Fundamental limits of scintillation detector timing precision

被引:68
作者
Derenzo, Stephen E. [1 ]
Choong, Woon-Seng [1 ]
Moses, William W. [1 ]
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Natl Lab, Div Life Sci, Berkeley, CA 94720 USA
关键词
scintillator; timing precision; timing resolution; Monte Carlo; photodetector; scintillation detector; TIME-RESOLUTION; DEPTH;
D O I
10.1088/0031-9155/59/13/3261
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper we review the primary factors that affect the timing precision of a scintillation detector. Monte Carlo calculations were performed to explore the dependence of the timing precision on the number of photoelectrons, the scintillator decay and rise times, the depth of interaction uncertainty, the time dispersion of the optical photons (modeled as an exponential decay), the photodetector rise time and transit time jitter, the leading-edge trigger level, and electronic noise. The Monte Carlo code was used to estimate the practical limits on the timing precision for an energy deposition of 511 keV in 3 mm x 3 mm x 30 mm Lu2SiO5:Ce and LaBr3:Ce crystals. The calculated timing precisions are consistent with the best experimental literature values. We then calculated the timing precision for 820 cases that sampled scintillator rise times from 0 to 1.0 ns, photon dispersion times from 0 to 0.2 ns, photodetector time jitters from 0 to 0.5 ns fwhm, and A from 10 to 10 000 photoelectrons per ns decay time. Since the timing precision R was found to depend on A(-1/2) more than any other factor, we tabulated the parameter B, where R = BA(-1/2). An empirical analytical formula was found that fit the tabulated values of B with an rms deviation of 2.2% of the value of B. The theoretical lower bound of the timing precision was calculated for the example of 0.5 ns rise time, 0.1 ns photon dispersion, and 0.2 ns fwhm photodetector time jitter. The lower bound was at most 15% lower than leading-edge timing discrimination for A from 10 to 10 000 photoelectrons ns(-1). A timing precision of 8 ps fwhm should be possible for an energy deposition of 511 keV using currently available photodetectors if a theoretically possible scintillator were developed that could produce 10 000 photoelectrons ns(-1).
引用
收藏
页码:3261 / 3286
页数:26
相关论文
共 50 条
  • [1] On the Fundamental Limits of the Position Resolution of Position-Sensitive Scintillation Detector Using Continuous Crystal
    Gal, J.
    Imrek, J.
    Kalinka, G.
    Kiraly, B.
    Nagy, F.
    Nyako, B. M.
    Molnar, J.
    Valastyan, I.
    Balkay, L.
    Kis, S. A.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3684 - 3689
  • [2] Scintillation Detector Timing Resolution; A Study by Ray Tracing Software
    ter Weele, David N.
    Schaart, Dennis R.
    Dorenbos, Pieter
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (05) : 1972 - 1980
  • [3] Systematic studies of scintillation detector with timing resolution of 10 ps for heavy ion beam
    Nishimura, S
    Kurata-Nishimura, M
    Morimoto, K
    Nishi, Y
    Ozawa, A
    Yamaguchi, T
    Ohnishi, T
    Zheng, T
    Chiba, M
    Tanihata, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 510 (03) : 377 - 388
  • [5] Intrinsic Energy Resolution of a Scintillation Detector
    Samedov, V. V.
    PHYSICS OF ATOMIC NUCLEI, 2021, 84 (11) : 1828 - 1835
  • [6] Intrinsic Energy Resolution of a Scintillation Detector
    V. V. Samedov
    Physics of Atomic Nuclei, 2021, 84 : 1828 - 1835
  • [7] Lightguide influence on timing response of scintillation detectors
    Bash, Z. Binaei
    Ghal-Eh, N.
    Bayat, E.
    Etaati, G. R.
    RADIATION PHYSICS AND CHEMISTRY, 2013, 82 : 1 - 6
  • [8] Fast Timing Silicon Photomultipliers for Scintillation Detectors
    Yeom, Jung Yeol
    Vinke, Ruud
    Pavlov, Nikolai
    Bellis, Stephen
    Wall, Liam
    O'Neill, Kevin
    Jackson, Carl
    Levin, Craig S.
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2013, 25 (14) : 1309 - 1312
  • [9] Ultrafast Waveguiding Quantum Dot Scintillation Detector
    Dropiewski, K.
    Minns, A.
    Yakimov, M.
    Tokranov, V.
    Murat, P.
    Oktyabrsky, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 954
  • [10] Heterogeneous γ-ray scintillation detector
    V. A. Terekhin
    Yu. I. Chernukhin
    S. I. Strel’tsov
    Atomic Energy, 2006, 101 : 583 - 587