks of electroinactive cation was obtained. The relative standard deviations (RSDs) of chromatographic peak height (nA) for these cations were all below 3.8%. The cations were detected conveniently in the linear concentration range of 6.0 x 10(-6)-5.0 x 10(-3) mol/L and their correlation coefficients were all above 0.99. The detection limits of the cations were 9.2 x 10(-6) mol/L for Li+, 3.4 x 10(-6) mol/L for Na+, 6.3 x 10(-7) mol/L for K+, 7.8 x 10(-7) mol/L for Rb+, 6.2 x 10(-7) mol/L for Cs+ and 6.2 x 10(-6) mol/L for NW, at a signal-noise ratio of 3. The method was quick, sensitive, simple and was successfully applied to the analysis of rainwater samples. The electrode was stable for a 2 week period of operation with no evidence of chemical or mechanical deterioration.
引用
收藏
页码:771 / 776
页数:6
相关论文
共 50 条
  • [1] Application of Nafion/Cobalt Hexacyanoferrate Chemically Modified Electrodes for the Determination of Electroinactive Cations by Ion Chromatography
    徐继明
    鲜跃仲
    施国跃
    李金花
    金利通
    ChineseJournalofChemistry, 2002, (08) : 771 - 776
  • [2] Amperometric detection studies of Nafion/indium hexacyanoferrate film for the determination of electroinactive cations in ion chromatography
    Q. Xu
    S. Zhang
    W. Zhang
    L. Jin
    K. Tanaka
    H. Haraguchi
    A. Itoh
    Fresenius' Journal of Analytical Chemistry, 2000, 367 : 241 - 245
  • [3] Amperometric detection studies of Nafion/indium hexacyanoferrate film for the determination of electroinactive cations in ion chromatography
    Xu, Q
    Zhang, S
    Zhang, W
    Jin, LT
    Tanaka, K
    Haraguchi, H
    Itoh, A
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 367 (03): : 241 - 245
  • [4] Determination of glutathione in vivo by microdialysis using liquid chromatography with a cobalt hexacyanoferrate chemically modified electrode
    Shi, GY
    Lu, JX
    Xu, F
    Sun, WL
    Jin, LT
    Yamamoto, K
    Tao, SG
    Jin, JY
    ANALYTICA CHIMICA ACTA, 1999, 391 (03) : 307 - 313
  • [5] ELECTRODES MODIFIED WITH COBALT HEXACYANOFERRATE
    JOSEPH, J
    GOMATHI, H
    RAO, GP
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1991, 304 (1-2): : 263 - 269
  • [6] MULTIFUNCTIONAL CHEMICALLY MODIFIED ELECTRODES WITH MIXED COBALT PHTHALOCYANINE NAFION COATINGS
    WANG, J
    LI, R
    TALANTA, 1989, 36 (1-2) : 279 - 284
  • [7] ELECTROCHEMICAL DETERMINATION OF ELECTROINACTIVE MEDICINES WITH NAFION-COATED ELECTRODES
    MATSUE, T
    AOKI, A
    UCHIDA, I
    OSA, T
    CHEMISTRY LETTERS, 1987, (05) : 957 - 958
  • [8] Electrooxidation and determination of dopamine using a Nafion®-cobalt hexacyanoferrate film modified electrode
    Castro, Suely S. L.
    Mortimer, Roger J.
    de Oliveira, Marcelo F.
    Stradiotto, Nelson R.
    SENSORS, 2008, 8 (03) : 1950 - 1959
  • [9] Simultaneous amperometric detection of electroinactive anions and cations in ion chromatography
    Xu, Q
    Xu, C
    Wang, YP
    Zhang, W
    Jin, LT
    Tanaka, K
    Haraguchi, H
    Itoh, A
    ANALYST, 2000, 125 (10) : 1799 - 1804
  • [10] Electrocatalytic oxidation of thiocholine at chemically modified cobalt hexacyanoferrate screen-printed electrodes
    Arduini, Fabiana
    Cassisi, Alessio
    Amine, Aziz
    Ricci, Francesco
    Moscone, Danila
    Palleschi, Giuseppe
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2009, 626 (1-2) : 66 - 74