Development of a System for Real-Time Monitoring of Pressure, Temperature, and Humidity in Casts

被引:7
作者
Tuan, Chiu-Ching [1 ]
Lu, Chi-Heng [1 ,2 ]
Wu, Yi-Chao [3 ]
Yeh, Wen-Ling [4 ]
Chen, Mei-Chuan [4 ]
Lee, Tsair-Fwu [5 ]
Chen, Yu-Jing [1 ]
Kao, Hsuan-Kai [4 ,6 ,7 ]
机构
[1] Natl Taipei Univ Technol, Grad Inst Elect Engn, Taipei 10608, Taiwan
[2] Chang Gung Mem Hosp Linkou, Dept Radiat Oncol, Linkou 33305, Taiwan
[3] Natl Taitung Univ, Interdisciplinary Program Green & Informat Techno, Taitung 95092, Taiwan
[4] Chang Gung Mem Hosp Linkou, Dept Orthoped Surg, Linkou 33305, Taiwan
[5] Natl Kaohsiung Univ Sci & Technol, Dept Elect Engn, Med Phys & Informat Lab, Kaohsiung 80778, Taiwan
[6] Chang Gung Mem Hosp, Bone & Joint Res Ctr, Linkou 33305, Taiwan
[7] Chang Gung Univ, Coll Med, Taoyuan 33305, Taiwan
关键词
cast fixation; fractures; swathing a cast; reliable reference data; nonlinear error; human interface; soft skin cushioning; cast wrinkles; APP; FRACTURES; BENEATH;
D O I
10.3390/s19102417
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cast fixation is a general clinical skill used for the treatment of fractures. However, it may cause many complications due to careless treatment procedures. Currently, swathing a cast for a patient can only be determined by a doctors' experience; however, this cannot be determined by the value of pressure, temperature, or humidity with objective and reliable equipment. When swathing a cast for a patient, the end result is often too tight or too loose. Hence, in this paper we developed a sensor for detecting pressure, temperature, and humidity, respectively. This could provide reliable reference cast data to help physicians to understand the tightness of cast swathing and to adjust the tightness of cast swathing instantly to alleviate a patient's complications caused by excessive pressure or overheating. In this paper, six pressure sensors and one temperature-humidity sensor are used to detect the pressure, temperature, and humidity in an arm swathed with a cast to confirm whether the tightness of the cast is fixing the fracture efficiently, while avoiding causing any damage by using excessive pressure. Currently, the variation in temperature and humidity can be detected by the inflammation of the wound, displaying secretions, and fever in the cast. Based on the experiments, the voltage and power conversion coefficients of the developed sensors could be compensated for by the nonlinear error of the sensor. The experimental results could be instantly displayed on a human interface, such as a smart mobile device. The average skin pressure in a swathed cast was 12.14 g and ranged from 5.0 g to 17.5 g. A few casts exceeded 37.50 g. The abnormal pressure of wrinkles produced during swathing a cast often ranged from 22.50 g to 38.75 g. This shows that cast wrinkles cause pressure on the skin. The pressure caused by cast wrinkles on bone protrusions ranged from 56.5 g to 84.4 g. Compared to other parts that lacked soft skin cushioning, the pressure of cast wrinkles that occurred in the ulna near the protrusion of the wrist bone increased averagely. The pressure error value was less than 2%, the temperature error was less than 1%, and the humidity error was less than 5%. Therefore, they were all in line with the specifications of commercially available products. The six pressure detection points and one temperature and humidity detection point in our newly designed system can accurately measure the pressure, temperature, and humidity inside the cast, and instantly display the corresponding information by mobile APP. Doctors receive reliable reference data and are instantly able to understand the tightness of the swathed cast and adjust it at any time to avoid complications caused by pressure or overheating due to excessive pressure.
引用
收藏
页数:12
相关论文
共 19 条
[1]   Are Orthopaedic Residents Competent at Performing Basic Nonoperative Procedures in an Unsupervised Setting? A "Pop Quiz" of Casting, Knee Arthrocentesis, and Pressure Checks for Compartment Syndrome [J].
Abzug, Joshua M. ;
O'Toole, Robert V. ;
Paryavi, Ebrahim ;
Sterling, Robert .
JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2016, 36 (01) :E10-E13
[2]  
[Anonymous], 2000, ARCHITECTURAL STYLES
[3]   Old and new risk factors for upper extremity deep venous thrombosis [J].
Blom, JW ;
Doggen, CJM ;
Osanto, S ;
Rosendaal, FR .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2005, 3 (11) :2471-2478
[4]  
Chen Y., 2018, THESIS
[5]   Effect of Pressure Applied During Casting on Temperatures Beneath Casts [J].
Deignan, Brian J. ;
Iaquinto, Joseph M. ;
Eskildsen, Scott M. ;
Woodcock, Cassandra A. ;
Owen, John R. ;
Wayne, Jennifer S. ;
Kuester, Victoria G. .
JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2011, 31 (07) :791-797
[6]  
DiPaola MJ, 2014, J PEDIATR ORTHOPED, V34, P643, DOI 10.1097/BPO.0000000000000191
[7]   Driving with a forearm plaster cast: patients' perspective [J].
Edwards, M. R. ;
Oliver, M. C. ;
Hatrick, N. C. .
EMERGENCY MEDICINE JOURNAL, 2009, 26 (06) :405-406
[8]  
Google, FAST LIGHTWEIGHT KEY
[9]   A comparison of the functional and radiological results of Paris plaster cast and ulnar gutter splint in the conservative treatment of fractures of the fifth metacarpal [J].
Gulabi D. ;
Avci C.C. ;
Cecen G.S. ;
Bekler H.I. ;
Saglam F. ;
Merih E. .
European Journal of Orthopaedic Surgery & Traumatology, 2014, 24 (7) :1167-1173
[10]   Factors contributing to the temperature beneath plaster or fiberglass cast material [J].
Hutchinson M.J. ;
Hutchinson M.R. .
Journal of Orthopaedic Surgery and Research, 3 (1)