Abnormal foreshortening of a Flow Re-Direction Endoluminal Device caused by in-stent thrombosis immediately after deployment br

被引:0
作者
Otsuka, Takafumi [1 ]
Izumi, Takashi [1 ,2 ]
Nishihori, Masahiro [1 ]
Tsukada, Tetsuya [1 ]
Goto, Shunsaku [1 ]
Ikezawa, Mizuka [1 ]
Kato, Naoki [1 ]
Nakano, Mizuki [1 ]
Uda, Kenji [1 ]
Yokoyama, Kinya [1 ]
Araki, Yoshio [1 ]
Saito, Ryuta [1 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Neurosurg, Nagoya, Japan
[2] Nagoya Univ, Grad Sch Med, Dept Neurosurg, 65 Tsurumai cho Showa ku, Nagoya 4668550, Japan
关键词
flow diverter; Flow Re -Direction Endoluminal Device; thromboembolic complications; INTRACRANIAL ANEURYSMS; PLATELET; EXPERIENCE;
D O I
10.18999/nagjms.84.4.884
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Thromboembolic complications are a concern in the treatment of cerebral aneurysms using a flow diverter. In this study, we report a case of abnormal foreshortening of a Flow Re-Direction Endoluminal Device (FRED) caused by in-stent thrombosis immediately after its deployment. A 72-year-old woman had a large cavernous carotid aneurysm, which caused ptosis and diplopia. FRED deployment was planned, and dual antiplatelet therapy was initiated 2 weeks before the procedure. Under systemic heparinization, FRED was deployed with local compaction over the aneurysm orifice. Cone-beam computed tomography subsequently revealed slightly poor wall apposition at the proximal side. While the balloon catheter was prepared for angioplasty, the stent became abnormally foreshortened, the proximal side slipped into the aneurysm, and the internal carotid artery became occluded. FRED was removed using a snare wire, and recanalization was obtained. The lumen of the removed FRED was filled with thrombus. The antiplatelet therapy was changed to triple regimen, and a Pipeline Flex embolization device was placed 1 month later. At that time, no thromboembolic complications were noted. It was considered that thrombotic occlusion was followed by foreshortening of FRED on the distal side because of antegrade blood flow. Multiple factors, such as increased mesh density by locally compacted stent deployment, slightly poor wall apposition, clopidogrel resistance, and the dual-layer structure of FRED, may have been involved in thrombus formation.
引用
收藏
页码:884 / 889
页数:6
相关论文
共 13 条
  • [1] Predictive value of platelet reactivity unit (PRU) value for thrombotic and hemorrhagic events during flow diversion procedures: a meta-analysis
    Ajadi, Ebunoluwa
    Kabir, Shaowli
    Cook, Aaron
    Grupke, Stephen
    Alhajeri, Abdulnasser
    Fraser, Justin F.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2019, 11 (11) : 1123 - +
  • [2] Increased risk in patients with high platelet amregation receiving chronic clopidogrel therapy undergoing percutaneous coronary intervention - Is the current antiplatelet therapy adequate?
    Bliden, Kevin P.
    DiChiara, Joseph
    Tantry, Udaya S.
    Bassi, Ashwani K.
    Chaganti, Srivasavi K.
    Gurbel, Paul A.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (06) : 657 - 666
  • [3] Immediate procedural outcomes in 44 consecutive Pipeline Flex cases: the first North American single-center series
    Colby, Geoffrey P.
    Lin, Li-Mei
    Caplan, Justin M.
    Jiang, Bowen
    Huang, Judy
    Tamargo, Rafael J.
    Coon, Alexander L.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (07) : 702 - 709
  • [4] TreatmentTreatment of Intra- and Extracranial Aneurysms Using the Flow-Redirection Endoluminal Device: Multicenter Experience and Follow-Up Results
    Drescher, F.
    Weber, W.
    Berlis, A.
    Rohde, S.
    Carolus, A.
    Fischer, S.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2017, 38 (01) : 105 - 112
  • [5] Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model
    Girdhar, Gaurav
    Andersen, Arielle
    Pangerl, Elizabeth
    Jahanbekam, Reza
    Ubl, Samantha
    Nguyen, Kevin
    Wainwright, John
    Wolf, Michael F.
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2018, 106 (12) : 3195 - 3202
  • [6] Flow Re-direction Endoluminal Device in treatment of cerebral aneurysms: initial experience with short-term follow-up results
    Kocer, Naci
    Islak, Civan
    Kizilkilic, Osman
    Kocak, Burak
    Saglam, Muzaffer
    Tureci, Ercan
    [J]. JOURNAL OF NEUROSURGERY, 2014, 120 (05) : 1158 - 1171
  • [7] The FRED Flow-Diverter Stent for Intracranial Aneurysms: Clinical Study to Assess Safety and Efficacy
    Moehlenbruch, M. A.
    Herweh, C.
    Jestaedt, L.
    Stampfl, S.
    Schoenenberger, S.
    Ringleb, P. A.
    Bendszus, M.
    Pham, M.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (06) : 1155 - 1161
  • [8] FRED Italian Registry: a multicenter experience with the flow re-direction endoluminal device for intracranial aneurysms
    Piano, Mariangela
    Valvassori, Luca
    Lozupone, Emilio
    Pero, Guglielmo
    Quilici, Luca
    Boccardi, Edoardo
    [J]. JOURNAL OF NEUROSURGERY, 2020, 133 (01) : 174 - 181
  • [9] Initial experience with implantation of novel dual layer flow-diverter device FRED
    Poncyljusz, Wojciech
    Sagan, Leszek
    Safranow, Krzysztof
    Rac, Monika
    [J]. VIDEOSURGERY AND OTHER MINIINVASIVE TECHNIQUES, 2013, 8 (03) : 258 - 264
  • [10] Thromboembolic complications with Pipeline Embolization Device placement: impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value
    Tan, Lee A.
    Keigher, Kiffon M.
    Munich, Stephan A.
    Moftakhar, Roham
    Lopes, Demetrius K.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2015, 7 (03) : 217 - 221