Fundamental limits of scintillation detector timing precision

被引:70
作者
Derenzo, Stephen E. [1 ]
Choong, Woon-Seng [1 ]
Moses, William W. [1 ]
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Natl Lab, Div Life Sci, Berkeley, CA 94720 USA
关键词
scintillator; timing precision; timing resolution; Monte Carlo; photodetector; scintillation detector; TIME-RESOLUTION; DEPTH;
D O I
10.1088/0031-9155/59/13/3261
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper we review the primary factors that affect the timing precision of a scintillation detector. Monte Carlo calculations were performed to explore the dependence of the timing precision on the number of photoelectrons, the scintillator decay and rise times, the depth of interaction uncertainty, the time dispersion of the optical photons (modeled as an exponential decay), the photodetector rise time and transit time jitter, the leading-edge trigger level, and electronic noise. The Monte Carlo code was used to estimate the practical limits on the timing precision for an energy deposition of 511 keV in 3 mm x 3 mm x 30 mm Lu2SiO5:Ce and LaBr3:Ce crystals. The calculated timing precisions are consistent with the best experimental literature values. We then calculated the timing precision for 820 cases that sampled scintillator rise times from 0 to 1.0 ns, photon dispersion times from 0 to 0.2 ns, photodetector time jitters from 0 to 0.5 ns fwhm, and A from 10 to 10 000 photoelectrons per ns decay time. Since the timing precision R was found to depend on A(-1/2) more than any other factor, we tabulated the parameter B, where R = BA(-1/2). An empirical analytical formula was found that fit the tabulated values of B with an rms deviation of 2.2% of the value of B. The theoretical lower bound of the timing precision was calculated for the example of 0.5 ns rise time, 0.1 ns photon dispersion, and 0.2 ns fwhm photodetector time jitter. The lower bound was at most 15% lower than leading-edge timing discrimination for A from 10 to 10 000 photoelectrons ns(-1). A timing precision of 8 ps fwhm should be possible for an energy deposition of 511 keV using currently available photodetectors if a theoretically possible scintillator were developed that could produce 10 000 photoelectrons ns(-1).
引用
收藏
页码:3261 / 3286
页数:26
相关论文
共 20 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   A Comprehensive & Systematic Study of Coincidence Time Resolution and Light Yield Using Scintillators of Different Size and Wrapping [J].
Auffray, E. ;
Frisch, B. ;
Geraci, F. ;
Ghezzi, A. ;
Gundacker, S. ;
Hillemanns, H. ;
Jarron, P. ;
Meyer, T. ;
Paganoni, M. ;
Pauwels, K. ;
Pizzichemi, M. ;
Lecoq, P. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (05) :3163-3171
[3]   The timing resolution of scintillation-detector systems: Monte Carlo analysis [J].
Choong, Woon-Seng .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (21) :6495-6513
[4]   Measurements of the intrinsic rise times of common inorganic scintillators [J].
Derenzo, SE ;
Weber, MJ ;
Moses, WW ;
Dujardin, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (03) :860-864
[5]   Experimental and theoretical studies of donor-acceptor scintillation from PbI2 [J].
Derenzo, Stephen E. ;
Bourret-Courchesne, Edith ;
Yan, Zewu ;
Bizarri, Gregory ;
Canning, Andrew ;
Zhang, Gaigong .
JOURNAL OF LUMINESCENCE, 2013, 134 :28-34
[6]   Effects of Ce concentration on scintillation properties of LaBr3:Ce [J].
Glodo, J ;
Moses, WW ;
Higgins, WM ;
van Loef, EVD ;
Wong, P ;
Derenzo, SE ;
Weber, MJ ;
Shah, KS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) :1805-1808
[7]   Concentration Effects in Eu Doped SrI2 [J].
Glodo, Jarek ;
van Loef, Edgar V. ;
Cherepy, Nerine J. ;
Payne, Stephen A. ;
Shah, Kanai S. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (03) :1228-1232
[8]   Time of flight positron emission tomography towards 100ps resolution with L(Y)SO: an experimental and theoretical analysis [J].
Gundacker, S. ;
Auffray, E. ;
Frisch, B. ;
Jarron, P. ;
Knapitsch, A. ;
Meyer, T. ;
Pizzichemi, M. ;
Lecoq, P. .
JOURNAL OF INSTRUMENTATION, 2013, 8
[9]   BARIUM FLUORIDE INORGANIC SCINTILLATOR FOR SUBNANOSECOND TIMING [J].
LAVAL, M ;
MOSZYNSKI, M ;
ALLEMAND, R ;
CORMORECHE, E ;
GUINET, P ;
ODRU, R ;
VACHER, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 206 (1-2) :169-176
[10]   New Approaches to Improve Timing Resolution in Scintillators [J].
Lecoq, P. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (05) :2313-2318